paper

Response of the GAGG(Ce) scintillator to charged particles compared with the CsI(Tl) scintillator

, , , , , , , , and

Published 12 October 2021 © 2021 IOP Publishing Ltd and Sissa Medialab
, , Citation T. Furuno et al 2021 JINST 16 P10012 DOI 10.1088/1748-0221/16/10/P10012

1748-0221/16/10/P10012

Abstract

GAGG(Ce) is a novel scintillator with a fast response and high light output without a hygroscopic nature. It is expected to be a useful detector for charged particles at high-counting rates. However, the response of the GAGG(Ce) scintillator to charged particles has not been fully examined. In the present work, the light output and energy resolution of the GAGG(Ce) scintillator were measured for protons and alpha particles at Ep=5–68 MeV andEα=8–54 MeV as well as gamma rays atEγ=662 keV from a 137Cs source. The results were compared with those of the CsI(Tl) scintillator. The scintillation efficiencies dL/dE of the GAGG(Ce) and CsI(Tl) scintillators were obtained and parametrized as a function of linear energy transfer dE/dx.

Export citation and abstract BibTeX RIS