Effects of Temperature on the Characterisation of a New Design for a Non-Articulated Prosthetic Foot

, , and

Published under licence by IOP Publishing Ltd
, , Citation Muhsin J. Jweeg et al 2018 IOP Conf. Ser.: Mater. Sci. Eng. 433 012064 DOI 10.1088/1757-899X/433/1/012064

1757-899X/433/1/012064

Abstract

This paper presents a new design for a non-articulated prosthetic foot made from high density polyethylene. The mechanical and physical properties of the proposed design are then compared with a Solid Ankle Cushioned Heel (SACH) foot based in a numerical and experimental investigation. The prosthetic foot failure characteristics at room temperature and at a relatively high temperature (60 °C) were determined using a fatigue foot tester and the dorsiflexion angle was measured using inclined compression. The numerical solution can be used to find the safety factor for the new design using SolidWorks software and ANSYS. From the results in this case, the maximum dorsiflexion angle at toe off for the SACH and new prosthetic foot are 4.7° and 6.3°, respectively. Fracture of the new prosthetic foot occurs at 1,1047,321 and 1,089,463 cycles for tests performed at room temperature (23 °C) and at (60 °C) respectively.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1757-899X/433/1/012064