Abstract
The voltage of thermocouple during process contribution is determined by the absolute thermal electromotive force coefficient in each wire and the temperature difference across the section. This paper described the development of a system as a local heating method for measurements of inhomogeneities in type T thermocouples with various temperature scanning using hot air. The system could control automatically measurement voltage, temperature, distance, and speed of the stepper motor for moving the hot air scanning along with the thermocouple and control with temperature monitoring. Both the measuring and reference junctions were using an ice bath kept at 0 °C. In this paper calculation of interpretation of the measurement results of inhomogeneity using movable heating zone system of type T thermocouples and the system could detected inhomogeneity of thermocouple and there have different characteristics of voltage and seebeck effect.
Export citation and abstract BibTeX RIS