High-voltage monitoring with a solenoid retarding spectrometer at the KATRIN experiment

, , , , , , , , , , , , , , , , , , , , , , and

Published 26 June 2014 © CERN 2014
, , Citation M Erhard et al 2014 JINST 9 P06022 DOI 10.1088/1748-0221/9/06/P06022

1748-0221/9/06/P06022

Abstract

The KATRIN experiment will measure the absolute mass scale of neutrinos with a sensitivity of mν = 200meV/c2 by means of an electrostatic spectrometer set close to the tritium β-decay endpoint at 18.6keV. Fluctuations of the energy scale must be under control within ±60mV (±3ppm). Since a precise voltage measurement in the range of tens of kV is on the edge of current technology, a nuclear standard will be deployed additionally. Parallel to the main spectrometer the same retarding potential will be applied to the monitor spectrometer to measure 17.8-keV K-conversion electrons of 83mKr. This article describes the setup of the monitor spectrometer and presents its first measurement results.