This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.
Paper The following article is Open access

Optimal operation of a 4-element injection lens in a hemispherical spectrograph: FDM/BEM simulation and experimental demonstration

, , , , and

Published under licence by IOP Publishing Ltd
, , Citation G Martínez et al 2015 J. Phys.: Conf. Ser. 635 052083 DOI 10.1088/1742-6596/635/5/052083

1742-6596/635/5/052083

Synopsis

We have investigated the voltage settings for the 4-element injection lens of a biased paracentric hemispherical deflection analyzer (HDA) with virtual entry aperture using numerical methods. The two lens electrode voltages were varied as free parameters while the electron optical properties of the analyzer were calculated from trajectories with test initial distributions for different pre-retardation factors in an effort to obtain improved energy resolution. The resulting lens voltages were then also tested on the existing HDA spectrometer at the Tandem Accelerator Laboratory, while simulated and measured line profile characteristics were compared, particularly at the best resolution working points.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/635/5/052083