Characterisation of a phantom for multiwavelength quantitative photoacoustic imaging

, , and

Published 10 June 2016 © 2016 Institute of Physics and Engineering in Medicine
, , Citation M Fonseca et al 2016 Phys. Med. Biol. 61 4950 DOI 10.1088/0031-9155/61/13/4950

0031-9155/61/13/4950

Abstract

Quantitative photoacoustic imaging (qPAI) has the potential to provide high- resolution in vivo images of chromophore concentration, which may be indicative of tissue function and pathology. Many strategies have been proposed recently for extracting quantitative information, but many have not been experimentally verified. Experimental phantom-based validation studies can be used to test the robustness and accuracy of such algorithms in order to ensure reliable in vivo application is possible. The phantoms used in such studies must have well-characterised optical and acoustic properties similar to tissue, and be versatile and stable. Polyvinyl chloride plastisol (PVCP) has been suggested as a phantom for quality control and system evaluation. By characterising its multiwavelength optical properties, broadband acoustic properties and thermoelastic behaviour, this paper examines its potential as a phantom for qPAI studies too. PVCP's acoustic properties were assessed for various formulations, as well as its intrinsic optical absorption, and scattering with added TiO2, over a range of wavelengths from 400-2000 nm. To change the absorption coefficient, pigment-based chromophores that are stable during the phantom fabrication process, were used. These yielded unique spectra analogous to tissue chromophores and linear with concentration. At the high peak powers typically used in photoacoustic imaging, nonlinear optical absorption was observed. The Grüneisen parameter was measured to be $\Gamma$   =  1.01  ±  0.05, larger than typically found in tissue, though useful for increased PA signal. Single and multiwavelength 3D PA imaging of various fabricated PVCP phantoms were demonstrated.

Export citation and abstract BibTeX RIS

Original content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

1. Introduction

Photoacoustic imaging (PAI) is a biomedical imaging modality currently of great interest as it can achieve the fine spatial resolution of ultrasound whilst benefiting from the spectroscopic specificity of optical techniques. It has the potential to recover 3D, high resolution maps providing information on physiology, pathology and function, but one hurdle remains: it is difficult to obtain reliably accurate quantitative estimates of the tissue chromophore concentrations. Quantification in PAI is non-trivial because photoacoustic (PA) images, while dependent on the presence of optical absorption for contrast, are proportional to the product of the absorption coefficient and the light fluence, neither of which is known a priori. As the fluence is a function of the absorption coefficient, the inverse problem of extracting the concentrations is non-linear and, as the optical scattering is also typically unknown, may be ill-posed (Beard 2011, Cox et al 2012). Many approaches have been proposed to address this challenge (Cox et al 2009, Pulkkinen et al 2014, Shao et al 2014, Gao et al 2015)—and extensively reviewed in Cox et al (2012)—but few have been thoroughly validated and optimised experimentally, much less implemented reliably in vivo. A quantitative photoacoustic imaging (qPAI) framework suitable for preclinical/clinical applications needs not only to give accurate and precise estimates, but also be valid for a wide range of scenarios and display robustness to experimental uncertainties and noise. Importantly, validity over the full range of conditions in which the method will be applied should be thoroughly tested and validated experimentally. One impediment to experimental validations is the lack of a well-characterised and detailed phantom material that has both acoustic and optical properties similar to tissue, which is proving to be ever more important as theoretical qPAI strategies improve towards quantification at higher-dimensionality and resolution.

This paper explores what properties the ideal phantom for quantitative PAI would have (section 2) and considers whether PVCP—currently used for non-quantitative PAI validations—could be a suitable material. Section 3 details the PVCP fabrication process. The characterisation of PVCP's acoustic and optical properties are described in sections 4 and 5 respectively, including a study on the effect of high peak power laser pulses typically used in PAI, section 5.3. The measurement of PVCP's Grüneisen parameter is described in section 6. Section 7 shows photoacoustic images obtained using PVCP as a phantom, and a discussion section (section 8) concludes the paper.

2. Phantoms for quantitative photoacoustic imaging

An ideal phantom for photoacoustic imaging (PAI) should possess: (i) tissue-realistic and controllable acoustic properties; (ii) tissue-realistic, stable and controllable optical parameters; (iii) tissue-realistic values for relevant thermoelastic properties; (iv) realistic and versatile architecture; (v) photo- and mechanical stability during the imaging acquisition procedure; (vi) long-term stability under storage and repeated use; (vii) reproducibility of fabrication.

Up to now, there has been no extensive review of PA phantoms, but some studies have focussed on characterising specific materials for quality control or system assessment in PA, though not directly on their use for quantification (Kharine et al 2003, Spirou et al 2005, Cook et al 2011, Bohndiek et al 2013, Vogt et al 2016). The studies involve a range of tests such as acoustic characterisation of sound speed, attenuation or backscatter, tuning of optical scattering, tuning of optical absorption and/or the construction of phantom geometries to assess image quality parameters such as signal-to-noise ratio (SNR), contrast, uniformity, spatial resolution, among others.

The characterisation of a phantom for qPAI extends beyond what is needed for semi-quantitative and image quality studies. To assess the suitability of a material for high resolution qPAI these further attributes are required:

  • (i)  
    Acoustics
    • (a)  
      Well-characterised broadband frequency dependency of sound speed and acoustic attenuation. PA-generated signals extend to several tens of MHz (Beard 2011) therefore acoustic properties need to be known at high frequencies to be properly accounted/corrected for. It is not crucial that they match soft tissue exactly as long as they can be correctly accounted for, since incorrect estimation compromises the success of high resolution quantification: erroneous sound speed leads to blurring or artefacts in the acoustic reconstruction, whilst incorrect estimation of acoustic attenuation affects the retrieval of the correct amplitude of the initial pressure distribution. Reference values for soft tissue can be found elsewhere (Duck 1990, Azhari 2010).
  • (ii)  
    Optics
    • (a)  
      Each absorber should ideally have an absorption spectrum that is spectrally unique and similar in magnitude and spectral behaviour to biological tissue chromophores or exogenous contrast agents typically found in vivo. Also, the number of different absorbers and their spatial distribution in the phantom material should be representative of relevant in vivo scenarios (Laufer et al 2007). Reference values for biological tissue and exogenous contrast agents can be found elsewhere (Luke et al 2012, Jacques 2013).
    • (b)  
      The absorbers should be suitable to be embedded/encapsulated in the phantom matrix material of choice;
    • (c)  
      The absorbing species should be suitable for qPAI, regarding: (1) ${{\mu}_{a}}$ linearity with species concentration; (2) ${{\mu}_{a}}$ addition linearity when in combination with other absorbing species; (3) Low propensity for permanent photobleaching (Laufer et al 2010, Avigo et al 2015); (4) Low propensity for peak power dependent transient effects such as ground-state bleaching (Laufer et al 2010); (5) Known effect of the solvent or matrix on the absorption spectrum (Laufer et al 2010, Cook et al 2011);
  • (iii)  
    Photoacoustic efficiency
    • (a)  
      Well-characterised Grüneisen parameter $\Gamma$ (thermoelastic efficiency) of the matrix material and its dependence with the inclusion of additives, e.g. optical absorbers (Laufer et al 2010). $\Gamma$ does not need to match soft tissue exactly (higher $\Gamma$ can actually bring useful gains in SNR) but should be known, since most suggested theoretical qPAI algorithms already assume $\Gamma$ is either known a priori or constant throughout the domain (Cox et al 2012). Knowing $\Gamma$ is also important as ground-truth in inversion frameworks where $\Gamma$ is included as an additional unknown to be estimated (Zemp 2010, Mamonov and Ren 2014). Typical $\Gamma$ values for some tissues can be found in Yao et al (2014).
    • (b)  
      Minimal fluorescent behaviour (low quantum yield) is desirable, such that the thermalised energy is maximised and confounding parameters in the qPAI problem are minimised.
  • (iv)  
    Suitability for imaging applications
    • (a)  
      Feasibility of manufacturing and constructing various phantom structures with interest for imaging purposes, beyond classical image quality assessment geometries.

