Development of an automated routine for the calibration of multi-point scintillation detectors for advanced dosimetry applications

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation B Lessard et al 2022 J. Phys.: Conf. Ser. 2167 012028 DOI 10.1088/1742-6596/2167/1/012028

1742-6596/2167/1/012028

Abstract

Multi-point scintillation dosimeters have the advantage to allow real-time measurements of dose with a good spatial resolution, but these detectors must be accurately calibrated to perform precise dose measurements. This study demonstrates, by means of simulations and an experimental validation, that when the calibration dataset is well chosen, it is possible to perform an automated calibration of such dosimeters, with an accuracy on the measured output factors reaching (0.5 ± 0.1) %.