Brought to you by:
Paper

Optical coherence tomography: an experimental validation for vascular imaging of saphenous vein bypass grafts

, , , , , , , and

Published 24 February 2016 © 2016 IOP Publishing Ltd
, , Citation Rebekah Wilson et al 2016 Biomed. Phys. Eng. Express 2 025002 DOI 10.1088/2057-1976/2/2/025002

2057-1976/2/2/025002

Abstract

Purpose. Optical coherence tomography (OCT) is a novel light-based intravascular imaging method with potential utility for quantifying vascular disease and the effect of therapy. OCT has been used to image vein grafts and neointima in clinical research studies but validity of OCT for vein graft imaging is limited by depth of field and tissue penetration. Experiments were carried out to validate the in-house developed new software (Medipass-iScan) and OCT vendor software comparing with a gold standard, photomicroscopy. Methods. Seven synthetic phantom tubes with varying inner diameters and eight saphenous veins were imaged with OCT. Imaging was performed five times on each phantom/vessel and sections of each were measured by photomicroscopy. OCT images were analyzed by iScan to measure the inner diameters and then compared with corresponding microscopy sections. Results. A Bland–Altman plot of differences between photomicroscopy and OCT measurements of phantoms, demonstrated evidence of limited bias (104 μm) and 95% limits of agreement, −100 and 308 μm. The mean variation of iScan OCT measurements from microscopy was 3.02% and that of the OCT vendor software 3.03%. Conclusions. OCT has high measurement accuracy of lumen diameter. iScan measurements of saphenous veins imaged by OCT had similar accuracy to vendor software, supporting its validity. Potentially, OCT may be used to measure saphenous vein dimensions ex vivo.

Export citation and abstract BibTeX RIS

Introduction

Coronary artery disease (CAD) is a major global cause of morbidity and mortality [1]. Coronary artery bypass grafting (CABG) is an evidence-based treatment for CAD that improves symptoms and survival, however the benefits of CABG are limited by saphenous vein graft disease [2, 3]. 10% of saphenous vein grafts have angiographic evidence of disease at 1 year and 66% have obstructive disease by 10 years [4], which is mainly secondary to the effects of arterial-level blood pressure and vascular remodeling [2]. While CT-angiography is useful for evaluating vein graft patency, the established method for quantifying the extent of saphenous vein graft disease is intravascular ultrasound (IVUS) [5, 6].

Optical coherence tomography (OCT) is a novel light-based intravascular imaging method that potentially might image saphenous veins with greater accuracy and precision. OCT involves an intravascular catheter, which can capture cross-sectional tomographic images by measuring back reflected light emitted from the probe with a tissue resolution of <20 μm. OCT has been used to image saphenous vein graft disease in patients with prior CABG. Adlam et al [7] concluded that when compared to IVUS, OCT delineated saphenous vein graft disease in CABG recipients including circumferential fibrous neointima and thin cap fibroatheroma [8]. Prati et al [8] concluded that when compared to current IVUS, OCT provided superior visualization and differentiation of the coronary artery lumen and wall interface [8].

Expert guidelines have identified a role for OCT to quantify the anti-atherosclerotic effects of therapeutic interventions in human coronary arteries [9]. Vein graft failure is a clinically important area of unmet need and new preventative therapies are urgently required [7]. OCT has been used to image vein grafts and neointima in clinical research studies but it is not validated for this purpose. Vein grafts are larger than coronary arteries and their adventitial border is typically more diffuse, thus light-based intravascular imaging with OCT could theoretically yield different results in these blood vessels. Furthermore, currently available OCT software has not been assessed for the purpose of measuring neointima or vein graft wall thickness.

