Brought to you by:
Paper

Novel TMS coils designed using an inverse boundary element method

, , and

Published 13 December 2016 © 2016 Institute of Physics and Engineering in Medicine
, , Citation Clemente Cobos Sánchez et al 2017 Phys. Med. Biol. 62 73 DOI 10.1088/1361-6560/62/1/73

0031-9155/62/1/73

Abstract

In this work, a new method to design TMS coils is presented. It is based on the inclusion of the concept of stream function of a quasi-static electric current into a boundary element method. The proposed TMS coil design approach is a powerful technique to produce stimulators of arbitrary shape, and remarkably versatile as it permits the prototyping of many different performance requirements and constraints. To illustrate the power of this approach, it has been used for the design of TMS coils wound on rectangular flat, spherical and hemispherical surfaces, subjected to different constraints, such as minimum stored magnetic energy or power dissipation. The performances of such coils have been additionally described; and the torque experienced by each stimulator in the presence of a main magnetic static field have theoretically found in order to study the prospect of using them to perform TMS and fMRI concurrently. The obtained results show that described method is an efficient tool for the design of TMS stimulators, which can be applied to a wide range of coil geometries and performance requirements.

Export citation and abstract BibTeX RIS

1. Introduction

Transcranial magnetic stimulation (TMS) is a non-invasive technique to stimulate the brain, which is applied to studies of cortical effective connectivity, presurgical mapping, psychiatric and medical conditions, such as major depressive disorder, schizophrenia, bipolar depression, post-traumatic, stress disorder and obsessive-compulsive disorder, amongst others (Wassermann 2008).

In TMS, strong current pulses driven through a coil are used to induce an electric field stimulating neurons in the cortex. The efficiency of the stimulation is determined by coil geometry (Salinas et al 2007), orientation (Laakso and Hirata 2012), stimulus intensity, depth of the targeted tissue and some other factors (Salinas et al 2009, De Geeter et al 2015), such as, pulse waveform and duration, although the effects of the latter two are still not well understood.

The TMS stimulator most commonly employed is the so called figure-of-eight or butterfly coil, but since the invention of TMS numerous coil geometries have been proposed to improved the performance and spatial characteristics of the electromagnetic stimulation (Deng et al 2013). A notable example is the work of Koponen et al (2014), who have recently presented method for designing spherical minimum-energy TMS coils, which relies on a description of the problem in terms of quasi-static continuous surface current density.

In general, the problem in TMS coil design is to find optimal positions for the multiple windings of coils (or equivalently the current distribution) so as to produce fields with the desired spatial characteristics and properties (Koponen et al 2014) (high focality, field penetration depth, low inductance, low heat dissipation, etc). This is an inverse electromagnetic problem of similar nature to that one found in magnetic resonance imaging (MRI) gradient coil design, where it is also required to determine the spatial distribution of quasi-static electric currents flowing on a conductive surface, which satisfies given requirements for the produced fields, electromagnetic energy, etc (Turner 1993).

Amongst all strategies developed over the last two decades to design MRI gradient coils, there is an especially successful group of techniques which incorporates the stream function within an inverse boundary element method (IBEM) (Pissanetzky 1992). This approach has been efficiently applied to produce MRI gradient coils with arbitrary geometry (Poole and Bowtell 2007, Cobos Sanchez 2010), allowing the inclusion of new coil features in the design process (Cobos Sanchez 2010a, Poole et al 2010).

In this work, a new method to design TMS coils is presented. It is based on an optimisation problem derived from the inclusion of the concept of stream function of a quasi-static electric current into a boundary element method (BEM). The proposed TMS coil design approach is a powerful technique to produce TMS coils of arbitrary shape, and remarkably versatile as it permits the prototyping of many different requirements and performance constraints.

The structure of this work is as follows. Firstly we review some of the most relevant requirements and parameters that describe the performance of a TMS coil. Secondly an outline of the stream function IBEM is presented in section 3, which allows us to formulate the optimisation problems in section 4. Finally we illustrate the validity of the proposed approach with the design and simulation of a set of TMS coils of different geometry subjected to different performance requirements, where the mechanical behaviour of each stimulator has been also evaluated when operating in an external static.

2. Method

2.1. TMS coil requirements and performance

In the following, some of the most important properties that assess the efficiency and performance of a TMS coil are listed.

