Abstract
A programmable quantum current generator based on the application of Ohm's law to quantum voltage and resistance standards has demonstrated a realization of the ampere from the elementary charge with a 10−8 relative uncertainty (Brun-Picard et al 2016 Phys. Rev. X 6 041051). Here, we report on improvements of the device leading to a noise reduction of the generated quantized current. The improved quantum current generator is used to calibrate different ammeters with lower measurement uncertainties. Besides, measurements of its quantized current using a calibrated ultrastable low-noise current amplifier have shown that the realizations of the ampere at PTB (Physikalisch-Technische Bundesanstalt) and LNE (Laboratoire national de Métrologie et d'essais) in the range ±50 μA agreed to −3.7 parts in 107 with a combined standard uncertainty of 3.1 parts in 107 (coverage factor kc = 1).
Export citation and abstract BibTeX RIS
Original content from this work may be used under the terms of the Creative Commons Attribution 4.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.
1. Introduction
Since 20 May 2019, the ampere definition has been related to the quantum description of the electrical transport and not to a mechanical force anymore. As the numerical value of the elementary charge e has been fixed, a current of one ampere is the flow of 1/1.602 176 634 × 10−19 elementary charges per second [2]. This new definition has a big impact on the realization of the ampere [3] which can be based on solid state quantum effects, and in particular on application of Ohm's law to the Josephson voltage standard and the quantum Hall resistance standard, which are realizations of the volt and the ohm, directly related to defining constants of the SI: the Planck constant h and the elementary charge e. Implementing experimentally Ohm's law to quantum electrical standards, we have demonstrated that currents could be quantized in terms of e within a relative uncertainty of 10−8 [1]. This 'mise en pratique' of the ampere combines a programmable Josephson voltage standard (PJVS), a quantum Hall resistance standard (QHRS) and a cryogenic current comparator (CCC). The high accuracy of the generated current relies not only on the universality of the Josephson effect and the quantum Hall effect but also on the CCC accuracy itself. The latter is used as a highly accurate current amplifier with a gain span of more than four orders of magnitude. Quantized currents from mA to μA levels can therefore be generated with the same expected accuracy.
For small currents, below 1 nA, single electron current sources offer another way for the quantum realization of the definition of the ampere by controlling the flow of electrons one by one at a clock frequency (for recent reviews [4–7]). However, their principle of operation is based on the single electron tunnelling which is a stochastic process, making the control of the electron flow difficult. Hence achieving an accuracy better than 10−7 remains very challenging [6, 8].
For currents above 1 nA, more recently different groups also proposed developments based on Ohm's law involving quantum electrical standards, either to improve current measurement uncertainty in the μA range below 10−7 by current-to-voltage conversion with a quantum Hall resistance array [9] or to improve the stability of an externally referenced current source down to 1 nA A−1 at 50 mA with a PJVS [10]. On the other hand, National Metrology Institutes apply Ohm's law to secondary voltage and resistance standards, traceable to the Josephson and quantum Hall effects, respectively. However, the uncertainties claimed in their calibration and measurement capabilities (CMC) for currents are about 10−6. In this context, the recent developments of the ULCA at PTB [11] based on Ohm's law tackle this limit for current levels up to 50 μA. The first version of the amplifier was specially focussed on small currents ⩽5 nA, since it was dedicated to amplify and measure the very small current generated by single-electron pumps, but recent developments increased the range up to 50 μA [12]. The predicted relative uncertainties for the ULCA devices from the nA range to the μA range is 2 parts in 107 for timely calibrations (kc = 1). All these developments have opened the way to new current comparisons targeting measurement uncertainties of few parts in 107.
The first demonstration of the PQCG showed very promising results, however, making it a primary current standard needs to ensure a very reliable operation. The reduction of noise and instability are essential for that purpose. At first, the paper details the noise characterizations and the modifications implemented to improve the set-up. It then reports on ammeter calibrations carried out following well established quantization criteria, that notably demonstrate an improvement of the measurement uncertainties in the mA range. Finally, a first direct comparison of the ampere realizations performed with measurement uncertainties of a few parts in 107 at 50 μA is presented.
2. Improving the stability and reliability of the set-up
2.1. The set-up
The principle of the programmable quantum current generator is presented in figure 1. It is based on a primary circuit, the programmable quantum current standard (PQCS), highlighted by the light yellow box, where the quantum reference current IPQCS is circulating. This circuit is magnetically linked by the cryogenic current amplifier to a second circuit that drives the output current of the PQCG, IPQCG. In the reference loop, at the superconducting pads of the PJVS, quantized voltage steps, with KJ = (2e/h), are generated when nJ Josephson junctions are dc biased and driven at a microwave frequency fJ. This voltage is applied to the QHRS through a double connection scheme [13, 14]. Doing so, and thanks to the properties of the quantum Hall effect [13, 14], the equivalent resistance R connected to the superconducting pads is equal to R = RH(1 + α), where RH is the quantum Hall resistance, which is equal to RK/2 at Landau level filling factor ν = 2 with RK = h/e2 the von Klitzing constant, and where is a second order correction due to the cable resistances r. This correction, which is significatively reduced compared to the usual first order correction of a simple connection corresponding to r/RH ∼ 3 × 10−4, can be determined with a relative uncertainty of a few parts in 109.
