STEM probe characteristics at large defoci for use in ptychographical imaging

, and

Published under licence by IOP Publishing Ltd
, , Citation K M Atkinson et al 2008 J. Phys.: Conf. Ser. 126 012092 DOI 10.1088/1742-6596/126/1/012092

1742-6596/126/1/012092

Abstract

A possible configuration for undertaking diffractive imaging microscopy in the conventional scanning transmission electron microscope (STEM) is to defocus the probe by a large distance in order to minimise the number of diffraction patterns required to reconstruct the ptychographical image. Although the characteristics of STEM probes have been well explored and measured near the beam crossover, they are rarely observed (or calculated) at very large defoci. In this paper we compare probes calculated under a variety of approximations with measured data from a JEOL 2010F in STEM mode. When the illumination aperture is narrow and the defocus is large (near parallel illumination useful for ptychographical imaging) the Ronchigram cannot be used easily to characterize aberration, We develop an alternative method of estimating the effective aperture size and condenser lens spherical aberration.