Optimization parameters of the heart pump design

, and

Published under licence by IOP Publishing Ltd
, , Citation V Markučič et al 2018 IOP Conf. Ser.: Mater. Sci. Eng. 393 012125 DOI 10.1088/1757-899X/393/1/012125

1757-899X/393/1/012125

Abstract

In this paper, research of optimization parameters of continuous flow heart pump is presented. The application of centrifugal pumps as heart assist devices imposes design limitations on the geometry of the heart pump. Geometry and pump parameters affect the performance and the hemocompatibility of the heart pump. The main quality assessment factor for heart pump is the pump hemocompatibility i.e., the amount of mechanical damage caused by a pump on blood cells. Besides stagnation zones and recirculation zones, wall shear stress is parameter that is used to predict pump hemocompatibility. Second important factor is minimal volume of heart pump with acceptable anatomical fitting. Additional factors are high efficiency and durability. The aim of the research is to propose the optimal design of bladeless centrifugal heart pump. The dimensionless optimization parameters of the heart pump design are derived from Navier - Stokes equation. In conclusion, dimensionless optimization parameters of bladeless centrifugal continuous flow heart pump are presented.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1757-899X/393/1/012125