This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy. Close this notification
NOTICE: Ukraine: Read IOP Publishing's statement.

Institute of Physics and Engineering in Medicine

IPEM's aim is to promote the advancement of physics and engineering applied to medicine and biology for the public benefit. Its members are professionals working in healthcare, education, industry and research.

IPEM publishes scientific journals and books and organises conferences to disseminate knowledge and support members in their development. It sets and advises on standards for the practice, education and training of scientists and engineers working in healthcare to secure an effective and appropriate workforce.

A photoelectric plethysmograph for the measurement of cutaneous blood flow

and

Published under licence by IOP Publishing Ltd
, , Citation A V J Challoner and C A Ramsay 1974 Phys. Med. Biol. 19 317

0031-9155/19/3/317

Abstract

The photodetector used is a cadmium selenide photoconductive cell and allowance is made for the temperature coefficient of the cell by applying a correction factor. Nonlinearity of the photocell response has been shown to be of negligible importance. Heating of the skin by the light source is greatly reduced by the use of fibre optic light guides and the influence on the recording of changes in oxygenation of haemoglobin is kept to a minimum by an infrared filter. The instrument has been shown to be sensitive, atraumatic and applicable to almost any skin area. Quantitative measurements of blood flow are not possible but accurate comparative measurements of changes in flow are easily obtained when the area of skin under study is compared with an adjacent control area.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/19/3/003