This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.
Brought to you by:
Paper

Improved motion contrast and processing efficiency in OCT angiography using complex-correlation algorithm

, , , , , , , and

Published 21 December 2015 © 2016 IOP Publishing Ltd
, , Citation Li Guo et al 2016 J. Opt. 18 025301 DOI 10.1088/2040-8978/18/2/025301

2040-8986/18/2/025301

Abstract

The complex-based OCT angiography (Angio-OCT) offers high motion contrast by combining both the intensity and phase information. However, due to involuntary bulk tissue motions, complex-valued OCT raw data are processed sequentially with different algorithms for correcting bulk image shifts (BISs), compensating global phase fluctuations (GPFs) and extracting flow signals. Such a complicated procedure results in massive computational load. To mitigate such a problem, in this work, we present an inter-frame complex-correlation (CC) algorithm. The CC algorithm is suitable for parallel processing of both flow signal extraction and BIS correction, and it does not need GPF compensation. This method provides high processing efficiency and shows superiority in motion contrast. The feasibility and performance of the proposed CC algorithm is demonstrated using both flow phantom and live animal experiments.

Export citation and abstract BibTeX RIS

1. Introduction

Given that many diseases are closely related to vascular abnormalities, a motion-contrast OCT angiography (Angio-OCT) has gained extensive attention [111]. By mathematically analyzing the temporal dynamics of light scattering, Angio-OCT is capable of contrasting the dynamic blood flow against the static tissue bed with high spatial resolution and motion sensitivity (down to capillary level) in a depth-resolved manner. Besides, Angio-OCT eliminates the requirement of exogenous contrast injection, and allows microvascular examination as frequently as required for clinical and scientific purposes.

To date, a variety of algorithms have been developed for extracting the flow signals [118]. Generally, each spatial position is OCT sampled several times with a certain time interval using repeated [6, 15, 17] or dense [16, 19] scanning protocols. Then the temporal changes in OCT intensity (or amplitude) [1, 46], phase [3, 13, 20] or complex-valued signals [9, 15, 18, 21] over such a time interval are analyzed with different processing algorithms, such as speckle variance [1, 4, 12], Doppler variance [9, 14, 16], phase variance [7, 13], differential calculation [11, 19] and correlation mapping [5, 22]. In consideration of the fact that the temporal dynamics can be amplified by increasing the time interval, an inter-frame analysis has been widely accepted for higher motion sensitivity [16]. In addition, compared with the methods which are purely based on the intensity or phase information, the complex-based method is potentially capable of a higher motion-contrast in theory for the combination of both the intensity and phase information. Hence, an inter-frame complex-based approach is particularly attractive for Angio-OCT [11, 15, 16, 18, 23, 24].

However, the motion-contrast Angio-OCT is inherently susceptible to the bulk tissue motions (BTMs), including global phase fluctuations (GPFs) and bulk image shifts (BISs) that are induced by involuntary cardiac and respiratory motions [17, 18, 25]. The influence of BTM is particularly serious in the inter-frame complex-based approach, due to the fact that the phase information is extremely sensitive to BTMs. The residual uncorrected BTM within an image sequence would generate motion artifacts. As reported in [18], in order to diminish motion artifacts in Angio-OCT, the complex-valued OCT data has to be processed sequentially by a cross-correlation algorithm for BIS correction, a phase operation for GPF compensation and then a time derivative (i.e. complex-differential (CD)) algorithm for flow signal extraction. Such a complicated procedure results in massive computational load, thus an efficient algorithm is desired for real-time processing and display.

The above-mentioned cross-correlation method is also suitable for the extraction of flow signals by evaluating the degree of similarity between adjacent B-frames [5, 22]. The correlation method is not involved with the static background artifacts, as pointed out in [19]. However, only the intensity (or amplitude) signals are adopted in the existing algorithms, i.e. intensity-correlation (IC) algorithm [5, 22], leading to a limited motion contrast.

