Abstract
The performance of three commercially available surface particle counters (SPCs)—a Tencor Surfscan 4000, an Aeronca WIS 200, and an Inspex EX20/20—was evaluated using latex spheres of 0.3–10 μm in diameter. The relationship between scattering cross section and sphere size was determined for each of the SPCs. In addition, the ratio between the numbers of particles detected by the Inspex and the Tencor was shown to decrease from 1 to 0.4 for increasing sphere size whereas the ratio for the Aeronca and the Tencor was determined to be 1.09(5). Interference effects for latex spheres of various diameters and significant variations of the apparent sphere size over the wafer surface have also been observed. Both phenomena are caused by the spherical shape and transparency of the spheres in combination with the specific designs of the SPC detection systems.