2.1.1. Stored magnetic energy

Minimum stored magnetic energy (or equivalently minimum inductance) is a crucial requirement in TMS coil design, as it enables the most rapid switching possible of the TMS fields (Koponen et al 2014).

2.1.2. Power dissipation

An ideal TMS stimulator should have a low power dissipation (or equivalently low resistance) in order to reduce the unwanted Joule heating. This is an essential issue in procedures such as repetitive TMS, where effectiveness and safety are remarkably reduced by coil overheating.

2.1.3. Mechanical forces

The high currents flowing in the TMS coil induce significant mechanical forces between coil elements, which are responsible for the audible click when a magnetic stimulator fires. So in designing TMS coils, it is important to consider the force and torque experienced to avoid damage or changes in the coil structure due to excessive mechanical stress.

Moreover, this issue is specially crucial in the concurrent application of TMS and functional MRI, where the operation of the coil within a large external field produces intense forces that can lead to mechanical failure of the TMS system.

In this particular case, the torque, m experienced by a surface current, J in the presence of the axial main magnetic field, ${{\mathbf{B}}_{0}}$ , is given by

Equation (1)

where the integration domain, S, is the coil surface, and $\mathbf{r}$ the vector pointing from the center of the coil to the corresponding current element.

2.1.4. Induced electric field

An ideal TMS coil should be able to stimulate a prescribed brain region and induce a minimum electric field in the rest of non target regions, to avoid safety risks to both the subject and the technician handling it.

More precisely, the spatial characteristics of the TMS electromagnetic stimulation can be described with the following metrics.

2.1.5. Penetration

Or depth, d1/2, is the radial distance from the cortical surface to the deepest point where the electric field strength is half of its maximum value on the surface.

2.1.6. Focality

There are several definitions of focality in the literature; in general, more focal stimulation means a smaller stimulation area with the maximum field. Here, we have employed the focality defined through the effective surface area (Deng et al 2013)

Equation (2)

where V1/2 is the volume inside the brain in which the stimulus is over $50 \% $ of the maximum. This metric takes into account that it is harder to have a focal stimulus deeper in the head.

2.2. Discrete quasi-static approximation

In this work, it is assumed that the range of frequencies of the employed TMS pulses is below 50 MHz, at these frequencies the electromagnetic properties of the body allow us to use a quasi-static approximation (Miranda et al 2003). Additionally, it is also considered that the currents flowing in the TMS coil are likewise in the quasi-static regime.

Under these conditions, and by using a boundary element method (BEM), the current under search can be modelled in terms of the nodal values of the stream function and elements of the local geometry (Cobos Sanchez et al 2010b).

Let us first assume that the conducting surface, S, on which we want to find the optimal current, is divided into Q triangular elements with N nodes, which are lying at each vertex of the element. The current density at a given point at the surface can be then written as

Equation (3)

where $\psi ={{\left[{{\psi}_{1}},{{\psi}_{2}},\ldots,{{\psi}_{N}}\right]}^{T}}$ is the vector containing the nodal values of the stream function, and $\boldsymbol{\jmath}^{{n}}$ , are functions related to the curl of the shape functions (Cobos Sanchez et al 2010b). The stream function nodal values $\psi \in {{\mathbb{R}}^{N}}$ are the optimization variables in this work.

By using this current density model, equation (3), the discretized expressions for the required physical magnitudes involved in the inverse problem can be obtained, and so matrix equations that transform ψ to the various coil properties (Poole et al 2010) can be then constructed

  • The magnetic field at a series of H points, $\left\{{{\mathbf{r}}_{1}},{{\mathbf{r}}_{2}},\ldots,{{\mathbf{r}}_{H}}\right\}$
    Equation (4)
    The coefficient ${{B}_{{{x}_{i}}}}(h,n)$ , is the xi  −  component of the magnetic induction produced by the current element associated to the nth-node in the prescribed hth-point.
  • The magnetic stored energy in the coil
    Equation (5)
    where L is the inductance matrix, a full symmetric positive definite matrix (Cobos Sanchez et al 2012).
  • The resistive power dissipation of the coil
    Equation (6)
    where R is the resistance matrix, a full symmetric positive definite matrix (Cobos Sanchez et al 2012).
  • The electric field induced in a series of M points inside of the conducting system (Cobos Sanchez et al 2010b), $\left\{{{\mathbf{r}}_{1}},{{\mathbf{r}}_{2}},\ldots,{{\mathbf{r}}_{M}}\right\}$
    Equation (7)
  • The net torque experienced by the surface current in an external static magnetic field, ${{\mathbf{B}}_{0}}$
    where the nth-column of the matrix $\mathbf{M}$ represents the torque vector produced by ${{\mathbf{B}}_{0}}$ in the current associate to the nth-node.

