Post-processing techniques using 3D Slicer for T1-weighted MRI analysis of radiochromic gel dosimeters

, , and

Published under licence by IOP Publishing Ltd
, , Citation Hannah J. Lee et al 2022 J. Phys.: Conf. Ser. 2167 012001 DOI 10.1088/1742-6596/2167/1/012001

1742-6596/2167/1/012001

Abstract

Radiochromic FXG gel dosimeters were investigated for end-to-end validation of TG-119 plans delivered in the pre-clinical Elekta 1.5 T-7 MV FFF MR-Linac. Due to the noise levels in the T1-weighted MR images, this preliminary study investigated post-processing techniques available in 3D Slicer and their impact on gamma passing rates. Binomial blur and discrete Gaussian image filters were identified as the most impactful on increasing gamma passing rates for T1-weighted MR images; however, any application of image filters should be implemented with caution on the interpretation of the results.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.