The reference current is circulating into the PJVS. It can also be written as IPQCS = nJ efJ(1 − α), which emphasizes the relation to the elementary charge. This current is divided into two separate branches in the double connection to the QHRS. In order to detect and amplify the sum of these currents, we use a CCC, which insures the ampere-turns balance of several superconducting windings by means of the SQUID-based feedback loop. Thanks to the introduction of two CCC windings of identical number of turns, NJK, into the double connection between the PJVS and the QHRS, the balance between the primary and secondary circuit, NJK IPQCS − NIPQCG = 0, leads to . The gain is exact within one part in 109 for all windings and within few parts in 1010 for the windings used in the present study [15]. The current IPQCG is generated by an external battery-powered low-noise current source which is servo-controlled by the feedback voltage of the SQUID that monitors the flux imbalance. An adjusted pre-balance current (or compensation current), Icomp, driven by the voltage Vout supplied by a digital-to-analog converter of the programmable current bias source of the PJVS, is added to the feedback current of the SQUID to reduce the set-point error due to the finite open loop gain of the SQUID electronics [1]. The current IPQCG circulates in the load which can be a resistance RLoad, an ammeter or the ULCA.
The PJVS used in this work is the PJVS♯A from reference [1]. It is based on 1 V Nb/Al/AlOx
/Al/AlOx/Al/Nb Josephson junction series arrays fabricated at the PTB [16, 17]. The QHRS is described in reference [18] and the CCC details can be found in [15]. More details are given in appendix
In our previous work [1], the correct operation of the PQCG was achieved by solving several technical issues. As the PJVS, the QHRS and the CCC are implemented in three different cryostats and connected by long shielded cables, it makes the PQCG very sensitive to external electromagnetic noise. The stability issues are more severe for high ampere-turns values. For instance, the SQUID became unstable for NJK > 2 when the PQCS circuit was connected. However, increasing NJK is relevant to improve the signal-to-noise ratio of the PQCG, which is given by the ratio of the flux generated into the SQUID by the primary circuit, NJK IPQCS/γCCC (γCCC being the flux to ampere-turn sensitivity of the CCC), to the total flux noise . To solve this problem, a damping circuit was implemented that stabilizes the CCC operation. Since it cannot be added directly on the reference circuit without introducing systematic errors, a series arrangement of a resistance RD and a capacitance CD has been connected to a spare winding of ND turns [19] as shown in figure 1. However, the damping circuit introduced increased noise in the 100 Hz–10 kHz bandwidth, the consequence of which was not analysed in detailed at that time. Besides, it remained impossible to increase NJK above 128. Moreover, the low-frequency noise was significantly higher than the SQUID specifications. Finally, the repeatability of the PQCG turned out to be restricted due to flux frequently trapped in the Josephson array during the measurement campaigns. In the following, we report on investigations about noise, stability issues and resonance damping, and we will present the improvements implemented to tackle the previous limitations of the PQCG.
2.2. Noise and stability issues
2.2.1. Analysis of the noise of the system
The SQUID is an ultra sensitive magnetic flux-to-voltage converter, and hence interfering electromagnetic signals are critical: both, low-level high-frequency noise and sharp transients, can be detrimental to the operation of the system. In the PQCG, the noise to which the SQUID is exposed can have different origins: (i) direct flux noise picked up by the SQUID, (ii) flux noise coming from the currents that circulate in the CCC windings due to Johnson–Nyquist noise emitted by the resistive elements connected to them, and (iii) external electromagnetic noise picked-up in the circuit through capacitive or inductive couplings. Moreover, the CCC is a complex system of 15 magnetically coupled superconducting windings with distributed capacitances and with inductances Li
∼ i2
LCCC, where LCCC is the inductance of the toroidal shield and i the number of turns. For example, for the windings of 2065 turns, L2065 = 85 mH and the equivalent parallel capacitance is 1 nF. The total flux noise can be deduced from the power spectral density , where SSQUID(f) and SCCC(f) are the respective contributions from the noise of the SQUID (intrinsic or captured), and from the noise of the CCC and the circuit connected to it. The latter is the most difficult to analyse. Thus, to have a better understanding, a model of the circuit is useful. Here we have used a simplified model of two magnetically coupled series-RLC circuits as a generic model to account for the main features of the spectra (see appendix
Figure 2(a) compares the noise spectra of the CCC not connected (bare CCC) and connected to the QHRS, without the damping circuit. For the present study, we have performed noise measurements on the output voltage of the dc-SQUID when operating it in internal feedback mode and with the highest bandwidth (0.75 V/ϕ0 sensitivity and 50 kHz bandwidth). We present the results in terms of the flux noise, , expressed in .
Download figure:
Standard image High-resolution imageFor the bare CCC, two resonances appear at 12.6 kHz and 36.4 kHz. They can be identified as the LC-resonances resulting from the magnetic coupling of the two identical windings of 2065 turns short-circuited by their parasitic capacitances (see figure 2(b) and appendix
The noise spectra when the QHRS is doubly connected to the windings of 160 and 1600 (red and green lines of figure 2(a)) have been measured to clarify the problem of the reduced stability when increasing NJK. In this case, at low frequency, the bottom white noise levels are well explained by considering both contributions of the SQUID noise and the Johnson–Nyquist noise of the QHRS , where k is the Boltzmann constant and T the temperature. The latter becomes dominant for 1600 turns and amounts to 15 μϕ0/. To avoid this increased noise, an optimum value of NJK would be about 500 turns. However, the most striking feature is the almost linear increase from 100 Hz up to 10 kHz, on which 50 Hz harmonics are superimposed. This extra noise, the magnitude of which scales with NJK, is explained by a coupling of the circuit with external electrical noise sources. It can be modelled by a voltage noise source coupled to the circuit through the capacitance to ground of the 15 m long shielded twisted pairs connecting the CCC and the QHRS [22] as shown schematically in figure 2(c). This capacitance has been measured and amounts to 4 nF. To account for the increased noise in the frequency range 100 Hz to 10 kHz, we have simulated the quantum circuit as a simple RL-circuit considering a single winding connected to the QHRS. A 4 nF capacitance in series with a noise source is connected in parallel to the inductance of the winding as shown in figure 2(d). The red and green hatched areas in figure 2(a) correspond to adjustments of the effective white noise voltage source to (30 ± 10) nV , for both NJK = 160 and 1600. The good agreement confirms that the SQUID instability observed for large values of NJK comes from the existence of external noise sources that couple to the quantum circuit and excite the high-frequency resonances especially as large NJK values are used. Moreover, when the battery powered programmable Josephson bias source is connected, its additional digital noise is sufficient to cause the system to become unstable even in internal feedback mode. In future we will improve this situation by reducing the length of the cable between the CCC and the QHRS. For the moment this length is fixed by the current location of the cryostats in the laboratory.