3. Problem formulation

By using the discretized current model presented in section 2, the TMS coil design problem can be posed as a constrained optimization, in which the goal is to find the optimal set of stream function nodal values, so as to achieve desired magnetic and electric fields in regions of interest, to minimize the magnetic energy stored and the power dissipated. According to these requirements we can define the following quadratic programming problem, in which a functional of several variables must be minimized

Equation (8)

It contains terms to control the magnetic field in a prescribed region of interest, $f\left(\psi \right)$ , the electric field induced in a second volume of interest, $e\left(\psi \right)$ , stored magnetic energy, $W\left(\psi \right)$ and power dissipation $P\left(\psi \right)$ .

The weights α, β, γ and δ are user-definable parameters that allow control of the different coil properties, and illustrates the trade-off between coil features (Cobos Sanchez et al 2012).

The f term represents the sum-of-squares of the error in magnetic field at a series of H points in a region of interest (ROI)

Equation (9)

where $\parallel \cdot {{\parallel}_{2}}$ is the l2  −  norm, and $b_{{{x}_{i}}}^{t}$ , xi  =  x, y, z, is an H-dimension vector that lists the desired values of the xi component of the magnetic field at each point. Analogously, the term e in equation (8) is sum-of-squares of the error in electric field at a series of M points

Equation (10)

where $e_{{{x}_{i}}}^{t}$ are M-dimension vector that lists the desired values of the xi, component of the electric field at each point of in a second region of interest (ROI2).

It is clear that induced electromagnetic fields increase with the norm of the solution, whereas incorporating minimum W and P terms has the effect of reducing $\parallel \psi \parallel $ . Therefore maximisation of the magnetic or electric field can be achieved by minimization of W and/or P given a sufficient accuracy of target field.

In this work, the target fields $b_{{{x}_{i}}}^{t}$ and $e_{{{x}_{i}}}^{t}$ , xi  =  x, y, z, are chosen to be null over a given region of interest when it is desired to minimize the stimulus associated to that field component; whereas, if we want to maximise the stimulation in a prescribed volume due to a particular field component, the corresponding target field is set to be constant.

It is also worth stressing that the final goal in TMS coil design is to find the wire arrangement that approximates the continuous current distribution. The conversion of the current solution into a conductor pattern can be achieved by contouring the stream function (Koponen et al 2014).

3.1. Forward problem

Having found the wire pattern, the value of the coil inductance is evaluated by using FastHenry © (Kamon et al 1994), a multipole impedance extraction tool, and assuming the coil wires have a 1 mm diameter.

Full details of the induced E-field of each TMS coil solution are acquired using an existing direct BEM (Cobos Sanchez et al 2009), which allows calculation of the electric field induced by the coil windings in conducting systems.

Moreover, sinusoidal variation of the electric and magnetic fields (f  =  5 kHz) has been adopted, and unless it is stated, an arbitrary coil current with peak value of 1 kA has been considered.

Additionally, depth and focality metrics (section 2.1.6) were employed to describe the spatial characteristics of the stimulation; where the human head has been modelled by a homogeneous sphere of 8.5 cm radius and isotropic conductivity. The cortical surface was assumed to be at a depth of 1.5 cm from the surface of the head (Deng et al 2013), so the cortex is described by a sphere of 7.0 cm radius.

Finally, in order to evaluate the mechanical behaviour when performing TMS and functional MRI concurrently, the net torque of each stimulator is calculated in three different scenarios, in which the main magnetic field, ${{\mathbf{B}}_{0}}$ , is considered to be parallel to each Cartesian axis and homogeneous; that is, ${{\mathbf{B}}_{0}}={{B}_{0}}{{\mathbf{\hat{x}}}_{i}},~{{\mathbf{\hat{x}}}_{i}}=~\mathbf{\hat{x}},\mathbf{\hat{y}},{{\mathbf{\hat{z}}}_{i}}$ .

