Paper The following article is Open access

Calibration of WLI Lateral Indication Error with 2D Micro/Nano Pitch Standard

, and

Published under licence by IOP Publishing Ltd
, , Citation Jiasi Wei et al 2021 J. Phys.: Conf. Ser. 1732 012138 DOI 10.1088/1742-6596/1732/1/012138

1742-6596/1732/1/012138

Abstract

White light interferometer (WLI) is one of the common machines that are used to characterizing surface structures. The indication errors of WLI are key metrologic characteristics, which may affect evaluation result of WLI. In this paper, a two-dimensional micro/nano pitch (2D pitch) standard is introduced to calibrate the lateral indication error of WLI. The calibrating experiments are carried out in four measurement positions. The results and the uncertainty is anlysized. And the expanded uncertainty of WLI in lateral measurement (x50) is evaluated as tens of nanometer.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/1732/1/012138