2.2.2. Damping
Given the coupling of external noise to the PQCS, it was therefore crucial to damp the high frequency resonances [1] to stabilize the SQUID operation. The main question is how to choose the parameters of the damping. Experimentally, we observed that high values of ND enhanced the effect of the damping. The results of a simulation shown in figure 3(a) illustrates the increasing efficiency of the damping on the high frequency peak for three different values of ND from 16 to 2065. The simulation is based on the same generic model for the CCC, i.e. the series coupled RLC-circuit of in figure 2(b), in which the values of the resistance and the capacitance of one of the circuits have been replaced by the values used in reference [1], RD = 1 kΩ and CD = 100 nF (see figure 3(b)).
Download figure:
Standard image High-resolution imageSince the first implementation of the damping circuit in the PQCG with ND = 1600 [1], the winding of 2065 turns has then been favoured [20] not only to achieve a better efficiency but also to allow the use of the 1600-turns winding for NJK. At intermediate frequency (100 Hz < f < 10 kHz), the damping circuit is responsible for a damped resonance much broader than the self-resonance of the bare CCC [1, 19, 23]. As a first approximation, the center of the broad peak is at and its amplitude, due to the Johnson–Nyquist noise of RD, is given by [1]. Once the winding has been chosen, ND is fixed and hence LD is fixed. The choice of the capacitance CD is made such that the broad peak is well below 10 kHz so that the damping is efficient at 12 kHz. The last parameter, RD, fixes the amplitude of the broad resonance and the quality factor . Noise measurements have been performed for the bare CCC for three different sets of damping parameters [RD, CD] (RD CD = 10−4 s), with ND = 2065 (see figure 4(a)). Three situations are represented corresponding to Q ∼ 2.7, 0.9 and 0.3 respectively. The flux noise has been calculated, for the corresponding parameters, in the case of the damping circuit magnetically coupled to a winding of 2065 turns figure 4(b), and they account very well for the complex shapes of the spectra in the three cases.
Download figure:
Standard image High-resolution imageFinally, we have chosen RD = 1 kΩ and Q ∼ 1 as a good compromise because it suppresses the high quality factor resonance at 12.6 kHz, such that the main goal has been reached, and it limits the amplitude of noise towards the low frequency part. However, the amplitude of the broad peak is about 1 m. Two possible improvements can be considered, increasing of the resistance by a factor of 2 to reach the critical value Q = 1/2 and/or placing the resistance at 4.2 K. The latter point will be detailed in the present paper. Note that the resonance at 35 kHz did not disappear, it has been broadened and slightly shifted to 36 kHz; for the moment, this is not clearly understood, and the model cannot account for it because only two windings are simulated. It is possible detrimental effect on the stability may be limited anyhow by the damping, which attenuates part of the high-frequency components of the signal during the fast reversals that are likely to excite this resonance.
More unexpectedly, figure 4(a) shows that the low frequency part of the spectrum is enhanced from about 7 to 20 at 1 Hz compared to the bare CCC. This may be a consequence of mixing-down effects due to the high level of broadband noise in the kHz range. The SQUID is used in flux-locked loop (FLL) in order to amplify and linearize the periodic V − ϕ transfer function of the SQUID. The SQUID is then locked at a given working point of the V − ϕ characteristic by the feedback loop. At low frequencies, within the FLL bandwidth, the FLL suppresses the effect of the noise on the SQUID, in other words, the flux 'seen by the SQUID' is constantly compensated by FLL. However, if the noise above the FFL bandwidth is sufficiently strong, then the linearized flux range can be exceeded, leading to non-linear distortions, and consequently to rectified noise which will be superimposed with the low frequency feedback signal. Hence these mixing down effects deteriorate the signal-to-noise ratio [24] but, unless the amplitude of the noise is much higher, the SQUID is still locked at its working point.
This hypothesis has been confirmed recently, as the noise level at low frequency retrieves the expected one, as soon as the peak in the intermediate bandwidth is reduced by a factor of 8.5 by placing the resistance RD = 1 kΩ at 4.2 K, as shown in figure 4(b). The integrated noise between 100 Hz and 10 kHz can be used to quantify the detrimental noise level. For instance, it is equal to 60 mϕ0 for RD at ambient temperature (black line), whereas it falls at 7 mϕ0 for RD at 4.2 K (light blue line).
2.2.3. External feedback
In external feedback mode, a CCC winding is used as the feedback path instead of the dedicated modulation coil of the SQUID (internal feedback mode). It has a big impact on the dynamic of the system and the problems of stability are even more severe.
Compared to the internal feedback, the stability of the FLL is ensured by choosing an operation mode (5S) with a reduced bandwidth of 500 Hz as suggested in the documentation of the SQUID [25], i.e. characterized by a cutoff frequency much lower than the high frequency resonances. However, despite both this change and the fine adjustment within 10−6 of the current ratio IPQCG/IPQCS, the system proved to be unstable (i.e. frequent unlocking of the SQUID) for high ampere-turns reversals, which are needed during the PQCG operation to eliminate voltage offsets. This instability is related to the periodic nature of the V − ϕ characteristic of the SQUID, with several equivalent working points, each separated by a flux quantum. Fast reversals of the signal can force the SQUID to jump between adjacent working points if large signals are applied and if they change faster than the FLL can track. These problems are very similar to those encountered in different CCC resistance bridges [15, 19, 26], where two current sources feed the resistors to be compared, each being connected in series with a ratio winding of the CCC. A special care is paid to keep the bridge balanced during current reversals, by matching the filters on both sides of the bridge [26, 27] and by using offset, in-phase and in-quadrature correction circuits [15]. Other authors have implemented wideband feedback [19]. However, in the case considered here, because the two circuits connected to the CCC, the PQCS circuit and the output circuit, are biased by a voltage source and a current source, respectively, these adjustments are more difficult.