2.1. Candidate materials for qPAI

The various materials that were considered in this study for qPAI are given in table 1, alongside some of their advantages and disadvantages. The relevant information was obtained either from the studies referred to above that assess the use of the material for PA quality control phantoms, or from reviews on acoustic (Culjat et al 2010) and optical phantoms (Pogue and Patterson 2006). It should be noted that nominally identical materials may exhibit slightly different properties, especially if different suppliers or preparation techniques have been used. Materials such as polyester resin and epoxy resin, despite being popular optical phantom options due to their optical transparency and long term stability (Firbank et al 1995, Pogue and Patterson 2006), are not appropriate for photoacoustics due to their high sound speed and attenuation, and the support of fast shear waves (Culjat et al 2010), and were therefore not included in the list.

Table 1. Advantages and disadvantages of several background media materials, in the context of photoacoustic imaging.

Matrix Advantages Disadvantages
Water/Coupling-gel • Easily obtainable; • Cannot be used to easily sustain an insert;
  • Largely optically transparent; • Does not allow tuning acoustic properties;
  • Well defined optical and acoustic properties; • Speed of sound varies significantly with temperature (Culjat et al 2010);
  • Sound speed similar to biological tissue, though somewhat lower (Culjat et al 2010) • Allows only reduced flexibility in architecture—shape, layering.
Hydrogels e.g. agar, bovine gelatin (Laufer et al 2010, Cook et al 2011) • Solidity for suspending inserts or scattering particles; • Can have dye absorbers added to it; • Concentration range typically used significantly affects sound speed (Laufer et al 2010);
• Well-defined inserts are short-lasting due to diffusion;
  • Largely optically transparent, with the precise levels dependent on the base material used; • Absorption of water; Water soluble dyes need encapsulation (Bohndiek et al 2013);
  • Tissue-like speed of sound. • Reaction with nickel and copper ions, changing optical absorption (Laufer et al 2010);
  • Relative ease of preparation. • Differing dye optical absorption in solution from absorption in gelatin (Cook et al 2011);
    • Cross-linking time;
    • Susceptibility to dehydration and bacterial growth in storage, thus requiring additional careful mitigating measures (Culjat et al 2010, Cook et al 2011);
    • High susceptibility to physical damage (Culjat et al 2010);
    • Low temperature stability at physiological temperatures, causing structural integrity loss;
    • Limited re-use capability, heavily conditioned by imaging, handling and storage conditions (Culjat et al 2010).
Polyvinyl alcohol (PVA) (Kharine et al 2003, Xia et al 2011) • Solidity for suspending scatterers or absorbers; • Extensive preparation (several freeze-thaw cycles spanning each several hours) (Culjat et al 2010);
  • Adjustable intrinsic ${{\mu}_{s}}$ ; • Sensitive to humidity (Culjat et al 2010);
  • Tissue-like speed of sound; • Inhomogeneities due to differential heating and cooling rates (Xia et al 2011).
  • Greater longevity and structural rigidity than hydrogels;  
Polyvinyl chloride plastisol (PVCP) (Spirou et al 2005, Bohndiek et al 2013) • Solidity for suspending scatterers or absorbers; • Non trivial preparation;
• Insoluble in water; • 180 °C temperature needed for preparation might cause dye denaturation so only pigments should be used;
  • Largely optically transparent; • Higher acoustic attenuation than tissue;
  • Tunable acoustic properties: hardener/softener; • Lack of a widely available supply chain by reference chemical suppliers.
  • Stable during storage, up to 6 months (Bohndiek et al 2013);  
  • Can be used and re-used for imaging.  
Silicone e.g. room-temperature vulcanising (RTV), polydimethylsiloxane (PDMS) (Avigo et al 2015) • Solidity for suspending scatterers or absorbers; • Limited by high acoustic attenuation (>9.8 @ 3 MHz,>39 @ 11 MHz) (Tsou et al 2008) and low speed of sound (<1000 ms−1) (Culjat et al 2010);
• Insoluble in water; • Organic dyes might not be suitable for addition (Pogue and Patterson 2006);
  • Can be embedded with inorganic optical scatterers and absorbers; • For RTV: Cost (Pogue and Patterson 2006);
  • Capacity to be embedded with acoustic scatterers like e.g. glass an plastic microspheres; • Hardening time.
  • Stable during storage;  
  • Can be used and re-used for imaging;  
  • Variable Young's modulus (hardness), englobing biological tissue levels;  
  • Relative ease of preparation.  
  • For PDMS: Machineability and capability of creating microfluidic channels  

2.2. PVCP as a candidate material for qPAI

In this work, PVCP was chosen as the phantom material to be further characterised for qPAI, given its promising long-term stability, optical transparency and tissue-like acoustic properties. Spirou et al (2005) first suggested its use for PA, and it has more recently been proposed as a strong candidate for commercial PA system characterisation and quality control (Bohndiek et al 2013, Vogt et al 2015a, 2016). PVCP has also been considered for other applications, e.g. as a phantom material in optical coherence tomography (Wróbel et al 2015) or ultrasound-guided needle insertion (Hungr et al 2012).

