Thermoluminescent response and relative efficiency of TLD-100 exposed to low-energy x-rays

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation I Gamboa-deBuen et al 1998 Phys. Med. Biol. 43 2073 DOI 10.1088/0031-9155/43/8/006

0031-9155/43/8/2073

Abstract

The dose-response of LiF:Mg,Ti (TLD-100) exposed to 15 and 35 kVp ( and keV effective energy respectively) x-rays and -rays has been measured in the dose interval from (1.2-5.4) Gy for x-rays, and from 0.14 to 850 Gy for -rays. In both cases the total TL signal and glow curve peaks 3 to 9 show supralinearity. The supralinearity function f(D) is similar for both x-ray beams, except for peak 8, where a 30% difference is observed. The maxima of f(D) for the total TL signal and peaks 5 to 8 are 2.1, 1.7, 6.4, 3.3 and 7.5 respectively for 8.1 keV x-rays and 3.7, 3.1, 13.6, 9.9 and 11.0 for -rays. The measured relative efficiencies for x-rays with respect to , for the total TL signal and peaks 5 and 7, were 1.04, 0.97 and 3.2 respectively.