Deep Contrast and Companion Detection Using the EvWaCo Test Bed Equipped with an Achromatic Focal Plane Mask and an Adjustable Inner Working Angle

, , , , , , , , , , , , , , , , , , , , , , , and

Published 2021 April 7 © 2021. The American Astronomical Society. All rights reserved.
, , Citation Mary Angelie Alagao et al 2021 AJ 161 208 DOI 10.3847/1538-3881/abe709

Download Article PDF
DownloadArticle ePub

You need an eReader or compatible software to experience the benefits of the ePub3 file format.

1538-3881/161/5/208

Abstract

The evanescent wave coronagraph uses the principle of frustrated total internal reflection (FTIR) to suppress the light coming from the star and study its close environment. Its focal plane mask is composed of a lens and a prism placed in contact with each other to produce the coronagraphic effect. In this paper, we present the experimental results obtained using an upgraded focal plane mask of the Evanescent Wave Coronagraph (EvWaCo). These experimental results are also compared to the theoretical performance of the coronagraph obtained through simulations. Experimentally, we reach a raw contrast equal to a few 10−4 at a distance equal to 3 λ/D over the full I band (λc = 800 nm, Δλ/λ ≈ 20%) and equal to 4 λ/D over the full R band (λc = 650 nm, Δλ/λ ≈ 23%) in unpolarized light. However, our simulations show a raw contrast close to 10−4 over the full I band and R band at the same distance, thus confirming the theoretical achromatic advantage of the coronagraph. We also verify the stability of the mask through a series of contrast measurements over a period of 8 months. Furthermore, we measure the sensitivity of the coronagraph to the lateral and longitudinal misalignment of the focal plane mask and to the lateral misalignment of the Lyot stop.

Export citation and abstract BibTeX RIS

1. Introduction

High-contrast imaging of exoplanets is essential in gaining a better understanding of the formation and evolution of planetary systems. With enough signal from the planet, high-contrast imaging reveals not only the spectrophotometry of its atmosphere but also the different interactions in the close environment of the star (Traub & Oppenheimer 2010; Mawet et al. 2012; Milli et al. 2016). However, direct imaging of exoplanets is a highly challenging goal because of the very high contrast and the very small angular separation existing between the planet and its parent star. In fact, contrasts range from C ≈ 10−3 for hot giant planets in the infrared to C ≈ 10−10 for Earth-like planets in the visible (Mawet et al. 2012), and a contrast between 10−6 and 10−7 in the near-IR is required to image warm Jupiter-like planets (Burrows et al. 1997). The typical angular separation is equal to 01 for Earth-like planets and 1'' for Jupiter-like planets, pertaining to a distance of 1 au, seen at 10 pc.

Regarding those two constraints, instrumental issues such as diffraction and scattering have serious adverse effects on the detection. To be able to measure such contrasts, it is required to eliminate diffraction effects and to limit (and control) wave-front errors.

The most recent ground-based coronagraphs installed on 8–10 m class telescopes combined a high-contrast performance coronagraph and a high-order adaptive optics to simultaneously (i) mask the central object, (ii) eliminate the diffraction effects, and (iii) control the wave-front errors. For example, the Spectro-Polarimetric High-Contrast Exoplanet Research (SPHERE) installed on the VLT obtained a contrast of a few 10−7 for a distance between 10 λ/D (05) and 30 λ/D (16) from the optical axis for a wavelength of 2.11 μm (K band). These numerical figures are based on the results of the high-contrast imaging of Sirius A (Vigan et al. 2015; Beuzit et al. 2019). Similarly, the measured 5σ contrasts during the first light of the Gemini Planet Imager (GPI), which is installed on the Gemini South Telescope in Chile, were 10−6 at 30 λ/D (12) and 10−5 at 8 λ/D (03) in the H band (λmid = 1.6 μm; Macintosh et al. 2014). In this framework, the aim of the EvWaCo project is to develop a new kind of simple and cost-effective coronagraph for ground-based telescopes with state-of-the-art performance comparable to the ones in SPHERE and GPI.

The Center for Optics and Photonics at the National Astronomical Research Institute of Thailand (NARIT) has developed a test bed to demonstrate in laboratory conditions the capability of the Evanescent Wave Coronagraph (EvWaCo) to reach comparable results at optical wavelengths (Alagao et al. 2017; Buisset et al. 2017). The suppression of starlight in EvWaCo is achieved by placing in contact a prism and a lens at the telescope focal plane. The prism is oriented at an angle of incidence such that light should undergo total internal reflection. However, when a glass medium, such as a lens, is placed in contact at the focus, the beam is transmitted by means of captured evanescent waves; this phenomenon is known as "frustrated total Internal reflection (FTIR)" (Rabbia 2003; Zhu et al. 1986; Gross 2005). In the case of an off-axis beam, the beam undergoes total internal reflection as soon as an air gap is met. This implies that with proper shaping of the materials in contact, the rejection profile can be adjusted. Furthermore, from the properties of the evanescent waves, we can theoretically show that the size of the mask inherently adapts itself to the wavelength, that is, the longer the wavelength, the larger the rejection profile (Alagao et al. 2017; Buisset et al. 2017). With these capabilities, we expect that the EvWaCo focal plane mask (FPM) will provide an achromatic rejection.