Some optical properties of PVCP have been previously characterised at single (Spirou et al 2005, Bykov et al 2011, Bohndiek et al 2013) and more recently at multiple wavelengths (Vogt et al 2015b). Absorption has been tuned through the addition of black pigment colour, whilst reduced scattering has been increased mainly through the dispersion of titanium dioxide (TiO2), though zinc oxide (ZnO) has also recently been suggested as an alternative scatterer, in the context of the development of an optical coherence tomography phantom (Wróbel et al 2015). Speed of sound and acoustic attenuation have been characterised at low frequencies for commercial PVCP formulations (Spirou et al 2005, Hungr et al 2012). Acoustic backscatter, the major source of contrast for conventional ultrasound, has been estimated for PVCP with added microspheres (Vogt et al 2016). It has also recently been shown that by using custom-made dispersions of PVC in various plasticiser types, a wider range of sound speed and acoustic attenuation values could be obtained (Vogt et al 2016). The use of PVCP—or in fact any material—for the assessment and validation of qPAI has nevertheless not been explicitly addressed and investigated.

3. PVCP fabrication

The fabrication of PVCP phantoms was based on a previously reported protocol outlined by Bohndiek et al (2013). The main steps in the procedure were: (i) prepare an oil bath, heated to  ∼200 °C, on a magnetic stirrer heating plate; (ii) pour 30 ml PVCP (Lure Flex Firm, Lure Factors, Doncaster, UK) into an Erlenmeyer flask alongside a magnetic stirrer bar; (iii) place the Erlenmeyer in the oil bath and allow it to stir continuously; (iv) turn on the vacuum line connected to the neck of the Erlenmeyer to eliminate air bubbles; (v) as viscosity and translucency increase, reduce stirring speed; (vi) once the mixture becomes liquid again and considerably transparent, restore initial stirring speed; (vii) when PVCP reaches  ∼180 °C and the solution is homogeneous (single phase), release the vacuum and swiftly pour PVCP into an alluminium or silicone mould; (viii) Leave it to cure for several minutes.

The addition of optical absorbers or scatterers were used to tune the properties of PVCP. These were mixed together with the PVCP mixture, placed in a sonicator for 10 min to ensure maximum homogenisation, and only then put in the oil bath.

Though the volume of PVCP fabricated and its level of hardness can be increased to suit specific needs, maintaining a homogeneous single phase over time becomes increasingly challenging.

4. Sound speed and acoustic attenuation

PVCP was characterised in terms of its frequency-dependent sound speed c( f ) and attenuation $\alpha (\,f)$ . By varying the composition through the addition of a softener agent (Softener, Lure Factors, Doncaster, UK) at either 5, 10 or 20%v/v, the potential to tune the acoustic properties was assessed. Samples of two thicknesses were manufactured (10 and 20 mm, two specimens each). Furthermore, two 2 mm thick samples were prepared for the softest formulation only to extend the range of frequencies. Spirou et al (2005) showed that optical additives such as TiO2 and pigment based absorbers did not significantly affect the acoustic properties, therefore these formulations were not fabricated.

c( f ) and $\alpha (\,f)$ were measured with a through-transmission substitution setup available at the National Physical Laboratory (NPL), UK (Zeqiri et al 2010). A broadband transducer (active diameter 12 mm, Medicoteknisk Institute, Denmark) was used as a source and driven by a single cycle electrical pulse with central frequency 9 MHz at a pulse repetition frequency of 1 kHz. Detection was performed with a broadband hydrophone (30 mm active element diameter bilaminar membrane hydrophone, Marconi). Both the hydrophone and the transducer were submerged in a water tank, and aligned.

For each manufactured specimen, measurements were made with and without the sample in the sample holder within the water tank (figure 1). In each case, four oscilloscope readings were taken, where each consisted of an average of 100 waveforms. This procedure was repeated with the hydrophone at 3 distances from the source.

Figure 1.

Figure 1. Through-transmission substitution setup used for sound speed and acoustic attenuation measurements.

Standard image High-resolution image

From the measured difference in phase of the detected acoustic wave when the sample is present, ${{\phi}_{m}}(\,f)=2\pi f\left(\frac{L-\Delta x}{{{c}_{w}}}+\frac{\Delta x}{{{c}_{m}}}\right)$ , or absent, ${{\phi}_{w}}(\,f)=2\pi f\frac{L}{{{c}_{w}}}$ , the sound speed of the material can be estimated (Zimmermann and Smith 1983, Zeqiri et al 2010):

Equation (1)

where $\Delta x$ is the sample thickness, f is the frequency, L the distance between transducer and receiver, cw( f ) the speed of sound in water (Bilaniuk and Wong 1993) and $\text{d}\phi (\,f)={{\phi}_{w}}(\,f)-{{\phi}_{m}}(\,f)$ the phase difference.

The same data can also be used to retrieve the acoustic attenuation coefficient ${{\alpha}_{m}}$ . When no sample is present, the measured amplitude is ${{V}_{w}}(\,f)\propto {{P}_{i}}{{10}^{-{{\alpha}_{w}}L}}$ , where Pi is the input pressure and ${{\alpha}_{w}}(\,f)$ the acoustic attenuation in water (Pinkerton 1949). Once the sample is put in place, the amplitude is given by ${{V}_{m}}(\,f)\propto T{{P}_{i}}{{10}^{-\left[{{\alpha}_{w}}(L-\Delta x)+{{\alpha}_{m}}\Delta x\right]}}$ , where T accounts for the transmission losses at the two water-material interfaces. The frequency-dependent transmission-loss (TL) is therefore given by:

Equation (2)

By computing transmission loss for two sample thicknesses $\Delta {{x}_{1}}$ and $\Delta {{x}_{2}}$ , ${{\alpha}_{m}}$ can be retrieved:

Equation (3)

The frequency-dependent sound speed of PVCP up to 20 MHz and the acoustic attenuation up to 15 MHz are shown in figure 2 and summarised in table 2. The table indicates the speed of sound c at 5 MHz and the attenuation behaviour in terms of power law parameters a and b ($\alpha (\,f)=a~{{f}^{b}}$ , where f is in MHz). Literature-reported values for PVCP formulations from other suppliers are listed as well for comparison. The reported uncertainty refers to one standard uncertainty, providing a coverage probability of approximately 68% (coverage factor k  =  1). The uncertainty evaluation has been carried out in accordance with UKAS requirements (UKAS 2012). Uncertainties in sample thickness, temperature, literature reference values for water, repeated measurement variability and inter- and intra-sample variability were accounted for. The speed of sound and attenuation were broadly comparable to tissue-like structures. Nevertheless, c was up to ∼10% lower than the typical soft tissue range (e.g. 1450 ms−1 for fat or 1590 ms−1 for liver (Azhari 2010)). The attenuation at 1 MHz was well within desired tissue range—between 0.6-1 among soft tissues—, nevertheless the power law b was above the typical range of b  =  1–1.35 (Azhari 2010), which might limit the depth to which high-resolution can be achieved. Increasing hardness through the increase of the relative volume of PVC particles in the plasticiser caused an increase in the sound speed and attenuation, although it remained below the speeds typical of tissue. Further increases in hardness through an even higher relative percentage of PVC particles would cause difficulties in the fabrication process. Recently however, it has been shown that the sound speed can be increased to more tissue-matched levels by using other types of plasticisers to suspend the PVC particles (Vogt et al 2015b, 2016), though at the cost of an increasing attenuation that, though suitable for breast tissue up to 9 MHz, would be too high to be representative of most other soft tissues and further limit depth resolution and SNR.

