Impact of the gas choice and the geometry on the breakdown limits in Micromegas detectors

, , , , and

Published 12 June 2023 © 2023 The Author(s)
, , Citation P. Gasik et al 2023 JINST 18 C06011 DOI 10.1088/1748-0221/18/06/C06011

1748-0221/18/06/C06011

Abstract

In this study we investigate the stability limits of Micromegas detectors upon irradiation with alpha particles. The results are obtained with meshes with different optical transparency and geometry of wires. The measurements are performed in Ar- and Ne- based mixtures with different CO2 content. We observe that the breakdown limit strongly depends on the gas and that a higher amount of quencher in the mixture does not necessarily correlate with higher stability. In addition, we observe discharge probability scaling with the wire pitch. This suggests that a Micromegas mesh cell can be treated as an independent amplification unit, similar to a hole in a GEM foil. The outcome of these studies provides valuable input for further optimization of MPGD detectors, multi-layer stacks in particular.

Export citation and abstract BibTeX RIS

Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.