This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.

The diffraction correction for absolute gravimeters

and

Published 11 September 2003 2003 BIPM and IOP Publishing Ltd
, , Citation D van Westrum and T M Niebauer 2003 Metrologia 40 258 DOI 10.1088/0026-1394/40/5/008

0026-1394/40/5/258

Abstract

The inherent curvature of the wavefronts in the laser beams used in most absolute gravimeters introduces a systematic reduction in the measured gravity value. This error depends strongly on the diameter of the laser beam, but is typically of the order of 1–5 µGal. This paper describes the origin of this reduction, and provides a correction based on the laser beam diameter as well as a practical method to measure the beam's diameter.

Export citation and abstract BibTeX RIS