The aim of this study was to assess the measurement accuracy and precision of the in-house developed iScan software and intra-vascular OCT vendor software for imaging saphenous veins. Using a developmental approach ex vivo, we initially created phantom tubes with similar dimensions to human conduit arteries and veins for ex vivo OCT imaging. The dimensions of the phantoms were customized to vary in diameter and thickness within the same tube in order to mimic saphenous veins experimentally. OCT tissue penetration is limited at 1–2.5 mm however Prati et al states that the 'tissue penetration of OCT rarely exceeds 1.0–1.3 mm.' Average coronary artery wall thickness is 1.2 mm [11] and average saphenous vein bypass graft wall thickness is 1.4 ± 0.5 mm [10]. Due to the limited tissue penetration of OCT of 1–2.5 mm, it is not currently adept to studying vessel remodeling such as in saphenous vein bypass grafts.

Through developing iScan we would like to further the depth of the wall that can be analyzed, by improving the sensitivity of detecting backscattered light. We begin by validating accuracy of diameter measurements. Optimizing this parameter first, enables the software to learn as we progress to depth measurements.

Additionally iScan makes all measurements automatically thus removing inter-user variation whereas vendor software can only automatically measure diameter, area and circumference, all other measurements are made by the user. This new software, iScan, would also allow us to directly compare OCT accuracy against IVUS, which is currently used in the clinical setting. Using the same software package to analyze both modalities would remove bias as OCT and IVUS vendor software may not be equally accurate.

Objectives

To measure the accuracy of OCT and iScan we had the following objectives: (1) to obtain OCT diameter measurements of the cross section of synthetic grafts and cadaveric vessels using both vendor software and iScan; (2) to validate the OCT vendor and iScan measurements using photomicroscopy; and (3) to compare the accuracy of iScan measurements with OCT vendor software measurements.

Materials and methods

Imaging experiments and analysis were conducted in the Beardmore Centre for Health Sciences in the Golden Jubilee National Hospital and photomicroscopy was performed in the Department of Pathology, Queen Elizabeth University Hospital, Glasgow. All of the experiments were performed from January 2013 to April 2014.

Phantoms

To allow for the validation of OCT, phantoms of known dimensions, which closely match that of a human coronary artery and mimic the optical properties of arterial tissue, were manufactured in house using precision moulds of specified dimensions that were custom manufactured in-house (NHS Medical Physics). The matrix material of the phantom was RTV silicone (Slygard 184) kit which was made in a ratio of resin:reactive 10:1 and alumina powder was added as the main scattering component. Carbon black was added in the wall for further attenuation of light. The silicone resin was mixed with 50 mg ml−1 of alumina powder and cast into the mould, this combination was based on research previously published by Bisaillon et al [11]. To ensure a uniform distribution of scattering material, the mixture was fornicated in an ultrasonic bath for 5 h before adding the reactive. It is cured at room temperature for 48 h.

Phantom version one (figure 1(a)) had a diameter 3.5 mm and a wall thickness of 1 mm throughout. There were two phantoms made to version 1 specifications. Version 2 (figure 1(b)) introduced a tube with a variable diameter, giving three zones: 4.5, 3.5 and 2.5 mm diameters with wall thicknesses of 0.5 mm, 1.0 mm and 1.5 mm respectively. The purpose of the variable diameter was to test the sensitivity of iScan to minor changes more characteristic of human vessels by mimicking the presence of stenosis. Version 3 (figure 1(c)) included different diameters (4.5, 3.5 and 2.5 mm) and different wall thicknesses of 1.25, 1.75and 2.25 mm. There were two phantoms made to version 3 specifications (3i and 3ii). Version 4 (figure 1(d)) introduced a 'plaque like bump' of 0.1 mm thickness. Version 5 (figure 1(e)) was created with 2 'plaque like bumps' for further testing. These plaques were designed to give a stronger OCT signal than the surrounding phantom, mimicking the characteristic signal of real plaques. All versions above were imaged but to facilitate clear comparison of the images for the reader, only version 3i has been shown in the figures section.

Figure 1.

Figure 1. (A) Version 1 phantom design n = 2. (B) Version 2 phantom design. (C) Version 3 phantom design n = 2. (D) Version 4 phantom design with one plaque like 'bump'. (E) Version 5 phantom design with two plaque like 'bumps'.

Standard image High-resolution image

Cadaveric vessels

Eight saphenous veins were obtained from cadavers at the Anatomy Department, University of Glasgow, which were fixed in formalin and stored in a safe and cool place throughout the study period. All vessels were approximately 80–100 mm in length and diameters of approximately 2–4 mm.

