Recent developments with a prototype fan-beam optical CT scanner

, and

Published under licence by IOP Publishing Ltd
, , Citation W G Campbell et al 2013 J. Phys.: Conf. Ser. 444 012066 DOI 10.1088/1742-6596/444/1/012066

1742-6596/444/1/012066

Abstract

The latest design of a prototype fan-beam optical computed tomography scanner is presented. A new beam creation system consists of a 635 nm laser diode module with variable, DC voltage-controlled beam intensity. A change in scanner alignment allows for the elimination of ring artefacts caused by data corruption that is spaced symmetrically across the detector array. These artefacts, as well as a pair of streaking artefacts caused by flask seams, are removed in sinogram space. A flask registration technique has been developed that allows for accurate, reproducible dosimeter placement. Protocol investigations with gel dosimeters have indicated the importance of: i) proper cooling techniques during gel manufacture, and ii) scanning the dosimeter while it is at room temperature. Latest reconstructions of a normoxic polymer gel dosimeter are presented as an indicator of current system performance.

Export citation and abstract BibTeX RIS