Characterization of Blood Properties from Coagulating Blood of Different Hematocrits Using Ultrasonic Backscatter and Attenuation

and

Published 7 September 2006 Copyright (c) 2006 The Japan Society of Applied Physics
, , Citation Chih-Chung Huang and Shyh-Hau Wang 2006 Jpn. J. Appl. Phys. 45 7191 DOI 10.1143/JJAP.45.7191

1347-4065/45/9R/7191

Abstract

The influence of hematocrit on the change of blood properties during coagulating was extensively investigated using ultrasonic integrated backscatter and attenuation. Measurements were performed with porcine blood at hematocrits ranging from 25 to 55% using a 10 MHz transducer. Results showed that both integrated backscatter and attenuation are able to sensitively differentiate various stages of blood properties during coagulating. The slopes of integrated backscatter (Sr, dB/S) and attenuation (αr, dB·cm-1·MHz-1·mS-1) are increased relative to hematocrit. The best fits for Sr and αr as a function of hematocrit (H) equal to Sr=0.0357+1.62e-0.108H and αr=0.0281+0.003H, respectively. Variations of clotting time (Ts) and reaction time (Tα), estimated respectively from ultrasonic integrated backscatter and attenuation, associated with clot formation are also increased with hematocrit. This study demonstrates that blood hematocrit is a substantial factor affecting viscosity and backscattering properties of blood during coagulation capable of being discerned by ultrasonic backscattering and attenuation.

Export citation and abstract BibTeX RIS