Figure 2.

Figure 2. Sound speed (a) and acoustic attenuation (b) for PVCP samples with various levels of hardness. The 2 mm soft PVCP slabs extend the frequency range for $\alpha (\,f)$ from 10 to 15 MHz.

Standard image High-resolution image

Table 2. Speed of sound and acoustic attenuation, within one standard uncertainty.

Composition Softener %v/v Sound speed c (m s−1) Power law pre-factor a (dB cm−1 MHzb) Power law exponent b Samples # Temperature (°C)
Hard 5 $1408.3\pm 0.8$ @ 5 MHz $0.67\pm 0.02$ $1.55\pm 0.01$ 4 (20.5–$\left. 20.8\right)\pm 0.03$
Medium 10 $1406.7\pm 0.7$ @ 5 MHz $0.62\pm 0.03$ $1.54\pm 0.02$ 4 (20.5–$\left. 20.8\right)\pm 0.03$
Soft 20 $1402.3\pm 0.7$ @ 5 MHz $0.55\pm 0.05$ $1.53\pm 0.01$ 4 (20.6–$\left. 20.9\right)\pm 0.03$
Spirou et al (2005)a $1400\pm 20$ @ 1 MHz $0.57\pm 1.01$ $1.51\pm 0.06$ 3
Madsen et al (2003)b 1395 1.05 dB cm−1 MHz−1 @ 4.5 MHz 1 22
Vogt et al (2015b)c 1380–1575 1–30 dB cm−1 MHz−1 @ 4.0 MHz

Note. Literature values are given for other PVCP formulations: aMF Manufacturing; bunknown origin; crange of custom formulations.

5. Optical properties

PVCP has intrinsic, relatively low, optical scattering and absorption which can be increased using additives.

5.1. Optical scattering

Optical scatterers can be used to tune the scattering behaviour towards more tissue-like levels. TiO2 powder (Titanium IV Oxide, Anatase, 232033, Sigma-Aldrich, Germany) is most commonly used and readily available (Spirou et al 2005) and was thus used to fabricate soft PVCP samples (2 mm thick) at 4 different concentration levels—0.25, 0.5,1 and 2 mg ml−1. Integrating sphere measurements coupled with the inverse adding-doubling (IAD) algorithm (Prahl et al 1993) were employed to determine the reduced scattering coefficient $\mu _{s}^{\prime}$ , and simultaneously the intrinsic absorption coefficient ${{\mu}_{a}}$ . Total transmittance (MT) and total reflectance (MR) measurements were carried out at four different spots on each sample (2 per side) between 400–2000 nm with a 100 mm integrating sphere mounted in a dual-beam spectrophotometer (Lambda 750, Perkin Elmer, Waltham, MA, USA). IAD is an inverse light modelling strategy that recovers $\mu _{s}^{\prime}$ and ${{\mu}_{a}}$ by iteratively finding the adding-doubling solution to the radiative transfer equation (RTE) under the assumption that the sample under study is a layer-like structure where each layer has homogeneous absorption and scattering. The scattering anisotropy factor, necessary as an input parameter, was set to g  =  0.6, based on Mie Theory calculations from the literature for TiO2 in PVCP (Bykov et al 2010).

The reduced scattering spectra of PVCP with added TiO2 are given in figure 3(a). Scattering for all samples was found to be comparable to biological tissue in terms of the order of magnitude expected (Jacques 2013). There is an approximate linearity of $\mu _{s}^{\prime}$ with increasing TiO2 concentration. It should be noted that the 1 mg ml−1 sample that was prepared had visible homogeneity issues likely caused by lack of proper dispersion of the mixture—which was subsequently rectified with increased mixing and sonicating—and its curve has therefore been omitted since it gave an erroneous portrayal of the expected behaviour (curve can be seen in Fonseca et al (2015)). The values obtained are in the same range as those in other $\mu _{s}^{\prime}$ characterisation studies (Bykov et al 2010, Vogt et al 2015b), but the absolute values differ among them. This is due to the fact that different TiO2 products can have different particle size distributions and even refractive indices (Phillips and Griffen 1981), and both parameters affect $\mu _{s}^{\prime}$ . When designing a phantom, care must therefore be taken when referring to scattering measurements using different TiO2 commercial products. Scattering characterisation should therefore ideally be performed per batch, even for products where the supplier specifies a mean particle size to a high degree of accuracy. Literature values should rather serve as a guideline for spectral behaviour and for preparing samples in the desired order of magnitude.

Figure 3.

Figure 3. (a) Reduced scattering coefficient spectra of PVCP with varying concentrations of TiO2. (b)–(c) Intrinsic absorption spectrum of soft PVCP. Variations at 850 nm are in fact artefacts due to the change of detectors within the spectrophotometer. The dashed lines correspond to the standard deviation between the estimated ${{\mu}_{a}}$ for the four slabs.

Standard image High-resolution image

5.2. Optical absorption

The intrinsic absorption spectrum of PVCP is shown in figures 3(b) and (c). PVCP has intrinsic absorption peaks at 910, 1190, 1400 and 1720 nm, probably due to vibrational energy transitions (overtones) in PVC (Goddu and Delker 1960). Overall, the absorption is low and analogous in range and magnitude to background biological tissue (Jacques 2013). A standardised preparation of the material is advised (Spirou et al 2005, Van de Ven and Erdman 2006) since the absorption may vary to a certain extent with the sample preparation process, namely due to factors such as heating rate and final heating temperature. For example, PVC suffers a significant decrease in optical transmission when held at temperatures in the 190–205°C range (Van de Ven and Erdman 2006).

The absorption of PVCP can be tuned through the use of additives. Pigment dispersions—where pigments are pre-dispersed in a paste and then mixed with softener—are advised as colorants (Berins 1991, Pigments and Additives Division: Clariant International Ltd 2007). Soluble dyes on the other hand are not suitable for use in plasticised PVC formulations since they migrate in the presence of the plasticiser (Pigments and Additives Division: Clariant International Ltd 2007). Three pigment-based absorbers (black, red and blue) from the same supplier as the PVCP (Liquid Colour, Lure Factors, Doncaster, UK) were therefore characterised. 2 mm thick slabs of soft PVCP were fabricated, with added absorbers at 0.5, 0.25 and 0.125%v/v concentration (serial dilution). Their absorption was characterised with the spectrophotometer, with four points being assessed per sample.

