The X-ray radiographic imaging technique is very important in medical, biological, inspection, material science, and other fields. However, it is not enough to obtain a clear X-ray image of samples with low absorbance materials, such as biological soft tissues. Then, we have used an X-ray phase-imaging method of an X-ray Talbot interferometer. In this method, X-ray gratings were required to have a narrow pitch and high aspect ratio structure. Therefore, we have developed and fabricated high-aspect-ratio X-ray gratings with a pitch of Equation or symbol description not available5.3μm, a height of Equation or symbol description not available30μm, and a large effective area of Equation or symbol description not available100×100mm using X-ray lithography and narrow electroforming techniques. In this paper, we discuss the fabrication process of X-ray gratings with a narrow pitch and high-aspect-ratio structure, and results of X-ray phase tomography using an X-ray Talbot interferometer with these X-ray gratings.