Myocardial Strain Imaging Based on Two-Dimensional Displacement Vector Measurement

and

Published 28 May 2004 Copyright (c) 2004 The Japan Society of Applied Physics
, , Citation Naotaka Nitta and Tsuyoshi Shiina 2004 Jpn. J. Appl. Phys. 43 3249 DOI 10.1143/JJAP.43.3249

1347-4065/43/5S/3249

Abstract

The abnormalities of myocardial wall motion caused by changes in wall stiffness often appear in the early stage of ischemic heart disease. Since the myocardium exhibits complex and large motion, a two-dimensional (2D) or three-dimensional (3D) assessment of stiffness distribution is required for accurate diagnosis. Although a 3D assessment is ultimately required, as a stepped approach for practical use, we propose novel methods for tracking the 2D motion using a one-dimensional (1D) phased array and for assessing myocardial malfunction by visualizing the invariant of a strain tensor. The feasibilities of the proposed methods were evaluated by numerically simulating the short-axis imaging of a 3D myocardial model. This model includes a hard infarction located between 1 and 3 o'clock, which is difficult to detect by conventional tissue Doppler and strain rate imaging, and the motions of the model were assigned by referring to actual myocardial motion. These results revealed that the proposed imaging methods clearly depicted the hard infarction area which conventional imaging could not detect.

Export citation and abstract BibTeX RIS

10.1143/JJAP.43.3249