Abstract
Radiotherapy treatment depends on the accuracy of the dose delivery to patients, the purpose of the study is to verify the dose in IMRT and VMAT technique in prostate cancer cases correspond to TPS dose using phantom base on ICRU No.50. The dose verification of the target and OAR was performed by placing the TLD Rod LiF100 and EBT2 Gafchromic film at slab hole of pelvic part of the Alderson RANDO phantom for prostate cancer simulation. The Exposed TLDs was evaluated using the TLD Reader Harshaw while EBT2 film was scanned using Epson scanner. The point dose measurements were compared between planned dose and measured dose at target volume and OAR. The result is the dose difference at target volume, bladder and rectum for IMRT and VMAT are less than 5%. On the other hand, the dose difference at the Femoral head is more than 5% for both techniques because the location of OAR already in low gradient dose. Furthermore, the difference dose of the target volume for IMRT technique tends to be smaller than VMAT either for TLD and EBT2 film detectors. From the measurement showed that the delivered dose on the phantom simulation match with ICRU No.50 criteria.