In its current form, the EvWaCo test bed comprises off-the-shelf components mounted on a passive support. This setup operates at ambient temperature without an adaptive optics setup to correct for the wave-front distortions induced by the optical quality of the components and by inhomogeneities of the air refractive index. In our previous paper (Buisset et al. 2018), we presented the performance of EvWaCo in unpolarized broadband light over the full I band (700 nm–900 nm) with a central wavelength λc = 800 nm and of spectral ratio Δλ/λc ≈ 25%. We used a large Lyot stop of diameter equal to 78% of the aperture stop, thus yielding a throughput close to 60%.

The measured contrast performance was a few 10−6 at distances from the optical axis in the focal plane between 10 λ/D and 20 λ/D, and a few 10−7 between 20 λ/D and 50 λ/D. The inner working angle (IWA) was close to 6 Airy radii along the x-axis and to 8 Airy radii along the y-axis. We have also demonstrated that the signal obtained from a companion 30,000 times fainter than the star, located at a distance equal to 15 λ/D from the center of the star image, can be detected with a signal-to-noise ratio (S/N) equal to 75 (Buisset et al. 2018).

2. The Evanescent Wave Coronagraph Project

These results lay the groundwork for the formation of the EvWaCo Project. The EvWaCo project is part of the Thai Space Consortium that was established in 2018 and aims to develop the human capacities, facilities, and technology for space observations and exploration in Thailand. In this framework, the EvWaCo project aims to develop state-of-the-art technologies for high-contrast observation with the medium- and long-term objectives of providing subsystems (coronagraphs, FPMs, masks, adaptive optics, or spectrographs) for future space missions dedicated to the direct detection and characterization of exoplanets.

Under this project, an EvWaCo prototype equipped with an adaptive optics system will be installed on the 2.4 m Thai National Telescope at horizon 2022. The prototype is currently under development at the NARIT Center for Optics and Photonics with the support of Centre de Recherche Astrophysique de Lyon (CRAL), Institut d'Optique Graduate School (IOGS), and Laboratoire Hubert Curien (LabHC). It is specified to observe in the R and I bands with a raw contrast of 10−4 at an IWA of 05. The limiting magnitude of the central object is 8. This coronagraph is set to observe the science cases briefly discussed hereafter.

The coronagraph will be used to directly image the circumstellar disks and other fine structures hidden in the bright stellar light. Among the suitable targets are the young stellar objects and pre-main-sequence stars (Watson et al. 2007; Tamura 2016). Another would be the direct imaging of brown dwarf binaries and M dwarfs. The brown dwarf binary HD 130948 BC was detected in the I band using the lucky imaging instrument, FastCam, installed at the 2.5 m Nordic telescope with an S/N ≈ 9. The binary stars were separated by 2561 ± 0007 (46.5 au) with a magnitude of ΔI ≈ 11.300.11 (C ≈ 3.10−5; Labadie et al. 2011). In the same spectral band, high-resolution images of 490 mid- to late-M dwarfs were also obtained using the FastCam installed at a 1.5 m telescope (Cortes-Contreras et al. 2017). These detections have been obtained without a coronagraph. Using a coronagraph would significantly increase the capability of detecting brown dwarf binaries and M dwarfs closer to the star. In addition, these targets were detected at the current observing wavelength of the EvWaCo.

Moreover, taking advantage of EvWaCo's capability to collect the light coming from both the star and the companion, the coronagraph will be used to validate the candidate host stars with a stellar companion or chance-aligned background star in the Kepler (Coughlin et al. 2016; Thompson et al. 2018), K2 (Yu et al. 2018; Mayo et al. 2018), and TESS (Stassun et al. 2018, 2019) surveys. Many of the transit-like signals are false positives (Morton et al. 2016). The false-positive scenarios are considered to be blended and unblended eclipsing binary stars, where the eclipsing binary may be physically associated with a primary target or maybe a chance alignment with the primary target. Follow-up photometric observations can be done to validate the contaminating sources.

The aim of this paper is to present both the experimental results and the simulations of the coronagraph. The experimental results are obtained using the new achromatic FPM with an adjustable IWA. The FPM comprises a dedicated optomechanical support designed to adjust the contact between the lens and the prism, thus modifying the shape of the mask and its rejection profile.

This paper is outlined as follows. We present an overview of the EvWaCo test bed in Section 3. This test bed has been upgraded to accommodate the capabilities of the new FPM. Section 4 presents mask reflection at varying pressure, and the experimental and simulation results for the mask reflection at two wavelengths: λ1 = 800 nm and λ2 = 650 nm. We also present the pupil irradiance distribution in this section. The raw contrast calculation is presented in Section 5. We also illustrate the contrast performance of EvWaCo by showing the gain provided by this coronagraph for the observation of a companion 25,000 times fainter than the star and located at a distance equal to 5.5 λ/D with an S/N equal to 11. In Section 6, we present the repeatability of the coronagraph performance in terms of the contrast and the IWA. Section 7 is dedicated to results obtained for determining the sensitivity of the coronagraph to the critical components such as the lateral and longitudinal misalignment of the FPM, and the lateral misalignment of the Lyot stop. In Section 8, we briefly discuss the gain in these recent experiments compared to the previous papers presented. Finally, we conclude the paper with a summary of the test-bed results and a description of our future plans toward the development of a full system that will include adaptive optics for ground-based observations with large and very large telescopes, and couple it with a spectrograph.

3. EvWaCo Focal Plane Mask

3.1. The Setup

Figure 1 shows an illustration of the current EvWaCo test bed. The setup comprises two sources: the star source and the companion source. The star source is made of a stable tungsten lamp that emits a collimated beam and is focused by an achromatic doublet lens on a 50 μm multimode fiber. This multimode fiber is connected to a single-mode fiber that directs the light toward the coronagraph bench.

Figure 1.

