Measurement of a PAGAT gel dosimeter by ultrasound computed tomography

, and

Published under licence by IOP Publishing Ltd
, , Citation S Khoei et al 2013 J. Phys.: Conf. Ser. 444 012083 DOI 10.1088/1742-6596/444/1/012083

1742-6596/444/1/012083

Abstract

In this work we used a 3D quantitative CT ultrasound imaging system to characterise polymer gel dosimeters. The system comprised of two identical 5 MHz 128 element phased-array ultrasound transducers co-axially aligned and submerged in water as a coupling agent. Rotational and translational movement of the gel dosimeter sample between the transducers were performed using a robotic arm. Ultrasound signals were generated and received using an Olympus Omniscan unit. Dose sensitivity of attenuation and time of flight ultrasonic parameters were assessed using this system.