Embedding either blue, black or red pigment absorbers in PVCP led to three unique spectra (figures 4(a)(c)). For all pigments, ${{\mu}_{a}}$ was found to be linear with pigment concentration (figures 4(d)(f)).

Figure 4.

Figure 4. (a)–(c) Absorption spectra of PVCP slabs with embedded absorbers, at different concentrations; [S]—spectrophotometer acquired data. Dashed lines represent the standard deviation of the four measurements on the same sample. (d)–(f) Linearity of spectroscopic ${{\mu}_{a}}$ with increasing concentration at the peak wavelength. (a) Black. (b) Red. (c) Blue. (d) Black. (e) Red. (f) Blue.

Standard image High-resolution image

5.3. Absorption at high peak power

PA imaging is usually performed using pulsed rather than continuously-modulated illumination. Efficient PA signal generation requires a condition of stress confinement to be satisfied, which requires the use of pulses of  ⩽10 ns duration. Since the phantom is meant for photoacoustic imaging, it needs to be able to cope with exposure to high peak power sources. In addition, for imaging systems that require repeated and prolonged pulsed exposure (e.g. reliant on raster scanning for PA acquisition), the ability to withstand the accumulation of average power is necessary.

To investigate the effect of using a high peak power source (as used in PA) rather than a low power continuous wave source (as used in the spectrophotometer), absorption coefficient measurements were made using photoacoustic spectroscopy (Stahl et al 2014) (figure 5), on 1%v/v 20 mm slabs of the same fabrication batch as the spectrophotometer samples. A fibre-coupled Nd:YAG pumped wavelength-tunable optical parametric oscillator (OPO) system (GWU, Spectra-Physics, Santa Clara, CA, USA) was used for excitation light delivery, tunable between 500–680 nm and 740–2100 nm and providing 7 ns long pulses at 10 Hz pulse repetition rate. A 10 mm  ×  10 mm  ×  50 μm polyvinylidene fluoride (PVDF) film bonded to a PMMA backing (Stahl et al 2014) was used for acoustic detection of the PA signal. The recorded PA signal $S(t)=\beta \Gamma{{\Phi}_{0}}{{\mu}_{a}}{{\text{e}}^{-{{\mu}_{a}}ct}}$ [V] can then be used to retrieve ${{\mu}_{a}}$ through curve-fitting to the exponential decay, where β is a scaling parameter related to system sensitivity and ${{\Phi}_{0}}$ is the incident fluence. Additionally, a different high peak power ns-pulsed laser source (SpitLight 600, Innolas, Krailling, Germany) was also employed to make optical transmission measurements, with the light detection being done by a powermeter (PowerMAX PM10V1, Coherent, Santa Clara, CA, USA). Three repeat measurements were made with the 2 mm slabs used for spectroscopy.

Figure 5.

Figure 5. Photoacoustic spectroscopy diagram. The inset shows the case where a liquid sample is imaged in a cuvette, instead of having a PVCP sample placed directly in the water bath.

Standard image High-resolution image

Figures 6(a)(c) show, for the three absorbers, that spectrometer-derived ([S]) and photoacoustic-derived ([P]) ${{\mu}_{a}}$ spectra differ. The red absorber shows the most notorious alteration, since the change in absolute absorption is accompanied by a clear alteration in the shape of the spectrum. The 550–580 nm plateau from the spectroscopic measurements does not appear in the photoacoustically-derived spectrum, giving way to a rather peak-like feature at 550 nm. In addition, the results of transmittance measurements performed with a high peak power source on the 2 mm slabs showed characteristics similar to the photoacoustic measurements (figures 6(a)(c), [T]). These results seem to indicate that the absorbers may be suffering photo-chemical or photo-physical changes in behaviour when exposed to high peak power nanosecond pulses (Khairutdinov and Serpone 1995, Marcano et al 2000, Berera et al 2009, Laufer et al 2010, Kislyakov and Yelleswarapu 2013) (∼1 MW peak power) that do not occur when exposed to the low power continuous-wave illumination of the spectrophotometer (∼W peak power).

Figure 6.

Figure 6. Comparison of extinction coefficient measured through: transmittance in spectrometer [S] (linear regression of results from samples at all concentrations) with cw-illumination, exponential fit of photoacoustic signal [P] and transmittance [T] with a high peak power OPO source. Measurement standard deviation is given. (a) Black. (b) Red. (c) Blue.

Standard image High-resolution image

Control measurements were first made with a CuCl2 solution (previously shown to have transient and permanent photostability (Laufer et al 2010)) on both the spectrophotometer and photoacoustic systems, to ensure both systems agreed and were working up to specification in the relevant absorption range. Intrinsic absorption of PVCP (no added pigments) around the 1720 nm peak was also in good agreement between systems. Measurements of water absorption between 1350 and 2000 nm (representing a range of ${{\mu}_{a}}$ up to 12 mm−1) were also made with the PA spectroscopy system to ensure that the exponential fitting methodology was accurate compared to literature reference values (Palmer and Williams 1974).

To further study the behaviour of the pigment-embedded PVCP when exposed to high peak power, photoacoustic absorption measurements were made as a function of illumination time, as a function of incident peak energy and as a function of pigment concentration.

The influence of the fluence on ${{\mu}_{a}}$ was assessed by obtaining photoacoustic absorption spectra at varying levels of incident fluence—100, 80, 60, 40 and 20% of ${{\Phi}_{\text{max}}}$ , where the non-attenuated energy is  ∼27 mJ at 500 nm and  ∼16 mJ at 680 nm. In a first experiment, one point on the sample was illuminated with increasing peak energy levels. In a second scenario, a different point was illuminated with decreasing peak energy levels. Finally, an experiment was performed where different points were illuminated with distinct peak energy levels—with minimised common history between points. These three modalities were employed to differentiate effects due to high peak power (leading to transient bleaching) from effects due to the accumulation of average power over an extended time (leading to permanent bleaching).

An increase in energy per pulse led to a decrease in ${{\mu}_{a}}$ (figure 7). The decreasing trend in ${{\mu}_{a}}$ when fluence is increased gradually over time (figures 7(a) and (d)) could be assigned to permanent photobleaching by accumulation of average power, nevertheless that would not explain why ${{\mu}_{a}}$ is partially restored at timepoint [t6] when energy is reset to its lowest level. It also would not explain why when one point is illuminated with gradually decreasing pulse energy the ${{\mu}_{a}}$ levels increase (figures 7(b) and (e)). Illuminating different points on a sample with distinct pulse energies also shows that ${{\mu}_{a}}$ increases as pulse energy is lowered (figures 7(c) and (f)). This suggests that there is a dependency of ${{\mu}_{a}}$ on peak pulse energy, likely due to the transient/reversible occurrence of ground-state bleaching (saturable absorption) (Berera et al 2009).