Figure 1. Illustration of the EvWaCo setup with the star and companion channels.

Standard image High-resolution image

The companion source comprises an LED that emits a beam at the central wavelength λc = 780 nm with a spectral bandwidth of Δλ/λ ≈ 3%. The light emitted by this LED is injected into a single-mode fiber. The beams emitted by the star and the companion fibers are recombined by a beam splitter BS, collimated by the lens L1, and transmitted through the aperture stop AS of diameter DAS. The beam transmitted by the AS is focused by the lens L2 on the FPM. The on-axis beam incident on the FPM is transmitted while the off-axis beam is reflected by the mask and directed toward the collimating lens L3 and the Lyot stop LS. The beam transmitted by the LS is focused by the lens L4 and directed by a folding mirror toward the CCD camera. This CCD camera is mounted on a translation stage for fine focusing. A photometric filter, which can either be I band or R band, is placed in front of the CCD.

A filter wheel, placed in front of the camera, houses a neutral density filter OD2 to measure the unsaturated PSF peak and an achromatic doublet to reimage the pupil plane on the camera. The camera used is an FLI Interline CCD camera cooled to a temperature of T = −146 C with a pixel size of 7.4 μm. During the contrast measurements, all of the optical components after lens L2 are enclosed in a black box to protect from stray light.

Figure 2 shows the spectral flux incident on the detector. It takes into account the irradiance due to the tungsten halogen lamp, single-mode fiber, transmission coating of the achromatic doublets, and the prism used. The relative spectral ratio is approximately equal to 20% in the I band and 23% in the R band.

Figure 2.

Figure 2. Profiles of the spectral flux incident on the detector at λc ≈ 800 nm (left) and λc ≈ 650 nm (right).

Standard image High-resolution image

The current EvWaCo test bed has been upgraded with a new coronagraphic mask built according to the design presented in Figure 3. The static prism S is mounted on the static support while the translating prism is mounted on the carrier moving along the high-precision linear slide. The pressure between the lens and the prism is adjusted by turning the micrometer that compresses a coil spring attached to the lens. The force over the contact area can be controlled from 0 N to 50 N with an accuracy equal to 0.02 N. We emphasize that the support of the static prism has been designed in order to keep the line of sight stable while adjusting the pressure between S and T with an accuracy better than 15''.

Figure 3.

Figure 3. Conceptual design of the EvWaCo coronagraphic mask.

Standard image High-resolution image

This coronagraphic mask is mounted on a support (Figure 4, left) that comprises five degrees of freedom: two translations in the horizontal plane, one rotation around the vertical axis, and two rotations around the horizontal axes. The tip-tilt orientation of the mask can be adjusted over a range larger than ±5° and with an accuracy better than 10''. This tip-tilt mount is mounted on two translating stages, each equipped with a micrometer that is used to position the FPM on the focus with an accuracy equal to 2 μm. Figure 4 (right) shows a close-up view of the coronagraphic mask.

Figure 4.

Figure 4. Image of the assembled FPM installed on the test bed (left) and a close-up view of the coronagraphic mask (right).

Standard image High-resolution image

4. Mask Response Profile and Pupil Irradiance Distribution

We measure the flat-field irradiance by placing an integrating sphere between the lens L1 and the aperture stop. A narrowband spectral filter with a central wavelength of λc is placed between the integrating sphere and the aperture stop. These components are only placed during flat-field measurements; otherwise, they are removed from the setup.

4.1. Variation of the Mask Response with the Applied Pressure

Figure 5 represents the variation of the shape of the coronagraphic mask for a force applying S on P, and varying from 0.10 N and 10.5 N for a source of central wavelength λ1 = 800 nm with an FWHM of 40 nm. At the time of these measurements, the reference is considered to be the case where the lens and the prism are placed in contact without applying any force.

Figure 5.

Figure 5. (Top) Variation of the shape of the mask response with respect to the increasing force applied on the static prism by the convex lens. These measurements were obtained in unpolarized light using a source of central wavelength equal to λc = 800 nm and Δλ = 40 nm. (Bottom) Cross-section along the x-axis (left) and the y-axis (right) of the reflection profile of the mask. As the force applied increases, the size of the mask enlarges, thereby increasing the IWA and the contact area between the prism and the lens.

Standard image High-resolution image

There is no force sensor in the setup so the force is calculated using the configuration of a sphere placed in contact with a flat plate (Hertzian contact stress; Budynas & Nisbett 2014). We calculate the contact stress (diameter of the hole) that corresponds to a specific contact pressure represented by the micrometer readings. This results in a force profile given the actuator reading, and thus relates the diameter of the contact area to the force/pressure applied. Figure 5 also shows the x and y profiles of the mask response to increasing pressure. It can be observed that as pressure on the prism increases, the size of the mask increases in terms of the FWHM and the contact area.

In this paper, the IWA is defined as the distance from the star center where the postcoronagraphic throughput is 50%. Thus, it is expressed in terms of λ/D, where D is the diameter of the aperture stop. Because the size of the mask directly impacts the IWA, by varying the force applied by the lens on the prism, the IWA can be finely tuned as well. For example, an FWHM of 222 μm corresponds to 3.7 λ/D at λc ≈ 800 nm.

The pressure of the lens has been adjusted until an IWA approximately equal to 3λ1/D is achieved along the minor axis for a source central wavelength equal to λ1 ≈ 800 nm. All of the experiments presented hereafter have been performed under this pressure unless specified otherwise.