Optical coherence tomography

Experimental OCT imaging was performed with the St. Jude Medical Ilumien system. The OCT Dragonfly catheter (2.7 French) was purged with 3 ml of H2O and then 1 ml of omnipaque contrast (300 mg I ml−1). The phantom or vessel was placed in the holding apparatus and submerged in a container filled with water. A six French sheath was placed 1 cm into the proximal end of the graft. A 0.014 inch guidewire was inserted into the end of the OCT catheter and then they were together passed through the sheath and advanced to the distal end of the graft. The graft was flushed with H2O to remove air bubbles.

The light of the OCT probe was aligned at the proximal end of the phantom and the system calibrated. The automatic pullback was commenced (20 mm s−1) at the same time as an 11 ml bolus of contrast was injected. The whole process was repeated a total of five times.

The St Jude OCT automatic detection tool measured the mean diameter measurements of OCT images.

Photomicroscopy

The measurements of the customized phantoms were known before testing but because of the possibility that the dimensions may vary during or after the manufacturing process, the actual dimensions of each phantom were acquired by photomicroscopy shortly after OCT had been performed. The same approach was adopted for the cadaveric vessels to provide a gold standard for comparison. In this way, the actual dimensions of each phantom and blood vessel were used for comparison with the OCT findings.

Five corresponding 1 mm sections measured by OCT were sliced from each graft by hand using a scalpel with the grafts fixed to a measuring scale. Practice and care was taken not to damage the geometry of the graft during slicing. These sections were measured by a photomicroscope (Olympus SX10) at ×2 magnification, using Nikon software 2.3 (NIS-Elements). The microscopy image (figure 2) was superimposed on a circular target to guarantee the line of diameter crossed the centre of the section. The microscope was calibrated and a minimum of ten measurements of inner diameter were taken at adjacent points on the inner circumference and the mean calculated. It is possible there was interobserver error, however this was low. The average standard deviation of the ten measurements was 5.4 μm (minimum 2.08 μm and maximum 9.97 μm).

Figure 2.

Figure 2. Photomicroscopy image of phantom version 3i, measurement lines seen in green and recorded in top left box in the image.

Standard image High-resolution image

Image analysis with iScan

In order to provide open access to the software for OCT research purposes and measure more parameters than the OCT vendor software, X.J of the Medical Devices Unit, NHS GGC, developed a novel imaging software suite (Medipass-iScan) [12] in-house. iScan was an ImageJ plugin using dynamic programming [13]. Dynamic programming is an optimization approach that simply stated, bypasses local minima which has been reported for the lumen and plaque detections for in vivo OCT images [1416]. Similar to the approach proposed by Wang et al [16], the developed software detected the smoothest lumen boundary which was close to the available data and which at the same time preserve the discontinuities. The central circular regions of the OCT images were identical for all the pullback images. The central circular regions were removed by typing in the pixel coordinates of the sensor and the diameter of the circle to the software. The OCT polar images were transferred to rectangular images. The dynamic programming was applied to the rectangular images; firstly, the guide wire was detected and removed from the images; and then the lumen boundaries were detected [16]. All OCT pullback images were recorded and digitally archived. Images that corresponded with the cross section of the graft measured by photomicroscopy were measured offline using iScan software. The tube properties derived were: inner diameter, luminal area and wall thickness.

Results

Analysis of OCT phantom images

Figure 3(a) presents an OCT image of phantom version 3i with all three zones visible on the longitudinal view at the bottom. The 0.014 mm guidewire as labeled resulted in the long shadow through the wall of the phantom; the circular structure with turquoise borders was the OCT catheter. The uneven reflections in the wall were indicative of the alumina powder, which was dispersed throughout the wall to reflect the light. Figure 3(b) shows the OCT image analyzed by iScan. Measurements of vendor images and iScan images were then compared to our gold standard, photomicroscopy. Figure 2 shows photomicroscopy measuring of a phantom.

Figure 3.

