Abstract
The SQUID microscope is the most suitable instrument for imaging magnetic fields above sample surfaces if one is mainly interested in field sensitivity. In this paper, both the magnetic moment sensitivity and spatial resolution of the SQUID microscope are analysed with a simple point moment model. The result shows that the ratio of SQUID sensor size to sensor–sample distance effectively influences the sensitivity and spatial resolution. In comparison with some experimental results of magnetic images for room temperature samples from our high-Tc SQUID microscope in an unshielded environment, a brief discussion for further improvement is presented.