In this study, as an extension of the IC algorithm, we present an inter-frame complex-correlation (CC) algorithm. Firstly, it is theoretically clarified that the CC algorithm is suitable for parallel processing of both flow signal extraction and BIS correction, and that it does not need GPF compensation. The motion contrast is compared with the existing IC algorithm. Then the feasibility and performance of the proposed CC method is demonstrated through the flow phantom and live animal experiments.

2. Materials and methods

2.1. System setup and scanning protocol

The imaging system used in this study was built based on the configuration of spectral domain OCT (SDOCT) as shown in figure 1. A broadband superluminescent diode (Superlum, Carrigtwohill, Ireland, Broadlighters D855-HP2) was used as the light source with a central wavelength of 850 nm and a full width at half maximum bandwidth of 100 nm, theoretically offering a high axial resolution of ∼3.2 μm in air. The output light was delivered into a 2 × 2 fiber coupler and then split into the reference arm and the sample arm. In the sample arm, an XY galvanometer was adopted for scanning a 3D volume, and an objective lens with the focal length of 40 mm was used to focus the probing light beam on the region of interest, yielding a measured lateral resolution of ∼15 μm. The OCT detection unit in our system was a high-speed spectrometer, equipped with a fast line-scan CMOS camera (Basler, Ahrensburg, Germany, Sprint spL4096-140k) providing a 120 kHz line-scan rate and 2048 active pixels. The spectrometer had a designed spectral resolution of 0.062 nm, providing an imaging range of ∼2.9 mm on each side of the zero delay line in air. The data processing unit in our system was a Dell Precision T5610 workstation installed with a single Intel Xeon Processor E5-2609 v2 and 16G RAM.

Figure 1.

Figure 1. Schematic of the imaging system used to collect the 3D spectral interferogram cube for OCT angiography. SLD: super luminescent diode; PC: polarization controller; CL: collimating lens; FL: focusing lens; OL: objective lens.

Standard image High-resolution image

As reported in [26], a general expression for the sensitivity of SDOCT is ${\rm{SNR}}=\eta {P}_{{\rm{s}}}\tau /{E}_{\upsilon },$ where $\eta $ is the quantum efficiency (∼38% in our system) of the detector, ${P}_{{\rm{s}}}$ is the sample arm power returning to the detection unit, $\tau $ is the integration time of detector (∼$8.4\;\mu {\rm{s}}$ in our system), and ${E}_{\nu }=hc/{\lambda }_{{\rm{c}}}$ (${\lambda }_{{\rm{c}}}$ is central wavelength of the light source, i.e. 850 nm in our system) is the photon energy. With an incident optical power of 2 mW on the perfect reflector sample surface, a theoretic signal-to-noise ratio (SNR) of ∼104 dB is predicted. The theoretic predication is in well agreement with the experiment measurement. As shown in figure 2, a system sensitivity of 100 dB was measured for the peak position at 0.5 mm with a calibrated sample arm attenuation of −52 dB. The experimental SNR in dB was calculated as twenty times the base-10 logarithm of the ratio of the A-scan peak amplitude to the standard deviation of the noise floor which was taken at the location of the A-scan peak by blocking the sample arm.

Figure 2.

Figure 2. SDOCT peaks from a calibrated −52 dB reflector at 0.25 mm increments.

Standard image High-resolution image

In this study, a step-scanning protocol with repeated B-scans was used to acquire 3D datasets. The volumetric scanning was composed of 512 A-scans in the fast-scan (X) direction and 200 steps in the slow-scan (Y) direction. Each 512 A-scans form one B-frame and five repeated B-frames were acquired at each step, resulting in a total of 1000 B-frames for each 3D dataset. The scanning range covers a 2 × 2 mm spatial distance in the X and Y directions. The B-frame rate was 190 frames s−1, determining a total acquisition time of ∼5.3 s for each volumetric dataset. The repeated scanning is designed to suppress the decorrelation noise in blood flow imaging [27].

