Abstract

Quantitative comparisons of dose distributions are an integral component of a medical physicist's responsibility of assuring high quality radiation therapy dose delivery. While overlays and other displays of multiple dose distributions are useful for such evaluations, quantitative evaluations require a mathematical comparison. The dose-difference is the most straightforward method for comparing two dose distributions, but it can show large differences in steep dose gradient regions, even for relatively small misalignments. A tool, termed y,was developed to take both dose and spatial difference into account. It does this automatically, evaluating distributions for dose difference and spatial discrepancies in regions of shallow and steep dose gradients, respectively. The tool has been used extensively in commercial dose measurement and evaluation software. This chapter describes the tool, some alternative techniques, and limitations of the tool.

Export citation and abstract BibTeX RIS