Abstract
The changes of the SAS program have been made as an attempt to adapt it to a two-detector system for the YuMO spectrometer. It is shown that applying weight algorithm technique for the raw data averaging procedure allows one to avoid a "sewing" problem resulting from using different detectors in rather wide q-range. The operating scheme and the background details of the experiment are discussed as well.
Export citation and abstract BibTeX RIS