4.2. Modeling the Coronagraphic Mask

To model this coronagraphic mask, the air thickness, d, is approximated by a polynomial equation given by Equation (1):

Equation (1)

where a1 and a2 are the coefficients of the polynomial of degrees 1 and 2, respectively, and ρ is the radial distance to the contact point in the prism hypotenuse plane. We calculate the theoretical reflection coefficients by applying the coefficients locally in the case of a thin, homogenous air gap with an optical index, n2, sandwiched between two optical indices, n1 and n3 (Buisset et al. 2017; Zhu et al. 1986). The coefficients of the polynomial are calculated by minimizing the sum of the squared difference between the simulation and the experimental data. We use a built-in function in Matlab called fminsearch that uses the Nelder-Mead simplex direct search (Lagarias et al. 1998) to perform the optimization. Using this optimization, the coefficients of the polynomial are found to be a1 = −1.10−3 and a2 = 36 μm−1.

Figure 6 compares the theoretical and experimental mask reflectivity for unpolarized light. The extracted profiles from the experimental data are normalized with respect to the median value of the calibrated image that is measured at a distance equal to 400 μm (13 λ1/D) from the optical axis of the coronagraphic mask. The horizontal error bars represent ±1 pixel while the vertical error bars represent the standard deviation ±σ of 101 intensity measurements.

Figure 6.

Figure 6. (Left) Approximation of the air gap at the contact area between the lens and the prism. (Middle) Comparison between the theoretical and measured mask response along the major axis of the mask (x-axis). (Right) Comparison between the measured mask response along the minor axis (y-axis). The error bars along the horizontal axis are equal to ±1 pixel while the error bars along the vertical axis are the standard deviation of the irradiance value from the 101 flat-field images obtained at the time of the acquisition of the experimental data.

Standard image High-resolution image

The mask profiles show that the FWHMs along X and Y are FWHMx ≈ 240 μm and FWHMy ≈ 180 μm, respectively. Consequently, the measured IWAs are as follows: IWAx ≈ 4.00 λ1/D and IWAy ≈ 3.00 λ1/D at the wavelength λc = 800 nm.

This enlargement of the mask along the X-axis is due to the angle of incidence on the coronagraphic mask prism hypotenuse as discussed in our previous paper (Buisset et al. 2017). We verified that the ratio ${\mathrm{IWA}}_{x}/{\mathrm{IWA}}_{y}=1/\cos (\theta )$ is equal to 1.33, which is compatible with an angle of incidence approximately equal to θ = 414.

At distances between −50 μm and +50 μm along the x-axis and between −30 μm and +30 μm along the y-axis, we notice that the theoretical mask transmission is slightly smaller than the experimental one. This difference is close to the error bar but could explain the discrepancy between the theoretical and the experimental raw contrasts at distances smaller than 3 λ/D as presented in Section 5 and in Figure 10.

At distances between +50 μm and +150 μm along the x-axis and between +70 μm and +125 μm along the y-axis, we notice a good match between the theoretical model and the experimental results, and the FWHM is correctly modeled. At distances greater than 150 μm in the x-axis and 120 μm in the y-axis, we notice a slight discrepancy between theory and experiment. However, at these distances, we expect that this small difference will only have a small impact on the raw contrast calculation.

4.3. Mask Response at Different Wavelengths

We, then, investigate the response of the new FPM to a different wavelength, λ2 = 650 nm. To measure the mask response at λ2, we used an LED emitting at λpeak = 617 nm and a spectral filter centered at λ2 = 650 nm with an FWHM equal to 40 nm.

Figure 7 shows the theoretical and experimental mask response profiles of the FPM obtained using the spectral filter with central wavelength λ2 and compared it to the mask response at λ1. The theoretical mask response is calculated by using the same polynomial coefficients obtained at 800 nm. We observe that the mask profiles are smaller at λ2 (blue curve) compared to the ones obtained at λ1 (pink curve). In terms of the IWA, the values are quite close. At λ2, IWAX ≈ 4.27 λ/D (FWHMX ≈ 208 μm) and IWAY ≈ 3.24 λ/D (FWHMY ≈ 158 μm).

Figure 7.

Figure 7. (First column) Experimental mask profiles along the x- and the y-axes at different central wavelengths: λ1 = 800 nm (pink curve) and λ2 = 650 nm (blue curve). (Second column) Theoretical mask profiles at these same wavelengths.

Standard image High-resolution image

These results show that the mask itself adapts well to wavelengths between λ1 and λ2 that correspond to the central wavelengths of the I and R bands of the Johnson-Cousins photometric filter. These first experimental observations of the quasi-achromatic behavior of the mask confirm the theoretical prediction for two wavelengths separated by 150 nm.

4.4. Pupil Irradiance Distribution

Figure 8 represents the irradiance distribution at the exit pupil (EP) plane when the PSF of the star source is placed off-axis (left) and when it is centered on the mask (right). These images were obtained by placing a lens, L5, inside the filter wheel (see Figure 1) in order to image the EP. The LS diameter is equal to 81% of the EP diameter and transmits approximately 65% of the incident light. The results show that the energy is mostly concentrated at the edges, forming a fine annular shape with a diameter that is approximately 2 mm, and an average extension width equal to 86 μm.

Figure 8.

Figure 8. Irradiance distribution at the exit pupil when the PSF is placed off-axis (left) and when it is centered on the mask (right). The images were obtained using a source of central wavelength λC = 780 nm and a spectral ratio of Δλ/λ ≈ 3%. The black circle represents the position of the Lyot stop. The Lyot stop diameter is approximately 81% of the pupil diameter.

