Electron Paramagnetic Resonance Spectroscopy for Gamma Knife Dosimetry

, , , and

Published under licence by IOP Publishing Ltd
, , Citation P Ramachandran et al 2020 J. Phys.: Conf. Ser. 1662 012026 DOI 10.1088/1742-6596/1662/1/012026

1742-6596/1662/1/012026

Abstract

Alanine chips exposed to high doses of radiation produces long lived free radicals that could be easily measured with electron paramagnetic resonance (EPR) spectrometers. In this study, the feasibility of using alanine dosimeters for performing rapid quality assurance of Leksell Gamma Knife (LGK) treatment plans was demonstrated. A 3D printed grid was placed inside the LGK spherical solid water phantom (SWP) for measurement of doses at isocentre and off-axis points. The EPR spectroscopy was performed on a Magnettech MS-5000 EPR/ESR spectrometer. A set of dose calibration curves were established prior to the use of alanine chips for LGK dosimetry. Absolute dose, transit dose and dose/timer linearity were performed with the alanine chips positioned at the centre of the LGK solid water phantom (SWP). Five patients of different sites were selected, and patient specific quality assurance (PSQA) was performed in the LGK SWP. The absolute dose measured with the EPR alanine dosimeter agreed well within 2% of the ion chamber results and PSQA results were within 2.1%. Alanine-based EPR dosimetry offers rapid dose measurement with high accuracy and can also be used as a dosimeter for Gamma Knife PSQA.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/1662/1/012026