Figure 7.

Figure 7. (a)–(c) Influence of varying peak power energy on the ${{\mu}_{a}}$ of blue PVCP samples. Results are shown as peak energy is gradually increased over a same location for timepoints [t1–5] and restored to its initial (lowest) value in [t6] (a), as peak energy is gradually decreased over a second location for timepoints [t1–5] and restored to its initial (highest) value in [t6] (b), and as peak energy is varied over distinct locations/spots on a same sample (c). (d)–(f) Influence of varying peak power energy on ${{\mu}_{a}}$ of PVCP samples embedded with either of the three colours. Red and black values are shown for 550 nm whilst blue values are shown for 610 nm.

Standard image High-resolution image

To assess ${{\mu}_{a}}$ as a function of extended illumination time, black, red and blue PVCP slabs were continuously irradiated with the Spectra-Physics OPO laser for 30 min, 10 Hz pulse repetition rate, at their respective peak absorption wavelength—500 nm, 550 nm and 615 nm. The average beam energy was 5 mJ. PA signals were averaged over a minute and stored.

The PVCP with embedded pigments displayed reasonable stability over a 30 min irradiation period (18 000 pulses)—data not shown. A decreasing ${{\mu}_{a}}$ trend was seen, but it did not exceed 5%. Given the timescale of the measurement and the accumulation of average power, the decrease is probably due to permanent photobleaching. Transient effects such as ground state bleaching or saturable absorption take place in the sub-nanosecond to nanosecond time regime (Khairutdinov and Serpone 1995) as a result of high peak power, thus their temporal onset and trend would not be noticed in this experimental setup.

To assess the linearity of PA-derived ${{\mu}_{a}}$ with increasing concentration, a new batch of samples was prepared. Red and blue 20 mm thick PVCP slabs were fabricated at 2, 1 and 0.5%v/v pigment concentration. 3 acquisitions were made per sample in the 500–680 nm range, each consisting of the averaging of 100 waveforms.

Figure 8 shows the level of linearity of the photoacoustically-derived ${{\mu}_{a}}$ with absorber concentration. Spectra have been normalised to the mean spectrum for 2%v/v absorber concentration, and if linearity holds should respect the 1/2 and 1/4 dilution factors. Only wavelengths where the coefficient of correlation of the exponential fit satisfies R2  >  0.98 are shown.

Figure 8.

Figure 8. Spectra at different absorber concentrations (0.5, 1 and 2%v/v) normalised to the spectrum at 2 %v/v. 1/2 and 1/4 concentration ratios are broadly respected. Error bars account for inter-acquisition standard deviation and for the standard uncertainty in the exponential fitting procedure. (a) Blue absorber, (b) red absorber.

Standard image High-resolution image

The blue samples display linearity with concentration whilst the red seem to display small deviations from absorption linearity in the 540–580 nm region: the lower the concentration, the lower the relative absorption at 570–580 nm compared to 540–550 nm. This coincides with the observations made when comparing spectrophotometer ([S]) to photoacoustic data ([P]) (figure 6(b)). Saturable absorption manifests itself not only at increasing peak power levels, but also when concentrations are lowered, since this lowers the ratio of absorbing molecules per unit volume to the number of photons.

6. Grüneisen parameter

The Grüneisen parameter $\Gamma$ quantifies the relationship between the optically absorbed and thermalised optical energy to pressure and has a direct impact on the PA amplitude. Its characterisation is therefore important in a qPAI context and was pursued in this study. The PA spectroscopy system described in the previous section was used, with a method comparative to water being employed to retrieve γ Stahl et al (2014). Briefly, PA measurements were made for water and the sample of interest, with fluence-normalised amplitudes defined by $S_{0,\text{water}}^{\prime}={{S}_{0,\text{water}}}/{{\Phi}_{0}}=\beta {{\Gamma}_{\text{water}}}{{\mu}_{a,\text{water}}}$ and $S_{0,\text{sample}}^{\prime}={{S}_{0,\text{sample}}}/{{\Phi}_{0}}=\beta {{\Gamma}_{\text{sample}}}{{\mu}_{a,\text{sample}}}$ respectively. The sensitivity parameter β was retrieved from the PA measurement of water, ${{S}_{0,\text{water}}}$ , through least-squares regression in the 1350–1850 nm region, where ${{\mu}_{a,\text{water}}}$ and ${{\Gamma}_{\text{water}}}$ are known from the literature (Haynes 2012, Palmer and Williams 1974). The derived β value, alongside ${{\mu}_{a,\text{sample}}}$ values retrieved through curve-fitting at the relevant wavelengths, were then used in the equation for ${{S}_{0,\text{sample}}}$ to retrieve ${{\Gamma}_{\text{sample}}}$ .

Unlike the case where a water-filled cuvette is placed in the water bath, when other samples are placed there is a considerable acoustic impedance mismatch at the interface with the water bath. This will result in a change in acoustic transmission, given by ${{T}_{\text{s},\text{w}}}=\frac{2~{{Z}_{\text{w}}}}{{{Z}_{\text{w}}}+{{Z}_{\text{s}}}}$ , where Zw and Zs are the characteristic acoustic impedances of water and the sample respectively. To correct for this error, the measured amplitude S was divided by the appropriate transmission ${{T}_{\text{s},\text{w}}}$ (Yao et al 2014). Method validation was carried out with methanol and ethanol, where the reference values were taken from the literature and are shown in table 3.

Table 3. Tabulated properties in the literature versus experimental results (subscript-exp). Table reproduced from Fonseca et al 2015, by permission of SPIE.

Material T (°C) Cp (J mol−1 K−1) c (ms−1) ${{\alpha}_{v}}$ (°C−1) ρ Z (MRayl) ${{\Gamma}_{\text{lit}}}$ Texp (°C) ${{\rm{\hat \Gamma}}_{\text{exp}}}$
Water 20 75.3 1482.3 $20.6\times {{10}^{-5}}$ 1 0.116    
  21.7 1487.47 Bilaniuk and Wong (1993)   1   0.12 $21.7\pm 0.1$
Ethanol 20 112.3 1159 $140\times {{10}^{-5}}$ 0.7893 0.91 0.772 $21.5\pm 0.2$ $0.78\pm 0.04$
Methanol 20 81.1 1116 $149\times {{10}^{-5}}$ 0.7917 0.88 0.733 $19.9\pm 0.3$ $0.71\pm 0.02$
PVCP 20 1402 ${{}_{\left(\text{exp}\right)}}$ 1.008 ${{}_{\left(\text{exp}\right)}}$ 1.41 ${{}_{\left(\text{exp}\right)}}$ $21.9\pm 0.1$ $1.01\pm 0.05$

Note. All literature values taken from Haynes (2012) unless otherwise indicated. T—temperature; Cp—specific heat capacity at constant pressure; c—sound speed; ${{\alpha}_{v}}$ —volume thermal expansion coefficient; ρ—specific gravity; Z—acoustic impedance $Z=\rho c$ ; ${{\Gamma}_{\text{lit}}}$ —Literature-derived Grüneisen coefficient $\Gamma={{\alpha}_{v}}{{c}^{2}}/{{C}_{p}}$ ; ${{\rm{\hat \Gamma}}_{\text{exp}}}$ —Estimated Grüneisen coefficient.