In the following, the proposed approach is demonstrated with the design of seven TMS coil examples, which have been chosen to elucidate the method and to illustrate some of the many different requirements that can be prototyped with this technique to design TMS coils of almost any geometry.

3.2. Coil 1: flat coil with minimum inductance and optimized By

The geometry of the first coil example is that of a rectangular flat form of $14~\text{cm}\times 7.5$ cm located at the xy  −  plane, and it is designed to have minimum stored-energy and maximise By in a prescribed region of interest (ROI), which is made up of 400 points inside a spherical region of radius 2 cm that is centred on the z-axis, and 4 cm below the conducting surface as shown in figure 1.

Figure 1.

Figure 1. (a) Schematic diagram showing the conducting surface for Coil 1, Coil 2 and Coil 3, along with the region of interest (ROI) in which the desired stimulation must be achieved. (b) Frontal view of the geometry used for Coil 1, Coil 2 and Coil 3, the red sphere is the ROI.

Standard image High-resolution image

The relevant functional describing this problem is a particular case of equation (8), which can be written as

Equation (11)

where $b_{y}^{t}$ has been set to have a constant value, in order to achieve an optimised stimulation in the ROI due to the y-component of the magnetic field.

3.3. Coil 2: flat coil with minimum power dissipation and optimized $|B|$

By using the same rectangular planar surface as in section 3.2, a slightly more complex coil design problem can be posed by maximizing the magnitude of the B-field in the same the prescribed spherical volume of interest (ROI), figure 1, while having minimum power dissipation (or equivalently minimum resistance). The functional describing this particular problem can be obtained by imposing $\beta =\gamma =0$ in equation (8); where $b_{x}^{t},b_{y}^{t}$ and $b_{z}^{t}$ are chosen to have a constant value to optimise the modulus of the magnetic field induced in the ROI.

3.4. Coil 3: flat coil with minimum power dissipation and optimized $|E|$

By using the same rectangular planar surface as in sections 3.2 and 3.3 we can explore the design of TMS coil with minimum power dissipation and an optimized magnitude of the E-field in the same the prescribed spherical volume of interest (ROI), figure 1(b). The functional describing this particular problem can be obtained by imposing $\alpha =\gamma =0$ in equation (8). The target fields $e_{x}^{t},e_{y}^{t}$ and $e_{z}^{t}$ are chosen to have a constant value to optimise the modulus of electric field induced in the ROI.

3.5. Coil 4: spherical coil with minimum inductance and optimized $|E|$

Koponen et al (2014) have shown the design of minimum-energy TMS coils wound on a spherical surface, this represents an interesting approach as this particular geometry is well matched to the head.

Here we investigate the design of a minimum inductance (or equivalently minimum stored energy) spherical TMS coil of radius 9 cm, which is designed to produce an optimised E-field in a spherical ROI of 2 cm radius and centred 5 cm above the centre of the conducting sphere, figure 2(a).

Figure 2.

Figure 2. (a) Schematic diagram showing the spherical conducting surface (for Coil 4 and Coil 5), along with the cortex, the region of interest (ROI) in which the desired stimulation must be achieved and the second volume of interest (ROI2) where the E-field must be minimized (only for Coil 5); (b) the geometry of the spherical Coil 6, where the red sphere is the ROI; (c) the geometry of the hemispherical Coil 7, where the red sphere is the ROI.

Standard image High-resolution image

The functional describing this particular problem can be obtained by imposing $\alpha =\delta =0$ in equation (8); where $e_{x}^{t},e_{y}^{t}$ and $e_{z}^{t}$ are chosen to have a constant value to optimise the modulus of electric field induced in the ROI.

3.6. Coil 5: spherical coil with minimum inductance, optimized $|E|$ and reduced unwanted stimulation

The undesired stimulation in non-target cortex regions can be reduced by introducing a second region of interest (ROI2), in which the E-field produced by the stimulator is minimized. This strategy is evaluated by designing a minimum stored-energy spherical TMS coil of radius 9 cm, constructed to produce an E-field which is both a maximum in a spherical ROI and minimum in a second ROI2. Both volumes of interest are of 2 cm radius and formed by 400 points, where ROI2 is concentric to the conducting sphere and ROI is shifted by 5 cm in the positive z-direction, figure 2(a). The functional of Coil 5 design problem is

Equation (12)

where the terms e and ${{e}^{\prime}}$ are sum-of-squares of the error in electric field at ROI and ROI2 respectively, whereas β and ${{\beta}^{\prime}}$ the corresponding weighting parameters.