This instability has been overcome by modifying the closed-loop feedback gain. Let us detail the original external feedback circuit as illustrated in figure 5(a). The SQUID output voltage (mode 5S, bandwidth 500 Hz) is connected to the input of a high impedance differential amplifier to avoid loading the 500 kΩ resistance at the output of the SQUID electronics with the capacitance of the long cable plugged into the external current source. However a parasitic capacitance CSQ of 1.6 nF loading the resistance has been measured at the output of the circuit connecting the SQUID to the isolation amplifier. The same closed-loop feedback gain of = = 0.75 V/ϕ0 as in the internal feedback mode of the SQUID has been targeted for a winding of N = 16 turns; therefore, a feedback resistance RFB = 1.5 MΩ has been chosen. A small capacitance of 200 pF in series with a 20 kΩ resistor is connected in parallel to RFB to partially compensate the phase shift caused by the 1 kHz low-pass filter (RF and CF) present at the output of the current source [15]. Nevertheless, a residual phase shift exists in the feedback circuit [28] that translates into a peak in the frequency response of the feedback loop (dark green line of figure 5(b)), according to the calculated response based on the circuit presented in figure 5(a). From the simulation, it can be deduced that the phase shift comes mainly from the parasitic capacitance, CSQ. This certainly explains the instability against current reversals in the PQCG. By increasing the closed-loop gain to 4.5 V/Φ0 for the PQCG, the peak in the frequency response has almost disappeared, as shown by the green line in figure 5(b). This smooth response of the feedback with a slightly reduced bandwidth has solved the problem of stability against high ampere-turns reversals. In the future, a possible improvement would be to reduce the parasitic capacitance. However, the impact of the higher closed-loop gain on the accuracy of the generated current has been tested in [1] and it revealed negligible.
Download figure:
Standard image High-resolution image2.3. Set-up improvements
To reduce the common mode noise on the PQCS circuit, a common mode torus has been placed on the output of the PJVS, on the current leads that supply the QHRS, as depicted in figure 1. A similar solution had been implemented on the output of the voltage-controlled current source [15]. Guided by the reduction of the noise at frequencies below 10 Hz while ensuring the stability even for the highest ampere-turns, i.e. for NJK = 160 and 1600 when the CCC is connected to the PQCS, we found that the noise and therefore the stability of the PQCG is very sensitive to the position of the ground connection. In [1], in order to reduce the effect of leakage currents redirected to ground, it was connected at 4.2 K at the low voltage pad of the Josephson array and on the current contact of the QHRS. However, the relative error due to the leakage currents has been estimated to be less than 10−11, and moving the position to room temperature does not change much from this point of view. On the contrary, the ground connection at room temperature allows testing different positions and their effect on the noise of the circuit. We found that the best position, reducing the low frequency noise from 25 μV/ to 7 μV/, is at the head of the CCC as shown in figure 1, on the voltage contact of the QHRS. This certainly short-circuits a noise source that couples directly into the windings. The presence of the high inductance of the common mode torus on the current lead might explain that the noise couples less on this lead.
However, in this configuration, the amplitude of the voltage steps of the PJVS were reduced due to an increased noise that perturbed the array. This has been resolved by adding a 1 μF highly isolated capacitance placed between the 0 V of the battery powered Josephson programmable bias source and the ground (see figure 1). This new configuration is much more reliable, the array is less subject to flux trapping than it was previously with the ground connection at low temperature. Moreover, a heating system has been implemented at PTB on the Josephson array chip (see figure 1), and allows to get rid of trapped flux within few minutes.
Finally, after all these modifications, the system has gained stability in internal feedback mode. As shown in figure 6, the PQCS (including the PJVS bias source) can be connected notably to NJK = 160 (purple line) and 1600 (grey line) and the low frequency noise in both configurations is close to the expected levels. The flux noise of the broad resonance integrated between 100 Hz and 10 kHz is below 10 mϕ0 in both cases which is close to the 7 mϕ0 for the bare CCC (see figure 4(b)). However, in external feedback mode, the stability issues still limit the operation to NJK = 160.
Download figure:
Standard image High-resolution imageCompared to reference [1], not only the noise in the intermediate frequency range has been reduced but also the low-frequency noise below 10 Hz. The reduction of the latter by a factor of 2 and the increase of NJK by a factor of 1.25, may lead to a maximum increase of the signal-to-noise ratio by a factor of 2.5 for the most accurate measurements. However, this potential gain may be reduced if the excess noise (1/f noise) becomes the dominant noise at very low frequencies [29]. Note that the PQCG experiments aiming for highest accuracy are performed at current-reversal frequencies below 100 mHz. Hence, new measurements are needed to confirm the improvement in Type-A uncertainty of the PQCG [1]. Nonetheless, the reduction of the noise in the intermediate frequency range has already improved the Type-A uncertainty of digital ammeter calibrations as will be seen in the next section.
3. Digital ammeter calibrations
Here we present the calibrations of two precision digital ammeters (DA), an HP3458A (DA♯1) and a Fluke 8508A (DA♯2), in the mA ranges. During the calibrations, the ammeter replaces RLoad in figure 1. However to avoid perturbing the SQUID with the digital noise from the DA, a highly insulated PTFE capacitor (100 nF) is added at the current measurement terminals. The low-potential input is connected to ground. Prior to the calibration campaign, DA♯1 has been adjusted by using a 10 kΩ resistor standard and a 10 V Zener voltage standard calibrated in terms of RK and KJ, respectively. DA♯2 has not been adjusted.