2.2. Processing algorithm

The depth-resolved complex analytic OCT spatial signal is reconstructed by performing Fourier transform of the raw spectral interference fringe signal. At each step of scanning, the complex analytic signal of the nth repeated B-frame is denoted as ${\tilde{A}}_{n}(z,x)={A}_{n}(z,x)\mathrm{exp}[{\rm{i}}{\phi }_{n}(z,x)],$ where, n is the B-frame index, and the coordinates ($z,x)$ are indices along the Z and X directions.

2.2.1. Parallel processing of BIS correction and flow signal extraction

Large BTMs, particularly the respiratory motion, may result in BISs between adjacent B-frames, which consequently cause a global decorrelation and hinder the identification of the dynamic flow regions. In this study, we follow the assumptions that the BISs are rigid with no deformation and rotation, and out-of-plane motion is also limited for well pre-process and fixation of the sample.

The processing flowchart of CC-Angio-OCT algorithm is depicted in figure 3. At each scanning step, adjacent complex-valued B-frames ${\tilde{A}}_{n}(z,x)\;$ and ${\tilde{A}}_{n+1}(z,x)$ are paired. The paired B-frames are up-sampled using cubic spline interpolation in both Z (axial) and X (lateral) directions, and then analyzed by the normalized cross-correlation algorithm

Equation (1)

where, * means the complex conjugate. The frame index n ranges from 1 to 4 in this work. P and Q are the window sizes in the Z and X directions, respectively, and p and q are the corresponding indices of the pixels within the window. Moving the 2D window across the entire frame, a cross-sectional CC map is generated. In the CC map, the value of each pixel, as an index of decorrelation, lies in the range [0, 1].

Figure 3.

Figure 3. Processing flowchart of CC-Angio-OCT. Adjacent complex-valued B-frames ${\tilde{A}}_{n}(z,x)\;$ and ${\tilde{A}}_{n+1}(z,x)$ are paired, and CC mapping is performed for BIS correction and flow signal extraction in parallel. A binarized structural mask is produced to remove the noise-induced decorrelation artifacts. Finally, a CC angiogram is generated by combining both amplitude and phase information, but without GPF compensation. Up-sampling is to achieve a sub-pixel alignment and then down-sampling is to restore the original size of dataset.

Standard image High-resolution image

BISs lead to a global decorrelation between adjacent B-frames. In order to eliminate the effect of BISs, the mean value $\bar{{{\rm{CC}}}_{n}(z,x)}$ of all the pixels is treated as an index of the global decorrelation and sub-pixel alignment can be achieved by minimizing the mean value $\bar{{{\rm{CC}}}_{n}(z,x)}.$ In the aligned CC map, the areas with high decorrelation can be classified as the dynamic flow regions. Then the aligned CC map is down-sampled to its original sampling to generate a cross-sectional CC angiogram. The up-sampling performed here is to achieve an alignment precision of sub-pixel and the followed down-sampling is to restore the original size of dataset. In the proposed CC algorithm, the flow signal extraction is completed in parallel with the BIS correction.

Additionally, in cross-sectional CC angiograms, the static tissue background with weak OCT signals also exhibit high decorrelation due to the large influence of noise, which may be misclassified as dynamic blood flow. To remove the noise-induced decorrelation artifacts, a 'structural mask', which is generated by performing a binary threshold on the OCT amplitude signal ${A}_{n}(z,x),$ is applied to the CC map. The threshold value is empirically chosen to be above the mean background value.

With the trade-off among SNR, resolution and computation time, the up-sampling factor in this study takes the value four and the window size P and Q in equation (1) take the value 12. In each scanning step, five repeated B-scans are performed, producing a total of four cross-sectional angiograms. The four angiograms are averaged to improve SNR. A total of 200 cross-sectional angiograms are obtained to reconstruct a 3D map of vascular network.