The target fields are chosen to have a constant value in the ROI and null in the ROI2, so as to produce a strong stimulation in the desired region, while reducing the undesired stimulation in the non target volume.

3.7. Coil 6: spherical coil with minimum inductance, optimized $|B|$ and improved focality

In this experiment we have studied the reduction of the size of the ROI, in order to improve the focality of the TMS coil. To this end, a minimum inductance spherical TMS coil of radius 9 cm, capable of stimulating with an optimised B-field a spherical ROI of 0.5 cm radius and 7 cm above the centre of the conducting sphere, figure 2(b), has been evaluated. This particular problem can be formulated by imposing $\beta =\delta =0$ in equation (8). The target fields $b_{x}^{t},b_{y}^{t}$ and $b_{z}^{t}$ are chosen to have a constant value to optimise the modulus of the magnetic field induced in the ROI.

3.8. Coil 7: hemispherical coil with minimum inductance and optimized $|E|$

Closed geometries, such as spherical shapes, suffer the significant disadvantage of offering no natural aperture for patient access. This known problem has already been addressed in Koponen et al (2014) work, where it is suggested to constraint the current distribution to have a negligible value at the region opposite to the target.

Here we propose the use of open geometries as an efficient solution to overcome this type of problem. More precisely, we have studied the design of a TMS coil with hemispherical shape with a radius of 9 cm, figure 2(c), which is also well matched to the head and allows a favourable patient access. It is also designed to produced a minimum inductance and an optimized E-field in a prescribed spherical volume of interest (ROI) of radius 2 cm and shifted by 5 cm in the positive z-direction, figure 2(c).

The functional describing this particular problem can be obtained by imposing $\alpha =\gamma =0$ in equation (8). The target fields $e_{x}^{t},e_{y}^{t}$ and $e_{z}^{t}$ are chosen to have a constant value to optimise the modulus of electric field induced in the ROI.

3.9. Choice of regularization parameter

In the construction of the functional equation (8) we must find the optimal choice of the regularisation parameters α, β, γ and δ. For the designs tackled here, they have been chosen so that the self-inductance and the resistance of the TMS coil are less than a prescribed value ($L\leqslant 10\mu $ H and $R\leqslant 100$ m$ \Omega $ for flat rectangular TMS coils, and $L\leqslant 18\mu $ H and $R\leqslant 200$ m$ \Omega $ for spherical and hemispherical coils).

It is worth noting, that this election of maximum values of R and L are just illustrative; for each particular problem, the election of the regularisation parameters can be studied so as to yield a solution with a concrete desired coil feature (Cobos Sanchez et al 2012).

4. Results

4.1. Planar TMS coils

A standard figure-of-eight coil has been chosen for sake of comparison, where each circular lobe has a diameter of 70 mm, containing 8 loops of 1 mm-thick wire, as can be seen in figure 3(a). We refer to this stimulator as Coil 0, and its performance parameters are found in table 1.

Figure 3.

Figure 3. Wire arrangements with 16 turns of (a) Coil 0, (b) Coil 1, (c) Coil 2 and (d) Coil 3. Red wires indicate reversed current flow with respect to blue.

Standard image High-resolution image

Table 1. TMS coil performance parameters: L is the inductance, R the resistance, S1/2 the effective surface area (focality), D1/2 the penetration or depth and Ncontours is the number of levels in which the stream-function is contoured to produce the wire paths. Simulated values of L and R were obtained using FastHenry © (Kamon et al 1994) using 1 mm diameter circular cross-section wire.

  L (μH) R ($\text{m} \Omega $ ) D1/2 (cm) S1/2 (cm2) Ncontours
Coil 0 13.3 127 1.2 16.1 16
Coil 1  9.3  70 1.2 13.9 16
Coil 2  6.9  56 1.3 15.9 16
Coil 3  7.1  57 1.3 16.2 16
Coil 4 17.4 174 2.5 48.0 18
Coil 5 10.1 155 1.6 31.7 18
Coil 6  5.0  73 1.4 14.3 18
Coil 7 14.7 125 2.3 41.4 18

The wire-paths of the solution to Coil 1, Coil 2 and Coil 3 problems are shown in figures 3(b)(d) respectively. All of them are two lobed TMS coils (red wires indicate reversed current flow with respect to blue), where it can be seen that the winding density is more concentrated over the region of stimulation for all cases.

