This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.

The MiniPET: a didactic PET system

, , , , and

Published 4 March 2013 Published under licence by IOP Publishing Ltd
, , Citation R Pedro et al 2013 JINST 8 C03005 DOI 10.1088/1748-0221/8/03/C03005

1748-0221/8/03/C03005

Abstract

The MiniPET project aims to design and build a small PET system. It consists of two 4 × 4 matrices of 16 LYSO scintillator crystals and two PMTs with 16 channels resulting in a low cost system with the essential functionality of a clinical PET instrument. It is designed to illustrate the physics of the PET technique and to provide a didactic platform for the training of students and nuclear imaging professionals as well as for scientific outreach. The PET modules can be configured to test for the coincidence of 511 keV gamma rays. The model has a flexible mechanical setup [1] and can simulate 14 diferent ring geometries, from a configuration with as few as 18 detectors per ring (ring radius ϕ=51 mm), up to a geometry with 70 detectors per ring (ϕ=200 mm). A second version of the electronic system [2] allowed measurement and recording of the energy deposited in 4 detector channels by photons from a 137Cs radioactive source and by photons resulting of the annihilation of positrons from a 22Na radioactive source. These energy spectra are used for detector performance studies, as well as angular dependency studies. In this paper, the mechanical setup, the front-end high-speed analog electronics, the digital acquisition and control electronics implemented in a FPGA, as well as the data-transfer interface between the FPGA board and a host PC are described. Recent preliminary results obtained with the 4 active channels in the prototype are also presented.

Export citation and abstract BibTeX RIS

Please wait… references are loading.