The following article is Open access

Results from the first cryogenic NaI detector for the COSINUS project

, , , , , , , , , , , , , , , , and

Published 8 November 2017 © 2017 The Author(s)
, , Citation G. Angloher et al 2017 JINST 12 P11007 DOI 10.1088/1748-0221/12/11/P11007

1748-0221/12/11/P11007

Abstract

Recently there is a flourishing and notable interest in the crystalline scintillator material sodium iodide (NaI) as target for direct dark matter searches. This is mainly driven by the long-reigning contradicting situation in the dark matter sector: the positive evidence for the detection of a dark matter modulation signal claimed by the DAMA/LIBRA collaboration is (under so-called standard assumptions) inconsistent with the null-results reported by most of the other direct dark matter experiments. We present the results of a first prototype detector using a new experimental approach in comparison to conventional single-channel NaI scintillation light detectors: a NaI crystal operated as a scintillating calorimeter at milli-Kelvin temperatures simultaneously providing a phonon (heat) plus scintillation light signal and particle discrimination on an event-by-event basis. We evaluate energy resolution, energy threshold and further performance parameters of this prototype detector developed within the COSINUS R&D project.

Export citation and abstract BibTeX RIS

Published by IOP Publishing Ltd on behalf of Sissa Medialab. Original content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1748-0221/12/11/P11007