Moreover, in order to produce a high stimulus in a given region, use of maximum $|B|$ and $|E|$ conditions can be used, which in general, lead to different solutions. Nonetheless, in the case of the flat designs tackled here, it has been found a similitude between the wire arrangements of Coil 2 (figures 3(c)) and Coil 3 (figure 3(d)). This fact is also confirmed when evaluating the characteristics of each TMS coil in table 1, where it can be seen that Coil 1, Coil 2 and Coil 3 provide a more efficient performance than Coil 0, which has remarkably high values of L and R compared to those obtained for the TMS coils design with the stream function IBEM.

The wire arrangements of the four rectangular TMS coils can also be found in figure 4 along with the colour map of the modulus of the E-field induced in the spherical cortex model (Coil 0 in figure 4(a), Coil 1 in figure 4(b), Coil 2 in figure 4(c) and Coil 3 in figure 4(d)).

Figure 4.

Figure 4. E-field induced at the surface of the spherical cortex for (a) Coil 0, (b) Coil 1, (c) Coil 2 and (d) Coil 3. Red wires indicate reversed current flow with respect to blue.

Standard image High-resolution image

As it is expected from their wire paths, there is a similarity between the electric field distribution induced by Coil 2 and Coil 3. Nonetheless it is worth noting that the maximum E-field induced by Coil 3 (about 53.0 V m−1) is significantly higher than that one found for Coil 2 (49.5 V m−1). The highest electric field induced at the cortex surface is produced by Coil 1, which is about 60.2 V m−1.

Similar facts can also be appreciated when evaluating the E-field induced inside the cortex by Coil 0, Coil 1, Coil 2 and Coil 3 when energized, as can be seen in figures 5(a)(d) respectively, where it is depicted the colour-coded $|E|$ maps on a xy-plane with z  =  4 cm and the grey line delineates the region that we want to stimulate (ROI). It is clear that these planar TMS coils generate an electric field that satisfies the initial requirements of stimulating the prescribed ROI.

Figure 5.

Figure 5. (a) Colour coded map for the E-field field induced in a xy-plane (z  =  4 cm) for (a) Coil 0, (b) Coil 1, (c) Coil 2 and (d) Coil 3. The grey line delineates the ROI.

Standard image High-resolution image

4.2. Spherical TMS coils

The wire-paths of the solution to Coil 4, Coil 5 and Coil 6 problems are shown in figures 6(a)(c) respectively, where red wires indicate reversed current flow with respect to blue. As expected, there is a higher density of winding turns over the region of stimulation. The performance parameters of each spherical TMS coil can be found in table 1.

Figure 6.

Figure 6. Wirepaths with 18 turns of (a) Coil 4, (b) Coil 5, (c) Coil 6 and (d) Coil 7. Red wires indicate reversed current flow with respect to blue.

Standard image High-resolution image

These wire arrangements can also be found in figure 7 along with the colour map of the modulus of the E-field distribution produced in the spherical cortex model by Coil 4 (figure 7(a)), Coil 5 (figure 7(b)) and Coil 6 (figure 7(c)). For all cases, the E-field induced pattern is a kind of a mirror image of the coil current distribution, forming two loops of eddy currents with reflection symmetry respect the x  =  y plane.

Figure 7.

Figure 7. E-field induced at the surface of the spherical cortex for (a) Coil 4, (b) Coil 5, (c) Coil 6 and (d) Coil 7. Red wires indicate reversed current flow with respect to blue.

Standard image High-resolution image

It can be seen that all the spherical TMS coils presented here satisfy the initial requirements of stimulating the prescribed ROI, where the highest values of the electric field are found.

Additionally, Coil 5 has been designed to mitigated the undesired stimulation in the prescribed non target cortex region (ROI2), in order to evaluate this fact, the electric field induced in the ROI2 by Coil 5 is depicted in figure 8(a), which can be compared to that one generated by its counterpart Coil 4 in figure 8(b). It can be seen how the values obtained for Coil 5 are lower than those produced in the case of Coil 4, being the reduction of the electric field induced greater than $70 \% $ at many points in the ROI2. The price we have to pay for the mitigation of the undesired stimulation is a significant reduction of the electric field induced in the target region ROI; and even the increase of the E-field induced in other head parts.

