An efficient method to improve the spatial resolution of laboratory X-ray diffraction contrast tomography

, and

Published under licence by IOP Publishing Ltd
, , Citation H Fang et al 2019 IOP Conf. Ser.: Mater. Sci. Eng. 580 012030 DOI 10.1088/1757-899X/580/1/012030

1757-899X/580/1/012030

Abstract

Lab-based diffraction contrast tomography (LabDCT) has recently enabled nondestructive 3D characterization of the crystallographic orientations and grain morphologies in bulk materials. Despite the wide accessibility and availability of lab-based X-rays, the current spatial resolution of LabDCT is only about 20 μm and has to be improved to make this technique a more versatile tool. Conventional LabDCT takes advantage of the Laue focusing effect, i.e. the diffracted beam from a sample illuminated by an incoming polychromatic beam, is focused when the source and the detector are symmetrically positioned. In this study we have investigated the possibility of increasing the spatial resolution of LabDCT by placing the detector at a distance beyond that required for Laue focusing. We first developed a forward projection simulation tool to elucidate the effects of placing the detector at a larger distance. Next, we performed LabDCT measurements on a partially recrystallized pure aluminum sample at different sample-to-detector distances. The results show that the diffraction spots can be magnified by a factor of about three compared to the current LabDCT setup. The benefits and limitations resulting from increasing the sample-to-detector distance are discussed.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1757-899X/580/1/012030