Figure 3. (A) OCT image of version 3i. (B) iScan analysis of OCT image of version 3i.

Standard image High-resolution image

Accuracy of phantom inner lumen mean diameter measurements

Linear regression of iScan OCT diameters compared to the photomicroscopy measurements demonstrated that the iScan OCT measurements of microscopy were closely related with R2 = 0.901.

The Bland–Altman plot (figure 4) represents the differences between the photomicroscopy and iScan OCT measurements, plotted against the average of photomicroscopy and iScan OCT (one sample t-test of the differences p < 0.001). There was evidence of limited bias (104 μm) and 95% limits of agreement: −100 and 308 μm.

Figure 4.

Figure 4. Bland–Altman plot of differences between the photomicroscopy and iScan OCT measurements for phantom grafts (n = 7). Mean difference = 104 μm. 95% limits of agreement = 308 μm, −100 μm.

Standard image High-resolution image

The mean variation of iScan OCT measurements from microscopy was 3.02% (min −26%, max 88%) with SD of 7.96%. The mean variation of OCT vendor measurements from microscopy was 3.03% (min −15.56%, max 15.62%) with SD of 3.94%.

Analysis of OCT saphenous vein images

Figure 5(a) presents an OCT image of a saphenous vein ex vivo. Figure 5(b) shows the OCT image analyzed by iScan. Figure 6 shows photomicroscopy of a section of saphenous vein.

Figure 5.

Figure 5. (A) OCT image of a saphenous vein. (B) iScan analysis of OCT image of saphenous vein.

Standard image High-resolution image
Figure 6.

Figure 6. Image of photomicroscopy of a section of a saphenous vein, corresponding to images seen in figure 7. ×2 magnification. Green lines are diameter measurements.

Standard image High-resolution image

Accuracy of saphenous vein mean diameter measurements

Figure 7 displays linear regression of iScan OCT diameters of saphenous veins ex vivo compared to the photomicroscopy measurements. It demonstrated that the image measurements of iScan OCT were positively related with R2 = 0.704.

Figure 7.

Figure 7. Linear regression of photomicroscopy diameters compared to the iScan OCT diameters for saphenous veins (n = 8). R squared = 0.704. R2 = 0.704. Y = 1.173 ×  −500.477.

Standard image High-resolution image

The Bland–Altman plot of saphenous vein imaging (figure 8) represents the differences between the photomicroscopy and iScan OCT measurements, plotted against the average of photomicroscopy and iScan OCT (one sample t-test of the differences p = 0.051). There was evidence of limited bias (116.24 μm), similar to the phantom grafts. 95% limits of agreement were −1754 and 1987 μm.

Figure 8.

Figure 8. Bland–Altman plot of differences between the photomicroscopy and iScan OCT measurements saphenous veins (n = 8). Mean difference = 116.24 μm. 95% limits of agreement = 1987 μm, −1754 μm.

Standard image High-resolution image

The mean variation of iScan OCT measurements of saphenous vein ex vivo from microscopy was 7.62% (min −65.02%, max 88.7%) with SD of 28.98%. The mean variation of OCT vendor measurements from microscopy was 16.74% (min −50.94%, max 79.63%) with SD of 25.73%.

Discussion

OCT vendor and iScan image analysis of phantom tubes and human saphenous veins imaged ex vivo exhibited good agreement with photomicroscopy, indicating a high degree of measurement accuracy. OCT may be acceptable for clinical purposes when imaging vein grafts. iScan software measurements of saphenous veins imaged by OCT had similar accuracy to vendor software supporting its validity for this purpose. Potentially, OCT may be used to measure saphenous vein dimensions in vivo [1722].

Linear regression analysis showed a strong positive relationship between OCT and photomicroscopy with 90% and 70.4% of the variance in OCT measurements explained by photomicroscopy measurements for phantom and saphenous veins respectively. Remaining variance may be explained by outliers due to iScan errors in detecting the inner lumen or distortion of the saphenous veins during preparation for photomicroscopy.