Before each calibration, we systematically verify the quantization of the PJVS by checking the current amplitudes of the quantized voltage steps, , of each segments at the operating frequency and at the optimum microwave power. The relative correction α = 1.6 × 10−7 ± 2.5 × 10−9 for NJK = 160 is calculated by taking into account the resistances of the double connection.
Each calibration point takes less than 10 min. An on-off-on cycle is used to subtract the offsets, the measured value is calculated from eleven cycles as shown in the figure 7(a). We tried to realize as much as possible similar acquisition times for the two DAs despite the different parameter settings and the different dead times (see appendix
Download figure:
Standard image High-resolution imageThe relative Type-A standard uncertainty of each calibration point is evaluated by experimental standard deviations of the mean. This is justified as the on-off-on cycle is effective in getting rid of the offsets and in making the data sets uncorrelated. This can be seen in figure 7(b) from the relative Allan deviations [30], σI
/I[DA♯1] and σI
/I[DA♯2], for two time series realized with τ0 = 28 s and 25.4 s for DA♯1 and DA♯2, respectively (see appendix
The relative combined uncertainties of the calibrations are reported in figures 8–10. The Type-B contribution due to the finite resolution of the DAs amounts to 3 × 10−8 at 1 mA. The combination of all other Type-B contributions related to IPQCG, which are due to the CCC gain G, the Josephson frequency fJ and the correction α, is estimated well below 10−8 [1] and is therefore negligible.
Download figure:
Standard image High-resolution imageDownload figure:
Standard image High-resolution imageDownload figure:
Standard image High-resolution imageFigure 8(a) reports the calibration results of DA♯1 performed using different settings of the PQCG. Results are expressed as the relative deviation ΔI/I = (Imeas. − IPQCG)/IPQCG between the measured current Imeas. and IPQCG, for two current levels: ±0.574 mA and ±0.918 mA. At 1 mA, the relative deviation ΔI/I is −1.98 × 10−6. The stability of ΔI/I over few hours is illustrated in the inset of figure 8 from the repeatability of two sets of calibrations points. The dispersion between two identical measurements is of the order of 6 × 10−7. A linear fit over all calibration points gives a very small slope of −0.7 × 10−7 mA−1 which appears to be not significant with respect to the dispersion of the calibration points. This result indicates that the DA is linear within the stability of the instrument [31].
During a calibration campaign, the quantization criteria demonstrated in reference [1] have to be checked to ensure that the PQCG delivers quantized currents. They consist not only of the usual quantization criteria of the QHRS and of the PJVS but also of more specific criteria. The quantization of the generated current is tested: (i) by varying the PJVS bias current, (ii) by testing different combinations of Josephson array segments corresponding to the same number of Josephson junctions and (iii) by generating identical current levels with different numbers of junctions nJ and scaling the gain accordingly: for example, dividing nJ and multiplying the gain by the same factor should not change the measured current value. Figure 8(b) details the quantization criteria that have been tested during the calibration presented in figure 8(a). As is expected for a quantized current, the dispersion of the results obtained during the tests of the quantization criteria is of the order of the dispersion due to the instability of the DA observed during the calibration, and is much lower than the manufacturer's specifications [31].
Here, we would like to point out a notable difference with the previous results of reference [1], where the possibility to calibrate DAs on different ranges extending from μA to a few mA has been demonstrated. There, the tests of the quantization criteria highlighted an unexpected noise issue. It is illustrated in figure 9, which reports the relative combined uncertainties uc of the calibration points from reference [1] (red squares) in the 1 mA range as a function of the number of Josephson junctions nJ. Indeed, a clear dependence on nJ was observed, demonstrating that the noise was dominated by the PQCG rather than by the DA. This was rather unexpected as the relative Allan deviation expected for the PQCG alone, shown in figure 7(b), is an order of magnitude below the relative Allan deviation of the DAs. We attributed this to the Johnson–Nyquist noise caused by the damping resistance RD at 293 K. The RC-filter formed by the 100 nF capacitance on the input resistance of the DA [1] corresponds to a cutoff frequency of 16 kHz, which does not prevent the noise generated by the damping circuit from perturbing the system and overcoming the noise of the DA. The new uncertainties observed with RD at 4.2 K, reported as blue triangles in figure 9, are independent of nJ, which confirms that the dominant contribution in the uncertainty now comes from the DA.
Figure 10(a) shows the results of the calibration of DA♯2 in the 2 mA range. IPQCG has been generated with the gain G = 160/4 and with nJ = 3072, 2048 and 1024, corresponding to current levels ±1.378 mA, ±0.918 mA and ±0.459 mA respectively. The dispersion of three calibration points (over 2 h) is about 1 × 10−7 at −0.9 mA [31]. This allows the unveiling of a small but significant non-linearity of DA♯2, revealed by a linear dependence of ΔI/I in the measured current range with a slope of +1.6 × 10−7 mA−1.
Figure 10(b) shows the comparison of the calibration of figure 10(a) with the calibration performed at LNE using secondary resistance and voltage standards with a relative uncertainty of (5 × 10−6 + 1 × 10−9/Imeas.). Both calibrations are in good agreement within the measurement uncertainties despite the 6 months period separating them. This might indicate a good long-term stability of the DA, better than the manufacturer's specifications, which would have to be confirmed in a further study.
These calibration examples show that the PQCG is a precious tool which allows characterizing thoroughly precision DAs with relative uncertainties close to 10−7 while reducing the traceability chain. By doing regular calibrations with the PQCG, a better knowledge of the short and long term stabilities of such precision DAs can be gained. Hence, if regularly calibrated by the PQCG, they could be used as secondary standards with uncertainties better than the manufacturer specifications with the goal of improving the CMC.
