3D Dose reconstruction: Banding artefacts in cine mode EPID images during VMAT delivery

and

Published under licence by IOP Publishing Ltd
, , Citation H C Woodruff and P B Greer 2013 J. Phys.: Conf. Ser. 444 012042 DOI 10.1088/1742-6596/444/1/012042

1742-6596/444/1/012042

Abstract

Cine (continuous) mode images obtained during VMAT delivery are heavily degraded by banding artefacts. We have developed a method to reconstruct the pulse sequence (and hence dose deposited) from open field images. For clinical VMAT fields we have devised a frame averaging strategy that greatly improves image quality and dosimetric information for three-dimensional dose reconstruction.