This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.

Simulation study of a highly efficient, high resolution X-ray sensor based on self-organizing aluminum oxide

, , , , , and

Published 26 January 2012 Published under licence by IOP Publishing Ltd
, , Citation J Muehlbauer et al 2012 JINST 7 C01095 DOI 10.1088/1748-0221/7/01/C01095

1748-0221/7/01/C01095

Abstract

State of the art X-ray imaging sensors comprise a trade-off between the achievable efficiency and the spatial resolution. To overcome such limitations, the use of structured and scintillator filled aluminum oxide (AlOx) matrices has been investigated. We used Monte-Carlo (MC) X-ray simulations to determine the X-ray imaging quality of these AlOx matrices. Important factors which influence the behavior of the matrices are: filling factor (surface ratio between channels and `closed` AlOx), channel diameter, aspect ratio, filling material etc. Therefore we modeled the porous AlOx matrix in several different ways with the MC X-ray simulation tool ROSI [1] and evaluated its properties to investigate the achievable performance at different X-ray spectra, with different filling materials (i.e. scintillators) and varying channel height and pixel readout. In this paper we focus on the quantum efficiency, the spatial resolution and image homogeneity.

Export citation and abstract BibTeX RIS

Please wait… references are loading.
10.1088/1748-0221/7/01/C01095