Standard image High-resolution image

5. Deep Contrast Measurements

5.1. Raw Contrast Performance

Figure 9 illustrates the calculation of the contrast adopted in this paper. We decided to take the definition of the raw contrast presented in Ruane et al. (2018). Thus, the raw contrast is calculated as

Equation (2)

where ηs (xo , λ) is the fraction of available starlight detected and ηp (xo , λ) is the fraction of companion light detected. The quantity ηs (xo , λ) is given by

Equation (3)

where PSFcoro(x, 0, λ) is the point-spread function (PSF) centered on the FPM recorded at the image plane and AP(xo ) is equal to 1 pixel. Meanwhile, the quantity ηp (xo , λ) is calculated as

Equation (4)

where PSFcoro(x, xo , λ) is the PSF recorded at a distance xo from the center of the mask. We assume that PSFcoro is invariant over the FOV of the coronagraph at distances higher than 7 λ/D where the mask transmission is close to 1.

Figure 9.

Figure 9. Qualitative schematic for calculating the raw contrast. The star and the planet are separated by an angle, θ. The coronagraph image shows the star centered on the coronagraphic mask together with a planet located at a distance, xo , from the star center.

Standard image High-resolution image

In Figure 9 (left), we show an image of the star residual irradiance PSFcoro(x, 0, λ) and the PSF of the companion PSFcoro(x, xo , λ) measured at a distance xo = 8 λ/D from the mask center. On the right, we illustrate the calculation of ηs and ηp from the cross-section of the PSFcoro(x, 0, λ) and PSFcoro(x, xo , λ), respectively.

In our case, the fraction of the available starlight is the normalized irradiance, Inorm, that is the ratio between the unsaturated off-axis PSF peak and the on-axis PSF. A discussion of the calculation of Inorm from the experimental data can be found in the Appendix. The light detected from the companion, on the other hand, depends on the transmission, T(x, λ), of the mask. The raw contrast is, then, calculated as

Equation (5)

We show in Figure 10 the normalized irradiance and raw contrast obtained using two spectral bands of the Johnson-Cousins photometric filter (Bessel 1990): I band and R band. The I-band filter has a central wavelength λc = 800 nm with a wavelength range between 700 nm and 900 nm and a spectral ratio Δλ/λC = 20%. Meanwhile, the R-band filter has a central wavelength of λc = 650 nm with a wavelength range between 550 nm and 800 nm and a spectral ratio Δλ/λC = 23%. We note that only the bandpass filter located in front of the camera during the acquisitions over the full I band or R band is changed.

Figure 10.

Figure 10. Experimental (column 1) and simulated (column 2) normalized irradiance distribution of the on-axis star over the full I band (row 1) and R band (row 2). The third column shows the calculated raw contrast for each passband obtained both experimentally and numerically.

Standard image High-resolution image

We superimpose on the same figure the results of the simulations done using the same model described in Buisset et al. (2018) represented by the dashed lines with an updated mask transmission. This model takes the following inputs: (i) spectral flux incident on the detector (see Figure 2), (ii) the air thickness profile (see Figure 6), (iii) the diameter of the AS and LS, and (iv) the focal length of L2, L3, and L4. The complex amplitude incident on the FPM is obtained by taking the Fourier transform of the complex amplitude transmitted by the AS while the complex amplitude incident on the CCD is obtained by taking the Fourier transform transmitted by the LS. A full description of this model can be found in Buisset et al. (2018).

Figure 10 shows the radial average of the raw contrast measured at radial distances between 1.5 λ/D to 20 λ/D (pink curve). We superimpose in the same graph the mask response curves obtained through the experiment and simulations. For both spectral bands, the coronagraph reduce the PSF peak by three orders of magnitude.

Between 3 λ/D and 4 λ/D, the raw contrast in the I band varies between 2.10−4 and 5.10−5 while it varies between 2.10−3 and 3.10−4 in the R band. Meanwhile, along the same distance, the theoretical raw contrast varies between 1.10−3 and 2.10−4 in the I band and between 3.10−3 and 2.10−4 in the R band. Between 4 λ/D and 5 λ/D, the raw contrast in the I band varies between 5.10−5 and 2.10−5, while it varies between 3.10−4 and 8.10−5 in the R band. On the other hand, the theoretical raw contrast at the same distance varies between 1.10−4 and 3.10−5 in the I band and between 2.10−4 and 4.10−5 in the R band. We thus notice a difference between the theoretical model and the experimental data. The potential origin of this discrepancy could be the mismatch between the theoretical and the experimental mask transmission as discussed in the previous section. Although the experimental raw contrast values between the two bandpass differs by an order of a magnitude, the theoretical results show that the raw contrast performance is similar in the R and I bands, thus confirming the theoretical achromaticity of EvWaCo. We suspect that the presence of low-order aberrations close to the IWA contributed to the discrepancy in the experimental raw contrast value.

Between 5 λ/D and 10 λ/D, the contrast decreases from C ≈ 2.10−5 to C ≈ 3.10−6 for the I band and from C ≈ 8.10−5 to C ≈ 9.10−6 for the R band. Along the same distance, the theoretical contrast is between C ≈ 10−5 to C ≈ 10−6 for both the I band and the R band. We believe that the origin of this difference between the experiment and the theoretical contrast is the presence of speckles in the experimental measurements. Overall, our theoretical results show that the similarity between the I and R bands is valid at every radial distance. This confirms one more time the theoretical achromatic behavior of the EvWaCo mask in unpolarized light over the two spectral bands.