4. ULCA comparison
4.1. Context
In this last section, we present the results of the direct comparison of the current measurements performed using the ultrastable low-noise current amplifier (ULCA) and the PQCG. The ULCA can be used for two purposes: current measurement or current generation, i.e., for current-to-voltage or voltage-to-current conversion. It is based on a two-stage design: a first stage provides a current gain GI of 1000 and a second stage provides a current-to-voltage conversion with a transresistance RIV (typically, 1 MΩ for the 'standard' ULCA). Thanks to this design, the device can be operated either in normal or extended mode [12] which differ by the factor of 1000 in current range; for example, in the case of the 'standard' ULCA, the 5 nA range (normal mode) and the 5 μA range (extended mode) are accessible. Several versions of the ULCA have been developed [12]. Among the second generation of ULCAs, the prototype of a high-accuracy variant (RI V -nominal value 100 kΩ) offers the highest current range compatible with the PQCG (±50 μA in the extended mode).
Thanks to its small size and compact design, the ULCA is perfectly suited as a travelling standard [32, 33]. The use of an ULCA as a travelling standard has been exercised first in an interlaboratory comparison between three national metrology institutes including LNE and PTB in 2015 with a two-channel 'standard' ULCA [32] operated in the 5 nA range. The calibration history of the travelling standard including the effects of lab-to-lab transports are crucial information; for instance, the history of the device which has been used in [32] has been recorded over more than five years [34]. However, comparably reliable information on the stability of the chosen high-accuracy ULCA, especially when travelling, was not available at the time of the first experiments reported below. The shipment from PTB to LNE was done by a commercial carrier and for the return transport, a PTB car has been used.
4.2. Description of the comparison and results
The comparison has been done in extended mode in the 50 μA range at two current levels 43.75 μA and 21.875 μA. Similarly to the calibration of the DAs, IPQCG is directly fed into the ULCA (i.e. into the current-to-voltage conversion stage). Different from the work with the DAs described in the previous section, no filter was necessary on the input of the ULCA. Introducing much less noise, the latter did not affect the SQUID's stability. The calibration of the total transimpedance ATR = GI RIV , which is performed at PTB in two steps, has been done before and after travelling to LNE. The gain GI is directly calibrated by means of a CCC and the transresistance RIV is traced back to the quantum Hall resistance by means of a CCC-based resistance bridge [32]. Note that for the present comparison with the ULCA operated in extended mode throughout, only the value of RIV is relevant.
The multiple connection of the PQCG has been implemented. However, to even further reduce the absolute value of the correction, we have 'short circuited' part of the cable resistances by using the 'local triple connection' on one side of the QHRS (dashed connection on figure 1). That is, an additional connection is made between the voltage pad of the QHR and the top of the cryostat rather than directly on the superconducting pad of the PJVS. From the measurements of the resistances of the cables, we have estimated the relative correction α = 1.35 × 10−7 ± 2 × 10−9.
The current measured by the ULCA is given by . It is determined from the value of the transresistance RIV
estimated at the time of the comparison from the calibration history determined at PTB (see appendix
In that way, we compare on one side the current measured by the combination of the ULCA and the DVM–both traceable to PTB's quantum Hall resistor and LNE's Josephson voltage standard, respectively—and the current generated on the other side by the PQCG—involving LNE's PJVS and QHRS—and estimated from the expression . Considering the degrees of equivalence of PTB and LNE in the international comparison BIPM.EM-K10.b [35] of Josephson array voltage standards at 10 V, this experiment can be interpreted as a comparison of the realizations of the ampere in the two laboratories. However, without the knowledge of the transresistance, this experiment can be used to determine the value RIV
from IPQCG and (see appendix
We have performed eighteen measurements within two days. We have applied exactly the same measurement sequence as for DA♯1. Figure 11 shows the relative deviation ΔI/I = (IULCA − IPQCG)/IPQCG over the two days. These measurements include the tests of the quantization criteria of the PQCG and the consistency check in terms of linearity of the ULCA. A summary of the parameters of the measurements is presented in table 1.
Download figure:
Standard image High-resolution imageTable 1. Parameters of the measurements during the comparison: number of Josephson junctions nJ, number of turns NJK and N, and tests realized.
±nJ | NJK | N | ±IPQCG | Tests |
---|---|---|---|---|
±2048 | 160 | 84 | ±43.750 μA | Nominal value |
±1024 | 160 | 84 | ±21.875 μA | Nominal value/2 |
±2048 | 160 | 42 | ±43.750 μA | nJ/2 and G × 2 |
±2048 | 160 | 84 | ±43.750 μA | Bias trim: ±0.1 mA and ±0.2 mA |
The Type-A relative uncertainty of each measurement point is <2 × 10−8. This low Type-A uncertainty allowed us to resolve a drift of 2.5 × 10−7 over the first 24 h followed by a period of stabilization to within 2.5 × 10−8, corresponding to the last 9 points in figure 11. The quantization criteria have been checked within the standard deviation of ⩽2.5 × 10−8. After stabilization, the linearity of the ULCA has been tested at ±21.875 μA and ±43.750 μA. The results are shown in figure 12. The error bars represent only the Type-A standard uncertainty. A slight deviation of the order of 10−8 can be observed but it is within the stability of the measurements.
Download figure:
Standard image High-resolution imageThe DVM gain error has been estimated from a two-weeks calibration campaign before the comparison. Unfortunately, we observed a deviation from the known linear drift when recalibrating the instrument one month after the comparison, which might have been due to a temperature regulation problem in the laboratory during this period. Therefore the relative Type-B uncertainty contribution due to this correction has been increased to cover this problem and amounts to 2.3 × 10−7.
