Abstract
The electrical conductivity of
as a function of
was investigated in the two gaseous systems
and
. The experimental results showed different conductivity values for these two different systems. The oxygen partial pressure was, therefore, checked carefully and it turned out that the control of
at reducing conditions is complicated by disequilibrium states when using
gas mixtures. A sample and an oxygen sensor that are positioned side by side may be exposed to
values that differ by up to 2 log units because of their different catalytic behavior with respect to the oxidation of CO. At intermediate
conditions, e.g., between
and
at
, low concentrations of
(CO) and
may result in
gradients of up to 10 log units per centimeter in both
and
systems.