This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.

Test bench for fibers of the LHAASO-WCDA time calibration system

, , , , , , , , and

Published 26 October 2017 © 2017 IOP Publishing Ltd and Sissa Medialab
, , Citation J.Y. Liu et al 2017 JINST 12 P10021 DOI 10.1088/1748-0221/12/10/P10021

1748-0221/12/10/P10021

Abstract

The Water Cherenkov Detector Array (WCDA), which is one of the main components of the Large High Altitude Air Shower Observatory (LHAASO), functions in surveying the northern sky for high-energy gamma ray sources at the energy range around of 100 GeV–30 TeV. The precision of the time measurement for shower particles hitting every detector in the array is directly associated with the detection sensitivity for the sources. The calibration precision of the time offsets among the detector cells should be less than 0.1 ns to obtain less than 0.1o pointing error of the detector to any point source. In this regard, "cross calibration" is employed for the detector array, with 180 bundles of fiber systems guiding the lights of LEDs to each PMT. A test bench comprising fast PMTs and an one-dimensional slide platform is set up in the laboratory to test the fiber bundles. The test bench and the test procedure is described in this paper, and the test results are presented.

Export citation and abstract BibTeX RIS

10.1088/1748-0221/12/10/P10021