Methanol, ethanol and water, were poured in turn into a cuvette which was then placed within a water bath. 3 acquisitions were carried out per substance, each acquisition being comprised by the averaging of 300 waveforms. 6 acquisitions were made on a 20 mm slab of PVCP for the 1700–1750 nm region.

Table 3 gives the results of the Grüneisen parameter characterisation. The estimated $\Gamma$ values for ethanol and methanol are consistent with the literature. $\Gamma$ of PVCP is higher in value than some tissues (Yao et al 2014), though this has as advantage that imaging can be done with higher SNR. Error ranges for the obtained $\Gamma$ estimates take into account inter-acquisition standard-deviation as well as the standard uncertainty of the least-squares regression process applied to each acquisition.

The variation in Grüneisen parameter with added blue or red pigments was also studied for 20 mm thick PVCP slabs with concentrations of 2, 1 and 0.5%v/v. These measurements were made at a single wavelength where intrinsic PVCP absorption is highest and pigment absorption is residual—1715 nm, which means that measured variations in ${{S}_{0,\text{sample}}}$ with pigment concentration will be mainly due to changes in $\Gamma$ rather than ${{\mu}_{a}}$ (Laufer et al 2010). Given that the same wavelength was used for all acquisitions, $\Gamma$ was retrieved through the relation (Stahl et al 2014):

Equation (4)

where β cancels out.

Experiments showed that there was a negligible change in Grüneisen parameter with pigment addition up to 2%v/v (data not shown, see Fonseca et al (2015)).

7. Phantom imaging for various constructs

To establish feasible manufacturing strategies and configurations for PVCP, three simple phantoms were developed. These were made with soft PVCP with 0.5 mg ml−1 of TiO2 (Sigma Aldrich, Germany), into a purchased silicone mould with 39 mm diameter cylindrical pockets. The motifs embedded were respectively: a simple sub-500 μm-wide line made of self-adhesive black vinyl film; a 'UCL' pattern made of self-adhesive black vinyl film; a pair of 2 mm thick soft PVCP square inserts with 2% and 8%v/v black pigment respectively. In terms of manufacture, a first layer of PVCP was poured and allowed to cure, on which the inserts of interest were then placed (figure 9). A second layer of PVCP was then deposited to fully encapsulate the inserts and allowed to cure.

Figure 9.

Figure 9. Phantoms are shown before the second layer of PVCP was poured.

Standard image High-resolution image

Photoacoustic imaging was performed using a Fabry–Pérot scanner system (Zhang et al 2008), with forward mode illumination at 500 nm being provided by an OPO system providing 6 ns pulses at 30 Hz (SpitLight 600, Innolas, Krailling, Germany). The beam diameter was approximately 1 cm. Imaging parameters were: spatial resolution of $\text{d}x=200~\mu $ m, $\text{d}y=200~\mu $ m, $\text{d}t=4$ ns, field of view $Nx\times Ny=24\times 22$ mm. A k-space back-projection algorithm (Treeby and Cox 2010) was used to reconstruct a 3D image of the initial pressure field from each set of acquired data. Speed of sound was set to c  =  1402 ms−1 and acoustic attenuation was not accounted for.

The reconstructed images obtained with the insert-embedded phantoms are shown in figure 10 as xy maximum-intensity projections (where the z-axis represents the depth direction). The main features can be perceived in all phantoms.

Figure 10.

Figure 10. xy maximum intensity projections of the obtained PA images. In (b), the 2%v/v and 8%v/v black PVCP inserts are positioned left and right respectively. (a) Line phantom. (b) Embedded squares phantom. (c) 'UCL' phantom.

Standard image High-resolution image

Given that PVCP is not prone to water absorption, it is also possible to create PVCP phantoms with wall-less channels (Bykov et al 2010, Vogt et al 2015a). This is advantageous since the characteristics of the absorbing regions of interest can be easily changed between experiments, by changing the injected dye or pigment. Fewer reconstruction artefacts will also be present when compared to phantoms with walled-channels (where the wall is a different material such as a polyethylene tube), unless the wall material is explicitly accounted for. A double-channelled phantom was built to illustrate its potential use for multiple wavelength PAI. A silicone mould was used, with two 1 mm needles traversing it. After PVCP was poured and cured, the needles were removed, leaving two 1 mm diameter channels in the sample that were then filled with a solution of 34.7 g l−1 copper chloride (CuCl$_{2}\centerdot \left[2{{\text{H}}_{2}}\text{O}\right]$ ) and 440.6 g l−1 nickel chloride (NiCl$_{2}\centerdot \left[6{{\text{H}}_{2}}\text{O}\right]$ ) respectively. Their spectra were evaluated with the spectrophotometer.

Photoacoustic imaging was performed in similar conditions as above. Imaging parameters were: spatial resolution of $\text{d}x=100~\mu $ m, $\text{d}y=100~\mu $ m, $\text{d}t=4$ ns, field of view $Nx\times Ny\times Nt=15$ mm $\times 15$ mm $\times 18~\mu $ s. Images were in this case acquired at multiple wavelengths, from 750 to 1000 nm in steps of 50 nm. Incident fluence variations with time and wavelength were corrected for. xy and yz maximum-intensity projection (MIP)s are shown in figures 11(a) and (b) respectively. Figures 11(c) and (d) compare the 850 nm normalised absorption spectrum measured in the spectrophotometer to a metric on the total intensity inside the tubes, found through a thresholding operation. The spectra are distinct, which is expected due to the effect of spectral colouring, to the reconstruction artefacts partially associated to limited-view and to the limitations in the accuracy of the tube segmentation and choice of total intensity as a metric.

Figure 11.

Figure 11. (a) xy MIP of a 3 mm slice for the channelled phantom, at 850 nm. Left: CuCl2; right: NiCl2; (b) yz MIP of the central 1 mm portion; (c)–(d) 850 nm normalised spectra for CuCl2 and NiCl2 respectively, comparing the spectrophotometer result ([S]) with the PAI intensity result ([PAI]).