Figure 8.

Figure 8. Three-dimensional plot of the colour-coded modulus of the E-field at the surface of the ROI2 induced by (a) Coil 4 and (b) Coil 5.

Standard image High-resolution image

Coil 6 solution induced a significant lower E-field in the cortex compared to those produced by the other spherical coils, nonetheless it fulfil the initial requirements of producing a more focal and superficial stimulation in the cortex, as the values of d1/2 and S1/2 show in table 1.

4.3. Hemispherical TMS coil

Figure 6(d) shows the wire path of the Coil 7, which is again concentrated over the region of stimulation. It is formed by two lobes of nine turns, which have been produce by contouring the optimal stream function with the same number of levels. The relevant parameters for Coil 7 are detailed in table 1, which indicate that the hemispherical Coil 7 have slightly better performance that than its spherical counterpart Coil 4.

The electric field induced in the cortex by this hemispherical TMS stimulator and the corresponding one produced by Coil 4 present a somewhat similar pattern, although the latter has a higher magnitude over a wider cortex surface, as it can be seen in figures 7(a) and (d).

5. Discussion

The TMS coil solutions obtained for rectangular flat geometries are double-coned coil with butterfly-wing shaped windings, of which symmetry depend on the field constraints. In general, minimization of only one field component yields highly symmetrical coils, where the symmetry axes of each solution depends on the particular component to be optimised; whereas minimisation of $|E|$ or $|B|$ changes the symmetries of the solution, as some components have more weights than other in the optimisation process.

Moreover, it has been also found that the increase of γ or δ leads to a spreading of wires, phenomenon specially remarkable when raising the minimum power dissipation condition, fact which can be employed to increase the efficiency of the TMS stimulator by allowing extra turns to be added or to reduce localized heating in the coil, as regions with maximum temperature coincide with those where the coil windings are closely spaced.

Furthermore, the induced E-field in the ROI by Coil 1 is higher that those produced by Coil 2 and Coil 3 (figure 4) which have been specifically designed to maximise the modulus of electric field in this prescribed volume, the main reason for this increased stimulation of Coil 1 is a significant degradation in the performance (higher inductance, resistance and less focal) compared to of Coil 2 and of Coil 3, as can be seen in table 1.

For all coil geometries tackled here, it has been found that there is a range of achievable values of focality and field penetration depth, which can be obtained by varying the size and position of the ROI, existing in any case, a clear trade-off between depth and focality. An example of this result can be found in figure 9(a), where it is depicted the variation of the focality versus the position of the ROI in the z axis for spherical coil of radius 9 cm with minimum inductance and optimized $|E|$ (Coil 4). Red line with red circles, blue line with squares and black line with diamonds show the focality for a spherical ROI of 2 cm, 1.8 cm and 2.2 cm radius, respectively (figure 9(a)). For the same spherical coils the dependence of the inductance with respect the focality is also be evaluated in figure 9(b), where it worth noting a remarkable increase of the inductance when designing more focal coils.

Figure 9.

Figure 9. (a) Focality versus ROI position in the z axis for spherical coil of radius 9 cm with minimum inductance and optimized $|E|$ . (b) Inductance versus focality for spherical coil of radius 9 cm with minimum inductance and optimized $|E|$ . Red line with circles, blue line with squares and black line with diamonds show the focality for a spherical ROI of 2 cm, 1.8 cm and 2.2 cm radius, respectively.

Standard image High-resolution image

TMS coils wound on a spherical surface offer excellent performance, allowing a wide range of achievable S1/2 (and d1/2), so they can be considered for both superficial and deep brain stimulation. However the lack of easy access to the coil's internal volume clearly makes such spherical coils of limited use for medical applications. Here we have suggested the adoption of hemispherical shapes for the coil former, as it offers open access to the coil's interior, although many other open geometries can be evaluated and tackled with the suggested approach which completely geometry-independent.

In order to study the stimulators mechanical behaviour when operating in external static magnetic fields, such as those found in MRI, we have evaluated the torque experienced by the designed TMS coils. The magnitudes of the torque vectors are listed in table 2 for each coil under three different orientations of the homogeneous main field. Those torque modules can provide useful information about any potentially dangerous turning forces may be exerted when the TMS coil is energised in a high-field environment.

