This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.
Brought to you by:
Paper The following article is Open access

Three-dimensional reproducibility assessment of radiochromic ClearView™ gel in custom vessel

, and

Published under licence by IOP Publishing Ltd
, , Citation Kevin Jordan et al 2022 J. Phys.: Conf. Ser. 2167 012013 DOI 10.1088/1742-6596/2167/1/012013

1742-6596/2167/1/012013

Abstract

Samples of the radiochromic hydrogel ClearView™ were characterized for 3D dosimeter performance in custom vessels and read with a modified Vista16 optical cone beam scanner as part of a small field dosimetry study. Uniform dose responses and small field depth doses revealed each sample was unique. As the gels aged over several weeks at 4 °C, the dose sensitivity decreased. Reconstructions of uniform dose distributions revealed unique features near the vessel interface. The central sub-volumes ranged from uniformity to within 2% throughout the sub-volume to a 12% gradient from base to top of gel. Comparisons of small field depth doses revealed that if the central volume had a uniform dose response, no correction was required. However, if a gradient in sensitivity from bottom to top was measured a correction may or may not be required. It is suspected that the filling protocol of the custom vessels is a factor in the observed variations in 3D dose response.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/2167/1/012013