5.2. Raw Contrast at Different IWAs

Figure 11 shows the experimental raw contrast obtained at different IWAs. We notice that the best raw contrast performance is always reached with the FPM with the smallest IWAs. It is important to mention that:

(i) The FPM transmission profile is quasi-Gaussian only when the contact area is small and when the IWA is close to 3 λ/D (minor axis). This quasi-Gaussian shape has the reputation of providing a better contrast performance than the sharp-edge transmissions obtained at larger pressure (Figure 5), and

(ii) The achromatic property of this coronagraph is maximum at smaller IWA. Indeed, the chromatic variation of the mask response is valid only at the distance where the mask transmission varies between 0 and 1. Hence, we deduce that masks with a small IWA will provide a quasi-Gaussian response that fully adapts itself to the wavelength. The masks with a large IWA will provide a top-hat transmission that does not change with the wavelength.

During astronomical observations, the size of the proposed FPM could thus be adjusted in order to provide:

(i) A small IWA close to 3 λ/D to reach the best contrast performance during excellent seeing conditions of star environment imaging and companion detection and characterization, and

(ii) A large IWA between 3 λ/D and 8 λ/D during bad seeing conditions to improve the low-order wave-front distortion measurement by a WFS installed on the star channel (Buisset et al. 2018) and to better correct the contrast degradation induced by the turbulence.

The large IWA could also be used to suppress the halo of extended objects with an acceptable loss of contrast performance, for example, in Cosmic Origins Science observations as mentioned in Ebbets et al. (2015).

Figure 11.

Figure 11. (Top) 2D map of the normalized irradiance distribution of the on-axis star over the full I band at different IWA . (Bottom) Radial average of the raw contrast curves. Refer to Figure 5 for the mask response at each IWA.

Standard image High-resolution image

5.3. Detection of a Companion

We consider that the coronagraph is used to detect a faint companion with its PSF located in the vicinity of the star PSF that is itself set on axis. The detection of the companion thus consists of extracting the companion signal from the background noise induced by the following contributors: star residual photon noise, speckle noise, stray-light photon noise, detector thermal noise, and readout noise. We assess the level of this background noise by calculating the standard deviation, σBackground, over an annular region starting with an inner radius R1 = 2.5 λ/D and an outer radius, R2 = 3.5 λ/D. This calculation is repeated over an annular region with inner radii equal to 3.5, 4.5, and 19.5 λ/D, and with an annular width that is always kept equal to 1 λ/D.

In Figure 12, we have represented the level of the background noise at distances between 3 λ/D and 20 λ/D (black dashed curve). Between 3 λ/D and 10 λ/D from the center of the mask, the background noise varies from σBackground ≈ 3.10−5 to σBackground ≈ 10−6. At distances greater than 10 λ/D, this noise decreases from σBackground ≈ 10−6 to σBackground ≈ 10−7. The background noise comprises the detector noise, the star residual photon noise, the stray-light photon noise, and the speckle noise.

Figure 12.

Figure 12. Irradiance distribution of the star with a companion, 25,000 times fainter than the star, located at 5.5 λ/D from the star center, without the coronagraph (first figure), with the coronagraph (second figure), and cross-section along the x-axis of the irradiance distribution of the star and the companion (third figure). In this experiment, the star source is in the I band and the companion source is an LED diode emitting at λC = 780 nm with a spectral ratio Δλ/λ ≈ 3.2%. The images are a median of 101 calibrated science images obtained at an integration time of 1 s per frame. The S/N calculated at the companion location is equal to 11, illustrating the detectability of the companion.

Standard image High-resolution image

We place the PSF of a companion, 25,000 times fainter than the star, at a location of 5.5 λ/D distance from the center of the star. Figure 12 also shows the irradiance distribution for both the star and the companion with and without the coronagraphic effect. Obviously, when the star is off axis, the companion is barely detected in the blinding light of the star. However, once the star is centered on the coronagraphic mask, the companion becomes visible.

The detectability of the companion is verified by calculating the S/N at the location of the companion. To calculate the background noise, σBackground, at the location of the companion xo , we take the standard deviation of a circle with a radius equal to 1 λ/D, centered at xo . Figure 12 shows the residual light profiles without (left panel) and with (right panel) the companion. We estimate that at the location of the companion, the S/N is equal to 11.

6. Repeatability and Stability of the Measurements

We verified the stability of the FPM and the EvWaCo setup, and the repeatability of the reported contrast by measuring the performance of the coronagraph over 8 months. In each measurement, the PSF was recentered on the FPM. The pressure between the lens and the prism was the same throughout the 8 months. However, the lateral position of the Lyot stop was readjusted from time to time so that it remains centered on the EP.

Figure 13 shows the x and y profiles of the residual light pertaining to the I band. The black curve represents the average of these measurements. We observe that the overall shape of the profile remains the same despite taking the measurements at different dates. Figure 14 shows the radial average of the curves for 10 sets of measurements. At the location of the shortest IWA, the average normalized irradiance is Inorm ≈ 10−4. The average normalized irradiance is equal to Inorm ≈ 6.10−5 ± 1.10−5 at 3.5 λ/D, Inorm ≈ 3.10−5 ± 7.10−6 at 4.5 λ/D, and Inorm ≈ 2.10−5 ± 5.10−6 at 5.5 λ/D. Note that these measurements were obtained on a passive setup without adaptive optics.

Figure 13.

Figure 13. Cross-section along the x-axis (left) and the y-axis (right) of the PSF images obtained over the full I band in unpolarized light. The different colors represent 10 normalized irradiance measurements done over 8 months. It can be observed that the overall shape of the X and Y profiles is retained over this time period.