Standard image High-resolution image

8. Discussion and conclusions

As mentioned in the introduction, a qPAI algorithm must give accurate estimates with low uncertainty, be applicable in various situations, and be robust to experimental uncertainties and noise. Its validity should be demonstrated over the full range of scenarios in which the method will be applied. Many of the reported qPAI strategies have not yet assessed or met all these criteria. Studies with phantoms with realistic and known properties would be very valuable for state-of-art qPAI propositions that have only yet been implemented and tested in simulation and that wish to extend their scope and optimise their performance in experimental scenarios (Saratoon et al 2013, Pulkkinen et al 2014, Shao et al 2014, Gao et al 2015). Likewise, phantom assessment would be of great aid to in vivo studies that have directly applied quantification algorithms of varying degrees of complexity and wish to assess their robustness, uncertainty and range of validity (Wilson et al 2014, Gottschalk et al 2015, Li et al 2016). With the advent of increasingly faster and higher resolution instrumentation (Deán-Ben et al 2015, Huynh et al 2016), as well as increasingly complex and efficient theoretical algorithms that allow high-dimensional and fine-resolution quantification both at single (Gao et al 2011, Saratoon et al 2013, Tarvainen et al 2014) and multiple wavelengths (Pulkkinen et al 2014), the need to find a phantom that is well controlled, characterised, tuneable and spatially-detailed is even more pressing and desirable.

In this paper, PVCP was characterised as a phantom material for qPAI assessment and validation studies. Its acoustic properties were assessed up to 15–20 MHz. Sound speed $c\left(1~\text{MHz}\right)=\left[1402.3\text{--}1408.3\right]$ ms−1 was within  ∼10% of soft tissue values and acoustic attenuation $\alpha (\,f)=\left[0.55\text{--}0.67\right]\,{{f}^{[1.53-1.55]}}$ dB cm−1 MHz−1 was analogous to soft tissues in the low frequency range, though higher than desired at increasing frequencies for most tissues except breast—as shown by the power law parameter $b=\left[1.53\text{--}1.55\right]$ . This sharp frequency-dependent behaviour may pose an issue given that many structures of interest in PAI generate characteristically high frequencies. Studies by Vogt et al (2016) have recently shown promise in finding alternative PVCP formulations—by varying the type of plasticiser used to disperse the PVC particles—that may provide more reasonable and tailored acoustic properties.

Intrinsic optical absorption and reduced scattering with added TiO2 as a scatterer were characterised between 400–2000 nm, corresponding to the full range of wavelengths that would be available in a typical OPO source for photoacoustics. Intrinsic absorption in the 400–1000 nm range was similar in magnitude to typical tissue background levels, around  ∼10−2 mm−1 order of magnitude, with higher absorption peaks above 1000 nm. Namely, the main peak at 1720 nm was made use of when performing the Grüneisen parameter characterisation. The reduced scattering was successfully tuned to tissue equivalent levels, around  ∼1 mm−1 order of magnitude.

Optical characterisation of three different absorbing pigments dispersed in PVCP revealed that they all displayed absorption linearity with concentration and distinct optical spectra, favouring their use as endogenous chromophore analogues. Nevertheless, further experiments focussing on the absorption behaviour at high peak power revealed differences compared to continuous-wave characterisation, possibly attributable to saturable absorption caused by ground-state depletion in the high-peak power regime (Piletic et al 2010). Since PA employs high peak power, absorber characterisation—including concentration linearity—for qPAI should always be done in this regime. The pigments displayed ${{\mu}_{a}}$ linearity with concentration even at high peak power. As for variation with incident peak power, the application may determine whether the variation is acceptable or not. In this study, a 5-fold variation in peak power led to up to about 13% variation in ${{\mu}_{a}}$ under the worst case scenario of direct surface incidence on the pigments. For the more realistic imaging phantom case of absorbing inserts at greater depth, where light has been attenuated and diffused, the variation is likely to be much less pronounced (as fluence lowers less saturable absorption phenomena occur—as seen in figure 7(f)) where the ${{\mu}_{a}}$ variation with fluence plateaus).

If a more stable absorption is required, one option is to explore different pigments. The particular pigments studied in this work were chosen to ensure compatibility to the matrix and resistance to the high temperature fabrication process—having come from the supplier as the PVCP—, nevertheless there is scope to explore in a wider-ranging photostability study pigments from different suppliers in the plastic materials colouring industry (Berins 1991, Pigments and Additives Division: Clariant International Ltd 2007). The main hurdle is short-listing the most promising variants from the large repository of colourants, especially since the type of tabulated features (such as hue, light and weather fastness, fastness to bleeding, tinting strength, particle size distribution, pigment dispersibility and minimum limiting concentration) do not give direct indication about their better or worse propensity for absorption non-linearities. The photostability assessment could in a first stage be done by comparing the behaviour of the absorbing species at low and high peak power, and by exploring the behaviour at high peak power as a function of concentration, time and power (Laufer et al 2010, Stahl et al 2014), as done in this study. For a more thorough and quantifiable characterisation between distinct and concurrent non-linear photophysical phenomena such as ground-state depletion, stimulated emission and excited state absorption (Alfano 2012), techniques such as pump–probe (time-resolved) spectroscopy could be employed (Piletic et al 2010, Märk et al 2015). These characterisation studies of various absorbers are not only important if a fully-characterised, stable and reliable qPAI phantom is desired, but also for in vivo studies, where a thorough knowledge of the photochemical and photophysical behaviour of the endogenous or exogenous chromophores of interest at high peak power is essential. Another option to create a phantom with more stable absorption is to use a PVCP phantom with wall-less channels filled with stable and well-characterised absorbing solutions, since this will circumvent the need to disperse absorbers in PVCP directly, with added advantages that a more vasculature-realistic geometry is achieved and that the absorbers can be changed quickly by flushing the solutions in and out the channels. The use of channelled phantoms was illustrated here by performing 3D multiple wavelength PAI.

Besides the channelled phantoms, fabricating and imaging (at a single-wavelength) different geometries with various embedded structures was also possible, showing the versatility of the material.

The Grüneisen parameter was characterised to be $\Gamma=1.01\pm 0.05$ for the soft formulation, its high value being beneficial in yielding a high SNR in qPAI acquisitions. Changes with increasing absorber concentration were not significant.

The final conclusion is that PVCP does have the potential to be a useful and versatile phantom material for qPAI, but for now its acoustic absorption may limit its use to lower frequency (larger scale) systems, and care must be taken because of the possibility of nonlinear optical behaviour at high peak powers for some absorbers.

Acknowledgments

The work presented was supported by a CASE studentship (grant EP/L505717/1) jointly funded by the UK's Engineering and Physical Sciences Research Council and National Physical Laboratory.

The authors would like to thank N Sadhoo for guidance on using the through-transmission substitution setup available at NPL.