Brought to you by:

Application of a magnetic fluid seal to rotary blood pumps

, , , and

Published 1 May 2008 IOP Publishing Ltd
, , Citation Y Mitamura et al 2008 J. Phys.: Condens. Matter 20 204145 DOI 10.1088/0953-8984/20/20/204145

0953-8984/20/20/204145

Abstract

A magnetic fluid seal enables mechanical contact-free rotation of a shaft without frictional heat and material wear and hence has excellent durability. However, the durability of a magnetic fluid seal decreases in liquid. The life of a seal applied to a rotary blood pump is not known. We have developed a magnetic fluid seal that has a shield mechanism minimizing the influence of the rotary pump on the magnetic fluid. The developed magnetic fluid seal worked for over 286 days in a continuous flow condition, for 24 days (on-going) in a pulsatile flow condition and for 24 h (electively terminated) in blood flow. The magnetic fluid seal is promising as a shaft seal for rotary blood pumps.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/0953-8984/20/20/204145