This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.
Paper The following article is Open access

Design and Application of Automatic Calibration Device for Multi-Channel Resistance Strain Gauge Indicator

, , , and

Published under licence by IOP Publishing Ltd
, , Citation Xiaoyin Hu et al 2021 J. Phys.: Conf. Ser. 2113 012040 DOI 10.1088/1742-6596/2113/1/012040

1742-6596/2113/1/012040

Abstract

In this paper, an automatic calibration device for multi-channel resistance strain gauge indicator is designed and its applicability and measurement accuracy are verified at laboratory. The calibration done by original resistance bridge calibrator is time-consuming for its manual operation and complex calibration process. With the intent to increase calibration efficiency, an automatic channel switch device was developed, and the resistance bridge calibrator was automated. The designed calibration device is completely computer controlled enabling a sequence of unmanned measurements. The calibration device was verified at laboratory that the maximum of error is 0.072%. It was applied to calibrate a 60-channel resistance strain gage indicator to approve its practical applicability. The result shows that the designed calibration device can realize automatic calibration and the efficiency is increased by 40%.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/2113/1/012040