Iterative reconstruction of PET images using a high-overrelaxation single-projection algorithm

, and

Published under licence by IOP Publishing Ltd
, , Citation Peter Schmidlin et al 1997 Phys. Med. Biol. 42 569 DOI 10.1088/0031-9155/42/3/009

0031-9155/42/3/569

Abstract

An iterative image reconstruction procedure is described which is able to calculate high-precision images within eight iterative steps. High initial overrelaxation parameters are used which drop down towards about one during continuing iteration. The parameters are determined pragmatically, postulating maximum gain of image quality during a sequence of iterative steps. Results with simulated and measured data show that parameters derived from one data set may be widely used for other data sets. The acceleration of iterative reconstruction achieved by the estimation of optimum overrelaxation parameters is important for large data sets, especially for fully 3D reconstruction.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/42/3/009