2.2.2. Immunity against GPFs

Equation (1) can be further expressed as

Equation (2)

where, ${\rm{\Delta }}{\phi }_{n}(z,x)={\phi }_{n+1}(z,x)-{\phi }_{n}(z,x)$ is the phase difference between the paired B-frames. Small BTMs and random environmental vibrations result in GPFs. Thus, ${\rm{\Delta }}{\phi }_{n}(z,x)$ can be written as

Equation (3)

where, ${\rm{\Delta }}{\phi }_{n}^{{\rm{F}}}(z,x)={\phi }_{n+1}^{{\rm{F}}}(z,x)-{\phi }_{n}^{{\rm{F}}}(z,x)$ is the flow-induced phase change. ${\rm{\Delta }}{\phi }_{n}^{{\rm{GPF}}}(z,x)$ is the GPF-induced phase change. Following the assumptions used in [21], ${\rm{\Delta }}{\phi }_{n}^{{\rm{GPF}}}(z,x)$ is considered as a constant within a limited 2D window, of which the size is $P\times Q$ pixels in equation (2). Substituting equation (3) into equation (2), we have

Equation (4)

From the expression above, we can know that the CC mapping is insensitive to the GPFs.

In this study, CD-Angio-OCT algorithm is also conducted for comparison,

Equation (5)

Different from the CC method, the CD is sensitive to GPFs. The phase term ${\rm{\Delta }}{\phi }_{n}^{{\rm{GPF}}}(z,x)$ is used to compensate the GPF, which is determined by a histogram-based phase selecting algorithm [3, 17, 25, 28, 29]. The phase compensation is conducted A-line by A-line, and consequently the procedure is extremely time-consuming. Exactly, in our system for each 3D volumetric dataset (i.e. 2048 × 512 × 1000 pixels), the processing time of the CD algorithm is ∼4455 s, while that of the CC algorithm is ∼401 s which is about 10 times faster.

2.2.3. Superior motion contrast

The dynamic scattering of the moving blood cells causes changes in both the amplitude and phase of OCT signals. In the conventional IC method, only the amplitude information ${A}_{n}(z,x)$ is utilized for extracting the flow signal, which can be expressed by

Equation (6)

Comparing the CC and IC algorithms (i.e. equations (1) and (6)), the only difference is the numerator of the equations. For clarity, ${A}_{n}(z+p,x+q){A}_{n+1}(z+p,x+q)$ $\mathrm{exp}[j{\rm{\Delta }}{\phi }_{n}^{{\rm{F}}}(z+p,x+q)]$ is denoted by a vector $M(p,q).$ As illustrated in figure 4, due to the influence of phase, the length of the vector sum $|{\sum }^{}M|$ is not larger than the sum of the vector length ${\sum }^{}|M|.$ Then, we have

Equation (7)

In the static regions, the flow-induced phase change ${\rm{\Delta }}{\phi }_{n}^{{\rm{F}}}$ approximates to zero in theory, while in the dynamic regions, the phase change ${\rm{\Delta }}{\phi }_{n}^{{\rm{F}}}$ is not constant and exhibits spatial fluctuations (e.g. parabolic distribution) within the lumen in reality. According to equation (7), it is reasonable to conclude that, by combining both the amplitude and phase information, the CC-Angio-OCT offers a higher motion contrast than the IC method does.

2.3. Flow phantom and animal preparation

To validate the feasibility and performance of the proposed CC-Angio-OCT algorithm, both phantom and in vivo animal experiment were conducted.

Figure 4.

Figure 4. Illustration of the superior motion contrast of the CC to IC algorithms in the polar coordinate. The pictorial description of the length of the vector sum and the sum of the vector length in the (a) static region and (b) dynamic region, respectively.

Standard image High-resolution image

The flow phantom was made of an agarose gel mixed with ∼5% milk to mimic the static scattering tissue background and a capillary tube with an inner diameter of 0.5 mm was embedded in this tissue-like phantom. Three percent milk solution was pumped into the tube at a constant rate by syringe pump (KDS 100 series, Stoelting Co., Wood Dale, Illinois) to simulate the flowing blood. The phantom was placed on a translation stage, and BISs in the X and Z directions were manually introduced during OCT scanning.