Table 2. Net Torque values of each TMS coil calculated with three different orientantions of the main magnetic field, ${{\mathbf{B}}_{0}}$ , which is considered to be homogeneous.

  Torque (NmA−1T−1) when ${{\mathbf{B}}_{0}}={{B}_{0}}\mathbf{\hat{x}}$ Torque (NmA−1T−1) when ${{\mathbf{B}}_{0}}={{B}_{0}}\mathbf{\hat{y}}$ Torque (NmA−1T−1) when ${{\mathbf{B}}_{0}}={{B}_{0}}\mathbf{\hat{z}}$
Coil 0 0.04 0.04 $3\times {{10}^{-7}}$
Coil 1 0.12 0.09 $5.6\times {{10}^{-6}}$
Coil 2 0.26 0.17 $2.2\times {{10}^{-4}}$
Coil 3 0.33 0.21 $2.0\times {{10}^{-4}}$
Coil 4 $9.6\times {{10}^{-2}}$ $7.5\times {{10}^{-2}}$ $7.2\times {{10}^{-3}}$
Coil 5 0.26 0.32 $1.1\times {{10}^{-2}}$
Coil 6 0.13 0.11 $2.4\times {{10}^{-2}}$
Coil 7 0.62 0.77 $9.1\times {{10}^{-2}}$

The net torque experienced by the planar stimulator Coil 1 is comparable to that one found in Coil 0. Nonetheless Coil 2 and Coil 3 torques values are slightly higher than that those experienced by the standard figure-of-eight for any of the three orientations available of ${{\mathbf{B}}_{0}}$ . This increase of the Lorentz forces can be related to the loss of symmetry of the wire pattern.

Moreover, Coil 4, Coil 5, Coil 6 and Coil 7 present also higher torques values than that one of Coil 0, specially the hemispherical TMS coil, which may make its operation at high currents and fields problematic.

There are alternative strategies that can be explored if it is required to balance the torque, for instance, in the case of hemispherical TMS coils, the addition of a short cylindrical extension to the hemispherical coil introduces currents loops in which current circulates in the opposite sense to the loops of the hemispherical part that naturally cancel the torque experienced. Furthermore, the stream function IBEM can also be used to include a torque-balancing constraint in the design process (Poole and Bowtell 2007).

6. Conclusion

In this work, we have introduced a novel method to design TMS coils based on the inclusion of the concept of stream function of a quasi-static electric current into a boundary element method. It is a completely geometry-independent technique; and thus can be applied to a wide range of TMS coil shapes, being particularly valuable for designs where the coil surface has low symmetry. The proposed approach is a remarkably efficient tool for TMS coil design as it permits the prototyping of many different requirements and constraints.

To illustrate the versatility and power of this approach, it has been used for the design of rectangular flat, spherical and hemispherical TMS coils subjected to different performance requirements, such as minimum magnetic stored energy or power dissipation; where it also has been found that there is a range of feasible values for focality and field penetration depth for each coil problem.

This wide interval of achievable values of focality for TMS coils design with stream function IBEM also suggests that this method can be a valuable strategy for the design of deep TMS stimulator.

The proposed technique allows the production of TMS coils capable of minimizing the stimulation in non target cortex regions, this is always at the expense of the decrease of the coil performance, and the increase of the E-field induced in other zones inside the coil. This redistribution can be used to allow larger pulse intensity while producing minimum stimulation in desired non target regions, at expense of increasing the field in less sensitive head regions.

The turning forces exerted when the designed TMS coils are energised in a high-field environment have also been theoretically studied. For hemispherical geometries, it has been found the need to consider the coil orientation with respect to the static field to avoid damage or changes in the coil structure due to excessive torques.

This method open up the possibilities for the design of TMS coil with new properties, which can be included in the optimisation process, such as zero net torque condition, or for instance, geometry constraints which can also be incorporated in the design process (Cobos Sanchez 2010) to improved patient comfort. Likewise, the stream function IBEM approach could allow the distinct head tissue layers properties to be taken into account in the design problem, which could be used to produce more realistic and adequate TMS stimulator. Supplementary data for this article can be found here stacks.iop.org/PMB/62/73/mmedia.

Acknowledgments

This work has been partially financed by the Spanish and Andalusian research programs TEC2013-48414-C3-01 and P12-TIC-1442.

Please wait… references are loading.
10.1088/1361-6560/62/1/73