Evaluation of the ability of a 2D ionisation chamber array and an EPID to detect systematic delivery errors in IMRT plans

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation Omemh Bawazeer et al 2014 J. Phys.: Conf. Ser. 489 012071 DOI 10.1088/1742-6596/489/1/012071

1742-6596/489/1/012071

Abstract

Two clinical intensity modulated radiotherapy plans were selected. Eleven plan variations were created with systematic errors introduced: Multi-Leaf Collimator (MLC) positional errors with all leaf pairs shifted in the same or the opposite direction, and collimator rotation offsets. Plans were measured using an Electronic Portal Imaging Device (EPID) and an ionisation chamber array. The plans were evaluated using gamma analysis with different criteria. The gamma pass rates remained around 95% or higher for most cases with MLC positional errors of 1 mm and 2 mm with 3%/3mm criteria. The ability of both devices to detect delivery errors was similar.