The typical uncertainty of the calibrations of RIV
at PTB is 6 × 10−9. During the comparison the internal temperature of the ULCA was recorded and was about 21 °C at LNE, whereas the transresistance calibration is given at 23 °C. This temperature difference was accounted for by a correction calculated from the −1.4 × 10−7 °C−1 temperature coefficient determined at PTB. The corresponding relative uncertainty contribution is 0.2 × 10−7. An additional uncertainty component has been added in the uncertainty budget, it accounts for the long term stability and the stability against transportation of RIV
. It has been estimated to 2 × 10−7 (see appendix
Table 2. Relative Type-B standard uncertainty (uB) budget of the comparison.
Contribution | uB |
---|---|
ULCA: RIV calibration | 6 × 10−9 |
ULCA: RIV temperature correction | 2 × 10−8 |
ULCA: RIV stability | 2 × 10−7 |
PQCG: cable correction | 2 × 10−9 |
PQCG: other contributions | <2 × 10−9 |
DVM: calibration | 2.3 × 10−7 |
Total | 3.10 × 10−7 |
For the final result of the comparison, we have used 9 data points obtained for ±21.875 μA and ±43.750 μA. The experimental standard deviation of the mean of the 9 points is 7 × 10−9. The final result of the comparison is expressed as the mean value of the 9 points within the total relative combined uncertainty (represented by the red dashed dotted lines in figure 11) (kc = 1):
In this comparison, the realizations of the ampere from both laboratories agree to −3.7 parts in 107 with a combined standard uncertainty of 3.1 parts in 107 (kc = 1). This result is reported in figure 11. Clearly there is no significant discrepancy between the two current values since the difference is lower than the expanded uncertainty of 6.2 parts in 107 (kc = 2).
5. Conclusion
To conclude, we have identified and estimated the noise sources of the PQCG with the support of an electrical model based on magnetically coupled RLC circuits. It turns out that the primary quantum circuit of the PQCG, which is formed by the PJVS, the QHRS and the CCC in series, is strongly coupled to noise via the connection to ground. This leads to an instability of the SQUID for windings of large number of turns, which is resolved by implementing a cryogenic damping circuit, an optimized SQUID feedback circuit and an adapted grounding scheme.
These improvements allowed us performing calibrations of commercial precision ammeters with a relative measurement uncertainty of a few parts in 107, only limited by their stability. These devices, which also demonstrated linearity within a few parts in 107, could be used as secondary current standards if they are periodically calibrated against the PQCG.
Furthermore, the improved PQCG was used to feed an ULCA with currents in the 50 μA range. The agreement we found, at a level of a few parts in 107, demonstrates the equivalence of the ampere realizations at LNE and PTB. Our results represent the state-of-the-art of currents in the range ±50 μA. However, there is room for improvements for future comparisons with the goal to reach 1 part in 107 or less. Improvements should include: (i) the use of transportable thermostat (thus avoiding thermal shocks during transport potentially responsible for discontinuity in the calibration value of the ULCA), (ii) timely calibration of the DVM before and after the series of measurements (or even during this series). Even better performance could be achieved by using a null detection scheme with a second PJVS [1]. Moreover, the range over which the currents can be compared can be extended down to tens of the nA [36] with the same expected uncertainties. All these new results open the way to a future metrology of the ampere that should be as efficient as those of the ohm and the volt.
Acknowledgments
The authors gratefully acknowledge support by M Mghalfi, J Azib and O N Ndir. This work was supported by the Joint Research Project 'e-SI-Amp' (15SIB08). It received funding from the European Metrology Programme for Innovation and Research (EMPIR) cofinanced by the Participating States and from the European Union's Horizon 2020 research and innovation programme.
Appendix A
A.1. PJVS, QHRS and CCC details
The Josephson array is subdivided into 14 smaller array segments, and follows the sequence 256/512/3072/2048/1024/128/1/1/2/4/8/16/32/64. The quantum Hall resistance standard, based on an eight-terminal Hall bar made of a GaAs/AlGaAs semiconductor heterostructure (LEP514), was produced at the Laboratoire Electronique de Philips [18]. The design of the CCC used here is detailed in [15]. It is composed of 15 windings with the following numbers of turns: 1, 1, 2, 2, 16, 16, 32, 64, 128, 160, 160, 1600, 1600, 2065, and 2065. It is equipped with a Quantum Design Inc. dc SQUID, operating with flux modulation at 500 kHz. The SQUID has a white noise level of 3 and a 1/f corner frequency fc = 0.3 Hz, where is the superconducting flux quantum. The flux to ampere-turn sensitivity of the CCC is γCCC = 8 μA turn/ϕ0. Note that in the present study no additional current divider [1] is used to finely adjust the gain between IPQCG and IPQCS.
A.2. Model used for the simulations
For the model of the noise coming from the CCC at high frequency, we consider two series-RLC circuits (R1, C1, L1, R2, C2, L2) coupled by a mutual inductance , where κ is the coupling factor. The parameters are given in the figure 13. The currents circulating in the two circuits i1 and i2 are written as a current vector:
Download figure:
Standard image High-resolution imageThe noise voltage density and of the voltage noise sources, e1 and e2, due to the Johnson–Nyquist noise of the resistances R1 and R2 are expressed in V and are given by: and respectively and are written as a vector:
The currents are deduced from the following system of linear equations:
where
and and .
The screening current δI on the toroidal superconducting shield due to the noise currents is deduced from the relation: , where and . The noise sources e1 and e2 being uncorrelated, the current noise amplitude is given by: . Finally, the flux noise contribution is .