Standard image High-resolution image
Figure 14.

Figure 14. Radial average of the normalized irradiance and the corresponding standard deviation of the 10 contrast measurements. At the location of the IWA, the average normalized irradiance is approximately equal to Inorm ≈ 6.10−5 ± 1.10−5 at 3.5 λ/D, Inorm ≈ 3.10−5 ± 7.10−6 at 4.5 λ/D, and Inorm ≈ 2.10−5 ± 5.10−6 at 5.5 λ/D.

Standard image High-resolution image

7. Sensitivity Measurements

7.1. Variation of the Normalized Irradiance Due to Tip-tilt

When observing the environment of stars with large telescopes equipped with adaptive optics, it is critical that the PSF is properly centered on the FPM; otherwise, it will contribute to the contamination of the faint companion signal (Lloyd & Sivaramakrishnan 2005). The measurement of the variation of the contrast performance is done by installing a pointing camera. A beam splitter is installed right after the aperture stop, and the beam reflected is focused by the lens L5 on a Point Grey Camera with a pixel size of 3.4 μm as shown in Figure 1.

First, the PSF is centered on the mask which serves as the reference image. Then, tilt is introduced by decentering the PSF on the FPM. This is done by controlling the lateral alignment of the star source fiber output face. We show in Figures 15 and 16 the irradiance distribution and the respective x and y profiles recorded by the CCD at the companion channel. The tilt is expressed in terms λ/D, where D is equal to the diameter of the aperture stop. The error bar corresponds to the standard deviation of the contrast values calculated from the 101 images obtained for each tilt.

Figure 15.

Figure 15. (Top panel) Evolution of the irradiance distribution recorded by the CCD camera at the companion channel as tilt along the x-axis is introduced in the test bed. The black cross indicates the center of the mask, and the white ellipse represents the location of the IWA. The images are shown at the same visualization level. (Bottom panel) Cross-section along the x-axis of the normalized irradiance taken at the following locations: ±4λ/D, ±5λ/D, and ±6λ/D. The error bars correspond to the standard deviation of the normalized irradiance from the 101 images acquired for each tilt.

Standard image High-resolution image

Along the x-axis, we notice that when the tilt is smaller than θX < 1 λ/D, the normalized irradiance at the IWA is better than 10−4. Meanwhile, both the normalized irradiance at 5 λ/D and 6 λ/D show slight variations. At ±3 λ/D, the cross-section along the y-axis, on the other hand, shows an asymmetry in the normalized irradiance profile. This is due to the fact that the PSF position set as the reference is slightly shifted along the y-axis. Despite that, we can deduce from Figure 16 that normalized irradiance is better than 10−4 for tilt angles smaller than θY < 0.2 λ/D.

Figure 16.

Figure 16. (Top panel) Evolution of the irradiance distribution recorded by the CCD camera at the companion channel as tilt along the y-axis is introduced in the test bed. The black cross indicates the center of the mask, and the white ellipse represents the location of the IWA. The images are shown at the same visualization level. (Bottom panel) Cross-section along the x-axis of the normalized irradiance taken at the following locations: ± 3λ/D, ± 4λ/D, and ±5λ/D. The error bars correspond to the standard deviation of the normalized irradiance from the 101 images acquired for each tilt.

Standard image High-resolution image

7.2. Variation of the Normalized Irradiance Due to the Defocus

The variation of the normalized irradiance due to defocus is done by translating the star source along the optical axis and measuring the normalized irradiance value. This displacement is expressed in terms of the PTV (peak-to-valley) wave-front error. The PTV wave-front error induced by each displacement is determined by making paraxial calculations using Zemax. Figure 17 (left panel) shows the wave-front error in terms of the λ PTV induced for each displacement. For each displacement, 101 images are acquired, and the normalized irradiance values are calculated over a circular annulus with an extension of 1 λ/D. The standard deviation of these values calculated for the 101 images is represented as error bars. At 3 λ/D, from the center of the mask, the normalized irradiance values are better than 10−4 for a wave front less than 0.65 λ PTV as shown in Figure 17 (right panel).

Figure 17.

Figure 17. The figure shows the normalized irradiance at specific radial locations from the center of the mask: 3.5 λ/D (red curve), 4.5 λ/D (yellow curve), and 5.5 λ/D (blue curve) expressed in terms of the λ PTV.

Standard image High-resolution image

7.3. Variation of the Normalized Irradiance Due to Lyot Stop Misalignment

The purpose of the Lyot stop is to reduce, if not eliminate, the residual starlight in the EP plane that is due to diffraction. Thus, if this Lyot stop is not properly centered, it causes the starlight to leak, thus degrading the performance of the coronagraph. In this section, we intentionally misaligned the Lyot stop along the x and the y directions, and measured the normalized irradiance. Figures 18 and 19 show how the contrast varies when the Lyot stop is not exactly positioned at the center of the EP. The position of the Lyot stop with respect to the center of the EP is marked by the black circle.

Figure 18.

Figure 18. The figure at the top shows the position of the Lyot stop with respect to the exit pupil. The figure above shows the normalized irradiance for each corresponding misalignment along the x-axis. We observe that light begins to leak as the Lyot stop is misaligned. However, a misalignment of 74 μm (0.02 DEP ) does not significantly affect the normalized irradiance at the IWAX ≈ 4.00 λ/D.

