Brought to you by:

Activity measurements of the radionuclide 99mTc for the VNIIM, Russian Federation and ENEA-INMRI, Italy, in the ongoing comparison BIPM.RI(II)-K4.Tc-99m and KCRV update in the BIPM.RI(II)-K1.Tc-99m comparison

, , , , , , , , and

©2016 BIPM & IOP Publishing Ltd
, , Citation C Michotte et al 2016 Metrologia 53 06014DOI 10.1088/0026-1394/53/1A/06014

0026-1394/53/1A/06014

Abstract

In 2014, comparisons of activity measurements of 99mTc using the Transfer Instrument of the International Reference System (SIRTI) took place at the VNIIM (Russian Federation) and at the ENEA-INMRI (Italy), with comparison identifier BIPM.RI(II)-K4.Tc-99m. Ampoules containing about 32 kBq of 99mTc solutions were measured in the SIRTI for three half-lives. There are now nine results in the BIPM.RI(II)-K4.Tc-99m comparison. This comparison is linked to the BIPM.RI(II)-K1.Tc-99m comparison and the key comparison reference value (KCRV) of the BIPM.RI(II)-K1.Tc-99m has been updated to include eligible results from BIPM.RI(II)-K4.Tc-99m. The degrees of equivalence between each equivalent activity measured in the SIR or the SIRTI and the KCRV have been calculated and the results are given in the form of a table. A graphical presentation is also given.

Main text

To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Export citation and abstractBibTeXRIS

undefined