C57BL/6 mice at 8–10 weeks of age were used in this experiment. Mice were anesthetized by intraperitoneal injection of 10% chloral hydrate (4 ml kg−1). The head was fixed in a stereotaxic frame (Stoelting, USA), and the scalp was retracted. The skull was thinned using a saline-cooled dental drill to generate a window of an area 3 mm × 3 mm and facilitate the optical penetration within the cortex at a wavelength of 850 nm. All animals were provided by the Experimental Animal Center and treated within the guidelines of the Institutional Animal Care and Use Committee of Zhejiang University.

3. Results

3.1. Phantom experiments

Figure 5 shows the results of the BIS correction and flow signal extraction in phantom experiments. Figure 5(a) is a representative cross-sectional OCT structural image of the phantom. The transparent tube can be visualized clearly. The regions inside and outside the tube represent the dynamic fluid and the static solid gel, respectively. Figure 5(b) shows the corresponding cross-sectional CC angiogram without BIS correction. Due to the BISs, both the static and dynamic regions exhibit high decorrection, and cannot be distinguished from each other. Using the sub-pixel alignment, sub-pixel shifts (3/4 pixel in the Z direction and 1/2 pixel in the X direction) are determined. The motion contrast appears in the aligned CC map without any further processing. Figure 5(c) presents the CC angiogram with the BIS correction conducted only in the Z direction, and figure 5(d) reports the result with the correction conducted in both Z and X directions.

Figure 5.

Figure 5. BIS correction and flow signal extraction in the phantom experiment. (a) Representative cross-sectional OCT structural image of the phantom. Cross-sectional CC angiogram (b) without BIS correction and after BIS correction (c) in the Z direction and (d) in both X and Z directions.

Standard image High-resolution image

With the BISs corrected, the immunity against GPFs of CC method was then tested in comparison with CD method. The left column in figure 6 presents the CD (a) and CC (b) angiograms without phase compensation, while the right one is the corresponding flow images after phase compensation. Comparing figure 6(b) with 6(d), it can be observed that the CD method is highly sensitive to GPFs. Due to the influence of GPFs, no motion contrast can be visualized before phase compensation in the CD method. In comparison, the CC method exhibits a high immunity against GPFs, referring to figures 6(a) and (c).

Figure 6.

Figure 6. Immunity against GPFs in the phantom experiment. Cross-sectional CC angiogram (a) without and (c) after phase compensation. Corresponding CD angiogram (b) without and (d) after phase compensation.

Standard image High-resolution image

Figure 7 presents a comparison of the motion contrast between the IC and CC methods. Figures 7(a) and (b) are the IC and CC angiograms, respectively. They are displayed in the same dynamic range and a motion-contrast improvement can be visualized in the CC flow image. The quantitative comparison is provided in figures 7(c) and (d). Figure 7(c) shows the representative signal profiles of the IC and CC methods, which are extracted from the same depth position indicated by the yellow dashed lines in figures 7(a) and (b), respectively. By comparison, although the CC profile exhibits a little higher background in the static part, the signals in flow part seem much more powerful. By calculating the mean value of the flow ($\bar{{\rm{Flow}}})$ and solid gel ($\bar{{\rm{Tissue}}})$ regions respectively, the motion contrast, which is defined as the ratio of $\bar{{\rm{Flow}}}$ and $\bar{{\rm{Tissue}}},$ can be quantified. Based on the statistical analysis of 200 flow images, the motion contrast of the CC method is much higher than that of the IC method (10.41 versus 6.64, ∼1.5 times) as reported in figure 7(d).

Figure 7.

Figure 7. Comparison of motion-contrast between IC- and CC-Angio-OCT using the phantom experiment. Representative cross-sectional (a) IC and (b) CC angiograms, respectively. (c) Signal profiles extracted from the same depth position marked by yellow dashed line in (a) and (b). (d) Error bar graphs representing mean and standard deviation of the contrast ratios of a total of 200 cross-sectional angiograms.

Standard image High-resolution image

3.2. In vivo experiments

Figure 8(a) is a representative cross-section of the OCT structural image, showing the depth-resolved morphological features of the mouse cerebral cortex, such as cranium and cortex. Figure 8(b) is the corresponding cross-sectional CC angiogram without phase compensation, in which the depth-resolved vasculature can be clearly visualized, as indicated by the arrows. Figures 8(c) and (d) are the cross-sectional CD angiograms before and after GPF compensation, respectively. It can be perceived that the CD angiogram is highly sensitive to the GPFs, and the residual static background is apparent in the cranium region.