The Bland–Altman plots of diameter demonstrated positive agreement and minimal bias for OCT, 104 μm and 116.24 μm respectively for phantom grafts and saphenous veins.

The mean variation of iScan OCT measurements of phantoms from microscopy was 3.02%, very similar to the OCT vendor mean variation 3.03%. However iScan had a larger range and standard deviation than vendor measurements and so consistency is an area for improvement in iScan.

The mean variation of iScan OCT measurements of saphenous veins from microscopy was 7.62%, while OCT vendor measurements was 16.74%. iScan and OCT had similar ranges of variation but iScan had a slightly larger standard deviation. This larger variation in veins as compared to phantoms (3.02% and 3.03%) may be explained by possible distortion of the lumen during cutting. This is discussed further under limitations.

OCT iScan versus OCT vendor: comparison of imaging results according to vessel diameter (and depth of field). We found that the mean variations of iScan OCT measurements from microscopy was 3.02% and that of the OCT vendor software 3.03%, and so were similar. This result implies that the accuracy of the iScan software application may be at least as good as a current commercially available analysis tool, however, iScan needs to be improved to reduce the variation of its measurements. We have also shown that as the saphenous vein diameter increases so does the variance with iScan, as compared with photomicroscopy reference measurements (figure 7).

Limitations

The phantom was cut at e.g. 30 mm along the phantom as seen by the naked eye but this may have actually been at e.g. 30.1 mm. The phantom was fixed with tape to a measuring scale and cut at 2 mm intervals with an average slice thickness of 0.5 mm. An automated cutting device would have removed this limitation but was not available.

The specimen lumen was in most instances circular during photomicroscopy, as could be verified by the circular target superimposed. However due to the preserving solutions used on the saphenous veins making them firmer than in vivo, the circularity of some sections were distorted during cutting. We tried to minimize this error by readjusting the section to the circular target and also by taking an average of the ten diameter measurements to reduce possible bias.

There were instances where iScan did not detect the entire border of the images lumen as artefacts e.g. guidewire reflections, were detected instead of the inner border. As diameter measurements were an average of five frames, this will have minimized a measurement error if for instance, the guidewire was detected instead of the lumen wall. The algorithm to rule out the guidewire shadow is being constantly improved upon, and this should not be an issue in future versions.

The mean difference of measurements of saphenous veins from microscopy was smaller for iScan than the vendor software however, the 95% limits of agreement was smaller with the vendor software than with iScan. We believe the increased variances or 95% limits of agreement with iScan were mainly due to the section alignment. The OCT measured sections were perpendicular to the guide wire; whereas the photomicroscopy measured sections were perpendicular to the middle axis of the vessel. Since the guide wire was not necessary aligned with the middle axis, the larger the diameter, the bigger the discrepancy of the sections measured.

Conclusion

This research has demonstrated that OCT is accurate when measuring inner diameter and this has been validated by photomicroscopy. OCT has also been shown to detect even the earliest stages of intimal thickening; this can be seen on OCT as bright tissue with a homogeneous thin rim [4]. However, there is currently no established cut-off value for what categorizes neointimal growth as pathologic [9]. OCT's accuracy makes it most suitable to measure neointima and so potentially OCT may be used to assess the burden of disease in vein grafts for clinical and research purposes.

Acknowledgments

We would like to thank the College of Medicine, Veterinary Medicine and Life Sciences at the University of Glasgow who provided a consumables budget for this study. We gratefully acknowledge The Pathological Society who kindly gave a summer research grant to R W, enabling further work on this area. Thanks go to Dr Quentin Fogg from the Anatomy Department of University of Glasgow for providing the cadaveric vessels and Shotts Abattoir for providing swine's hearts. We also thank the staff in the Beardmore Centre for Health Sciences in the Golden Jubilee National Hospital. Professor Berry was supported by a Senior Fellowship from the Scottish Funding Council. This research was supported by a grant from the Medical Research Council (MRC Reference: G1001147/1).

Disclosure

Professor Berry and the University of Glasgow hold research and consultancy agreements with St Jude Medical.

Please wait… references are loading.
10.1088/2057-1976/2/2/025002