Abstract

In the presented work, we investigated several digital methods of a discrimination signals from fast neutrons and gamma quanta. The experimental setup consists of a Pu-Be neutron source, a scintillation detector with an organic para-terphenyl monocrystal, and a digitizer (CAEN DT5730, 500 MS/s). Mixed waveform sequences were stored and then separated by pulse shape. Four methods were used for signals separation. Comparison of the traditional and the new methods of Figure of Merit (FOM) calculation is given. FOM = 1.5 was obtained in our setup for the minimum threshold value. A scintillation detector with a para-terphenyl crystal was used to measure neutron yield in the neutron generator with carbon nanotubes.