Abstract
Brachytherapy is a radiotherapy treatment where encapsulated radioactive sources are introduced within a patient. Depending on the technique used, such sources can produce high, medium or low local dose rates. The Monte Carlo method is a powerful tool to simulate sources and devices in order to help physicists in treatment planning. In multiple types of gynaecological cancer, intracavitary brachytherapy (HDR Ir-192 source) is used combined with other therapy treatment to give an additional local dose to the tumour. Different types of applicators are used in order to increase the dose imparted to the tumour and to limit the effect on healthy surrounding tissues. The aim of this work is to model both applicator and HDR source in order to evaluate the dose at a reference point as well as the effect of the materials constituting the applicators on the near field dose. The MCNP5 code based on the Monte Carlo method has been used for the simulation. Dose calculations have been performed with *F8 energy deposition tally, taking into account photons and electrons. Results from simulation have been compared with experimental in-phantom dose measurements. Differences between calculations and measurements are lower than 5%.The importance of the source position has been underlined.
Export citation and abstract BibTeX RIS