Standard image High-resolution image

Figure 18 reveals that a misalignment of 74 μm (0.02 DEP ) along the x direction does not significantly impact the normalized irradiance at the location of the IWA (IWAX ≈ 4.00 λ/D, IWAY ≈3.00 λ/D). The same phenomenon is observed for a misalignment of 84 μm (0.02 DEP ) along the y direction as shown in Figure 19. These results show the possibility of enlarging the Lyot stop to 90% the size of the pupil diameter (currently, it is at 81%) at the expense of a tighter tolerance of the Lyot stop alignment.

Figure 19.

Figure 19. The figure at the top shows the position of the Lyot stop with respect to the exit pupil. The figure below shows the normalized irradiance for each corresponding misalignment along the y-axis. We observe that light begins to leak as the Lyot stop is misaligned. However, a misalignment of 84 μm (0.02 DEP ) does not significantly affect the normalized irradiance at the IWAY ≈ 3.00 λ/D.

Standard image High-resolution image

8. Discussion and Conclusion

Our recent results demonstrate that EvWaCo has now entered the regime of small-angle coronagraphy (IWA ≤4 λ/D; Mawet et al. 2012) over a wide spectral band [550 nm—900 nm]. It is important to note that the results have been obtained by using a setup on a passive support without isolation, in a nonvacuum laboratory environment, and without adaptive optics to correct the wave-front distortions induced by surface errors of the optical components and local air turbulence. These results show a clear improvement in the performance we have previously reported (Buisset et al. 2018) in terms of the raw contrast, the spectral band, and the IWA. First, the spectral band has been enlarged from the range of [700 nm, 900 nm] to [550 nm, 900 nm]. Second, the IWA is now fully tunable between 3 λ/D and 7 λ/D—a capability that is unique. We emphasize that the current IWA has been reduced by a factor of 2 compared to the previous one, which was at 6 λ/D. Finally, we have shown that this mask is able to reach a raw contrast close to a few 10−4 at 3 λ/D over the full I band and at 4 λ/D over the full R band. A series of contrast measurements over 8 months confirms its stability and the repeatability of the measurements. In terms of misalignment, the normalized irradiance measured at a distance close to 4 Airy radii is better than 10−4 even for tilt angles as large as 1 λ/D and defocus (0.65 λ PTV) at the IWA. The Lyot stop can be moved by at most ±74 μm without affecting the normalized irradiance that is also compatible with common mechanical tolerances. Finally, we have shown that a companion, 25,000 times fainter than the star, can be detected with good S/N at 5.5 λ/D from the star center.

The results presented in this paper show that the best contrast performance is reached for the minimum contact area that provides a fully achromatic quasi-Gaussian mask transmission. Therefore, the objective of the FPM future upgrades will consist of improving the mask optomechanical design to (i) reduce the size of this contact to a negligible value, (ii) keep the current performance and stability, and (iii) increase the maximum available mask size.

The future development of EvWaCo will aim to develop a full prototype that will include adaptive optics and demonstrating on-sky performance. The full prototype will demonstrate the capability of reaching contrasts that are in line with the laboratory results during ground-based observations. The first step will consist of installing an EvWaCo prototype on the 2.4 m Thai National Telescope at horizon 2022.

This work is carried out at the NARIT Center for Optics and Photonics. We thank the mechanical engineers of the Operations Department of NARIT for the fabrication of optical component supports in the test bed.

The Thai Space Consortium (TSC) is an association composed of the National Astronomical Research Institute of Thailand (NARIT), Siam Photon, Geo-Informatics and Space Technology Development Agency (GISTDA), Suranaree University of Technology (SUT), King Mongkut's University of Technology North Bangkok (KMUTNB), King Mongkut's Institute of Technology Ladkrabang (KMITL), and National Innovation Agency (NIA). T.S.C. is funded by PMU-B, Office of National Higher Education Science Research and Innovation Policy Council. The research is supported by the Program Management Unit for Human Resources & Institutional Development, Research and Innovation, NXPO (Grant number: B05F630115). We also thank the Chulalongkorn University's CUniverse (CUUAASC) grant for the support.

Finally, we are very grateful for the invaluable insights from the reviewer that helped us improve the quality of our paper.

Appendix

A.1. Data Acquisition Procedure

In the experiment, we obtained two sets of off-axis PSF images, PSFcoro(x, xo , λ), placed at a distance of xo ≈ 30λ/D. In the first set, we installed an optical density OD2 in front of the detector with the following transmission on each bandpass: tI band = 2.10−3 and tR band = 1.3. 10−3. We adjusted the integration time so that the peak of the star PSF is close to the maximum dynamic range of the camera and recorded N = 101 frames of the same integration time. Then, we recorded 11 dark images and 11 bias images at an integration time equal to 1 s and 0 s, respectively. We calculated the master dark frame and the master bias frame as the median of the 11 single dark frames and 11 bias frames, respectively. We subtracted the master bias and the dark frames from every single frame of the off-axis PSF and calculated the median of the 101 images. We measured on this median image the peak value SStar,peak of the (unsaturated) off-axis PSF.

On the second set, we removed OD2 and measured the signal SStar (x, y) of the PSF with the same integration time, same number N of images, and identical preprocessing method. This image of the off-axis PSF has now a saturated core but a high S/N in the PSF wings.

To obtain PSFcoro(x, 0, λ), we centered the PSF on the occulting mask and recorded the signal SResidual (x,y) at the same integration time. Thus, the normalized irradiance is calculated using Equation (6):

Equation (6)

To calculate the raw contrast, we divided Inorm by the mask transmission obtained from the experiment. For example, from Figure 10, at 3 λ/D, the normalized irradiance is approximately at Inorm ≈ 1.10−4 in the I band. Thus, the corresponding raw contrast at this location is equal to 2.10−4.