Figure 8.

Figure 8. Immunity against GPFs in the mouse cortex in vivo. (a) Representative cross-sectional structural image. (b) Corresponding cross-sectional CC angiogram without compensation of GPFs. Cross-sectional CD angiograms (c) before and (d) after compensation of GPFs.

Standard image High-resolution image

Figure 9 presents a comparison of the motion-contrast between the IC and CC-Angio-OCT of the mouse cortex in vivo. Figures 9(a) and (b) are the representative cross-sectional angiograms using the IC and CC algorithms, respectively. They are displayed in the same dynamic range. Higher contrast of flow signals can be clearly visualized in the CC method. The signal profiles of the same flow area (indicated by the white arrows in figures 9(a) and (b)) are extracted along the same position (marked by the yellow dashed lines in figures 9(a) and (b)), as plotted in figure 9(c). The blood flow and static tissue regions can be distinguished. The motion contrast, i.e. the ratio of $\bar{{\rm{Flow}}}$ and $\bar{{\rm{Tissue}}},$ can be quantified. Totally, 100 flow areas are likewise selected randomly from 200 cross-sectional angiograms for statistical analysis, as plotted in figure 9(d). The mean values (mean ± standard deviation) of motion contrast are 48.78 ± 5.82 and 73.29 ± 5.54 in IC and CC methods, respectively, demonstrating ∼1.5 times improvement in motion contrast of the proposed CC method.

Figure 9.

Figure 9. Comparison of motion-contrast between IC- and CC-Angio-OCT using the mouse cortex in vivo. Representative cross-sectional (a) IC and (b) CC angiograms. (c) Signal profiles extracted from the same depth positions marked by the yellow dashed lines in (a) and (b). Peaks of the profiles correspond to the flow signals indicated by white arrows in (a) and (b). (d) Motion contrast distribution of 100 random flow signals with CC contrast value as the x-axis and IC contrast value as the y-axis.

Standard image High-resolution image

Figure 10 presents a 3D CC angiogram to demonstrate the feasibility of the CC algorithm for in vivo applications. Figure 10(a) is the 3D rendering of vasculature. The corresponding en face maximum intensity projection view is displayed in figure 10(b). Figure 10(c) shows a representative cross-sectional structural image overlaid with the corresponding blood flow signals, which was extracted from the position indicated by the yellow dashed line in figure 10(b).

Figure 10.

Figure 10. (a) 3D rendering of the mouse cortex vasculature. (b) Corresponding en face angiogram. (c) Representative structural cross-section overlaid with blood flow signals at the position marked by the yellow dashed line in (b).

Standard image High-resolution image

4. Discussion

Other groups of the inter-frame complex-based methods can be found in the Doppler variance [16] and the complex differential variance algorithms [24]. In these two algorithms, the extraction of the flow signals is mainly based on the variance computation [16, 24], and the global intensity fluctuations would induce static background artifacts. In contrast, the proposed CC algorithm is mainly based on the cross-correlation computation, and is relatively free from the static artifacts, as discussed in [19].

The proposed CC-Angio-OCT provides improved performance of motion contrast and bulk motion correction. Nevertheless, there still exist several limitations in the current study. The correlation-based (IC and CC) methods are prone to the noise-induced decorrelation artifacts in the low-SNR area. As shown in figure 11, due to the scattering attenuation, OCT SNR decreases gradually versus depth (referring to the curve in figure 11(b)). As the SNR decreases, the random noise would dominate in both the amplitude and phase of the complex-valued OCT signal. As depicted in figure 11(c), the CC values of static region increase more rapidly than those of dynamic region as the SNR decreases. When SNR is lower than ∼6 dB (as indicated by the red arrow), the static curve nearly overlaps the dynamic curve in large CC values, which implies that the low-SNR area would be misclassified as dynamic flow area, resulting in the noise-induced decorrelation artifacts. Similar phenomenon could also be observed in the IC curves. To remove such noise-induced decorrelation artifacts, a binary 'structural mask', as described in section 2.2.1, is thus overlaid to the IC or CC map. In contrast, the differential-based (CD) algorithms are less sensitive to this kind of noise, which can be spotted in figure 6(d) in comparison with 6(c).