A.3. Calculation of the expected relative Allan deviation for the PQCG in figure 7
The Allan deviation allows the analysis of the noise by differentiating noise types according to the exponent of its power dependence with time. White noise manifests itself by a τ−1/2 dependence. In the white noise regime, the Allan deviation σw is related to the flux power spectral density Sϕ by:
where τ is the sampling time.
We want to relate the flux power spectral density measured for the PQCG to the relative Allan deviation σI /I (=σϕ /ϕ) expected for a series of current measurements obtained from on-off-on cycles that allows us to subtract the offsets. For that, we calculate the standard deviation of an on-off-on cycle by adapting a calculation done for polarity reversals by Drung et al [11].
The value of the current calibration point, corresponding to one on-off-on cycle, is calculated from the average of the 'on' values for which we subtract the average of the 'off' values. We suppose that the standard deviations, σon and σoff, for the 'on' and 'off' polarities, can be calculated from the white noise standard deviations (see equation (2)) corresponding to the duration, τon and τoff = 2τon, of the 'on' and 'off' polarities, respectively.
The white noise uncertainty for the on-off-on cycle, σcycle, is then given by:
with .
In figure 7, we plot the relative Allan deviation as a function of the sampling time τ = ns τ0, so we want to express σcycle as a function of τ0. Following [11], we define the effective measuring time τe, i.e. the sampling time without the waiting time after polarity reversals, here: τe = τ0 − 2τw = τ0(1 − ρ), with ρ = 2τw /τ0 the rejection rate. For the on-off-on cycle, τon = τe/4. We rewrite σcycle as a function of τ0,
We introduce the factor A' defined as , such that:
Finally, the Allan deviation for a sequence of ns cycles is:
All the parameters used for the calibration measurements and for the Allan deviation are given in table 3. We tried to realize as much as possible similar acquisition times for the two DAs despite the different parameter settings and the different dead times. As can be seen from table 3, the DA♯1 has been used in the mode AUTOZERO ON (internal offset errors due to bias currents of the amplifiers and thermal emfs are subtracted), which is the most accurate mode. However compared to the other DA, which does not have this function, the measuring time has been doubled for DA♯1, in order to have the same effective measuring time in a given polarity. This explains the difference of the timing parameters for the two DAs.
Table 3. Parameters used for the calibration measurements and for the Allan deviation measurements.
Parameters | DA♯1 cal | DA♯2 cal | DA♯1 'Allan' | DA♯2 'Allan' |
---|---|---|---|---|
NPLC | 10 | 64 | 10 | 64 |
AUTOZERO | ON | — | ON | — |
Points per pol | 30 | 5 | 15 | 3 |
τon (s) | 12 | 6.4 | 6 | 3.8 |
τw (s) | 2 | 5 | 2 | 5 |
τ0 (s) | 52 | 35.6 | 28 | 25.4 |
facq. = 1/τ0 (mHz) | 19 | 28 | 36 | 39 |
τe | 48 | 25.6 | 24 | 15.4 |
ρ | 0.08 | 0.28 | 0.14 | 0.39 |
♯ of points | 11 | 11 | 71 | 71 |
Total duration (s) | 572 | 392 | 1988 | 1800 |
To calculate the relative Allan deviation expected for the PQCG at IPQCG = 0.918 mA, we have to divide A' by the flux seen by the SQUID due to IPQCS, i.e. . The flux is 460 ϕ0 (calculated with the parameters NJK = 160, N = 4, IPQCS = 23 μA and the flux sensitivity 8 μA turns/ϕ0). We calculate based on a large range of estimates for at 20 mHz, which spans from 8 μϕ0/ to 50 μϕ0/ taking into account the all cases from white noise to 1/f noise.
Based on the previous calculation, it is possible to use the fitting parameters of the relative Allan deviation σI /I of the current measurements of the 2 DAs presented in figure 7, to estimate the equivalent current noise and eventually to compare with the constructor specifications (table 4).
Table 4. Results of the calculation of for the PQCG when at 20 mHz is equal to 8 μϕ0/ or 50 μϕ0/ and results of the calculation of for DA♯1 and DA♯2 from the values of obtained from the fit of figure 7.
Instrument | |||
---|---|---|---|
PQCG | a | 8 μϕ0/ | — |
PQCG | a | 50 μϕ0/ | — |
DA♯1 | b | — | 0.8 nA a |
DA♯2 | b | — | 0.3 nA a |
aDenotes a result from a calculation. bDenotes a result from a fit.
A.4. Determination of the RIV
Figure 14 shows the relative deviation of RIV from the 100 kΩ nominal value measured at PTB (black dots) as a function of time over a period of more than two years including the period of the comparison. Before transportation, the transresistance followed an exponential decay at PTB (red dashed line in figure 14). However, the first calibration after returning to PTB showed a deviation of 4.8 × 10−7 from the predicted value, which might be due to mechanical and thermal shocks during transportation. Unfortunately, we cannot clearly assign the discontinuity to the transport to or from LNE. This impacts the estimation of the transresistance, and hence the value of IULCA, at the time of the comparison. Here, to make the less assumptions possible, we have considered a linear interpolation at the date of the comparison, between the value of the last calibration before transportation and the first after return to PTB. The estimated uncertainty is calculated from a rectangular probability distribution of total width equal to the difference between these two values (7 × 10−7). The relative uncertainty on the estimation of RIV is therefore 2 × 10−7. From the measurements done at LNE, the transresistance RIV can be estimated from ΔI/I = (RIV − 100 kΩ)/100 kΩ. This value can be compared to the value obtained at PTB. It has been added in figure 14. The error bar reflects the total combined uncertainty without the contribution of the stability of the RIV and is therefore dominated by the contribution due to the calibration of the DVM. On a general point of view, in principle, the transresistance could also be directly calibrated by means of the CCC-based resistance bridge presented in [15] and traced back to the quantum Hall resistance at LNE.
Download figure:
Standard image High-resolution image