Figure 11.

Figure 11. Influence of SNR on the performance of motion contrast. (a) Representative cross-sectional OCT structural images of the phantoms. The left part of the red vertical line is dynamic milk solution and the right part is static solid gel. (b) Changes of SNR versus depth. (c) Changes of mean CC, IC values versus SNR. Totally, 62 depth positions were evenly selected for SNR, CC and IC calculation between Depth 1 (190th pixel) and Depth 2 (800th pixel).

Standard image High-resolution image

In spite of its immunity to the GPFs, the proposed CC method is sensitive to the phase noise which is determined by the system sensitivity, i.e. $~1/\sqrt{{\rm{SNR}}}$ [30]. As shown in figure 11(c), in the high-SNR areas, both the IC and CC values of static region are very small and approximate zero, which indicates a similar low static background in both CC and IC angiograms (referring to equation (7)). However, as the SNR decreases, the CC values increase faster than IC values for the static region due to the influence of phase noise on the CC method, and in consequence, the CC angiogram exhibits a stronger background in the static tissue region, which is consistent with the results shown in figures 7 and 9. Fortunately, for the regions of interest (e.g. SNR above 15 dB), the influence of phase noise is limited and the CC method still offers a better performance of motion contrast than the IC method does by combining both the amplitude and phase information.

Mainly due to the limited speed of two-dimensional scanning, data acquisition, etc, of the OCT system, the correction of BISs in the current study are confined to inter-frame analysis with the assumption of no out-of-plane BTM. However, in reality, the BISs are generally 3D and the correction performance is constrained. With the advances in ultra-fast OCT [31], it is promising that the scanning protocol of stepped repeated B-scan implemented in this study could be substituted with that of repeated dense C-scan. Through inter-volume analysis of repeated volumetric datasets, a more efficient 3D correction of BISs becomes possible. In the meantime, the extraction of blood flow signals can also be extended to three dimensions. Since the time interval between adjacent B-frames within a single C-volume is sufficiently short in ultra-fast OCT, GPFs could be regarded as a constant within each small volume element and then a 3D spatial window may be applied in the CC method for OCT angiography, which has potential to provide higher motion contrast compared with our current study.

5. Conclusion

In this study, we propose an inter-frame CC-Angio-OCT with improved motion contrast and higher processing efficiency. Specifically, the proposed method is capable of correcting BISs in parallel with the extraction of flow signals on the basis of the same mathematical model of correlation-based algorithm. Although the proposed method is phase-sensitive, it is free of GPF artifacts, which further releases the computation load of complicated GPF compensation for each A-line. In addition, with combination of both amplitude and phase information of the moving blood cells, the CC method presents approximately 1.5 times greater motion contrast in angiograms than the IC approach does. The feasibility and performance of the proposed method have been demonstrated by imaging with a tissue-like flow phantom as well as by visualizing the microvasculature of the mouse cerebral cortex in vivo. As the ultra-fast OCT technology matures, we believe that the performance of the proposed CC-Angio-OCT would highly benefit from the incorporation of 3D inter-volume analysis within the current study.

Acknowledgments

We acknowledge financial supports from National Natural Science Foundation of China (61475143, 11404285, 61335003, 61327007 and 61275196), Zhejiang Province Science and Technology Grant (2015C33108), Zhejiang Provincial Natural Science Foundation of China (LY14F050007), National Hi-Tech Research and Development Program of China (2015AA020515), Fundamental Research Funds for the Central Universities (2014QNA5017), and Scientific Research Foundation for Returned Scholars, Ministry of Education of China.

Please wait… references are loading.
10.1088/2040-8978/18/2/025301