This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.
Brought to you by:
Paper

Direct reconstruction of the source intensity distribution of a clinical linear accelerator using a maximum likelihood expectation maximization algorithm

, , and

Published 13 January 2016 © 2016 Institute of Physics and Engineering in Medicine
, , Citation P Papaconstadopoulos et al 2016 Phys. Med. Biol. 61 1078 DOI 10.1088/0031-9155/61/3/1078

0031-9155/61/3/1078

Abstract

Direct determination of the source intensity distribution of clinical linear accelerators is still a challenging problem for small field beam modeling. Current techniques most often involve special equipment and are difficult to implement in the clinic. In this work we present a maximum-likelihood expectation-maximization (MLEM) approach to the source reconstruction problem utilizing small fields and a simple experimental set-up. The MLEM algorithm iteratively ray-traces photons from the source plane to the exit plane and extracts corrections based on photon fluence profile measurements. The photon fluence profiles were determined by dose profile film measurements in air using a high density thin foil as build-up material and an appropriate point spread function (PSF). The effect of other beam parameters and scatter sources was minimized by using the smallest field size ($0.5\times 0.5$ cm2). The source occlusion effect was reproduced by estimating the position of the collimating jaws during this process. The method was first benchmarked against simulations for a range of typical accelerator source sizes. The sources were reconstructed with an accuracy better than 0.12 mm in the full width at half maximum (FWHM) to the respective electron sources incident on the target. The estimated jaw positions agreed within 0.2 mm with the expected values. The reconstruction technique was also tested against measurements on a Varian Novalis Tx linear accelerator and compared to a previously commissioned Monte Carlo model. The reconstructed FWHM of the source agreed within 0.03 mm and 0.11 mm to the commissioned electron source in the crossplane and inplane orientations respectively. The impact of the jaw positioning, experimental and PSF uncertainties on the reconstructed source distribution was evaluated with the former presenting the dominant effect.

Export citation and abstract BibTeX RIS

1. Introduction

Advanced radiotherapy techniques, such as stereotactic radiosurgery (SRS) and intensity modulated radiation therapy (IMRT), commonly utilize small radiation fields (<$2\times 2$ cm2) in order to improve target coverage and reduce the dose to surrounding organs at risk (OAR). However, accurate commissioning of beam models for small fields appears still to be challenging. Most detectors often present significant perturbations in small fields due to volume averaging or the presence of high density materials. During beam model commissioning of linear accelerators, these perturbations may result in mis-configuration of crucial beam parameters in the treatment planning system (TPS). Systematic errors introduced at this step will eventually propagate as dosimetric errors at the patient dose calculation stage.

In order to address this issue, one approach is to apply detector-specific correction factors directly on the measurements (Alfonso et al 2008). Several researchers have shown that corrections for output factors can be derived by Monte Carlo (MC) methods (Cranmer-Sargison et al 2011, Francescon et al 2012). Similar corrections may be needed for dose profiles as well, especially in the tail region (Papaconstadopoulos et al 2014a, Francescon et al 2014). The MC-derived correction factors have been validated by experimental methods (Pantelis et al 2012). However, deriving such corrections is a lengthy process and requires MC expertise as well as detailed knowledge of the design and materials of the detector. Furthermore, each time a new detector is produced, new correction factors are needed. It is clear that alternative pathways for determining the appropriate beam parameters would offer a great service in the dosimetry of small fields.

One of the most important beam parameters is the source size and shape of clinical linear accelerators. In MC beam models, the term source most often refers to the electron spatial distribution incident on the target. In other beam models, it may refer to the bremsstrahlung radiation that is produced in the target, often referred to as the primary x-ray source or focal-spot. In both cases, the source size is most often characterized by the full width at half maximum (FWHM) and the shape is assumed to be a 2D elliptical Gaussian distribution. Sterpin et al (2011) showed that for a nominal 5.5 and 18.0 MeV incident electron energy, the primary x-ray source is located approximately at a depth of 0.13 and 0.25 mm respectively from the top of the target. It was also shown in that study that the photon source FWHM agrees within 0.06 mm to the electron source FWHM for typical source sizes of 0.5–1.5 mm. This important conclusion implies that if direct reconstruction techniques could be developed for the x-ray source, then the electron source could be reproduced as well within the previously stated level of accuracy. Other beam parameters, such as the energy of the incident electron source, have a limited effect in small fields. Furthermore, methods for determining the incident electron energy or even unfolding the full energy spectrum of linear accelerators have been suggested (Ali et al 2012).

As the field becomes small, the primary x-ray source, as seen from the point of measurement, starts to be partially occluded (Aspradakis and IPEM 2010). Photons originating from other scatter sources, such as the flattening filter and primary collimator (extra-focal spot), are almost completely blocked and their dosimetric impact is minimized. The magnitude of the source occlusion effect not only depends on the spatial extent of the primary x-ray source, but also on the exact position of the collimators that define the field (Scott et al 2009). Thus, any model attempting to reproduce the source occlusion effect should reconstruct both parameters.

Several methods have been suggested for measuring the x-ray source directly. These methods may involve a multiple slit γ camera (Lutz et al 1988), determining the inverse Abel transformation of the derivative of fluence profiles (Treuer et al 1993), output ratios (Zhu et al 1995) or a micro-leaf collimator (MLC) (Treuer et al 2003). Perhaps the most widely known and accurate technique is using a rotating slit collimator and a CT reconstruction algorithm (Munro et al 1988, Jaffray et al 1993, Caprile and Hartmann 2009). Other researchers applied a similar method using a moving slit (Loewenthal et al 1992, Sham et al 2008). However, the experimental set-up of a slit collimator technique is difficult to implement and perform routinely in a clinical environment. Furthermore, Chen et al (2011) exhibited that the accuracy of this technique depends strongly on the choice of slit width, height and distance to the source.

An interesting finding by some of the previous investigations was that the x-ray source may not follow a Gaussian functional form (Sham et al (2008) and Chen et al (2011)). Furthermore, the source distribution may vary with time (Jaffray et al 1993) or as currents in the magnets of the electron beam are altered (Munro et al 1988). Despite the above facts, there is currently no requirement for the physicist to directly measure the source as part of the TPS commissioning or of the linac quality assurance procedure.

The purpose of this work is to suggest a method, including a model and an experimental procedure, that would allow the direct reconstruction of the primary x-ray source distribution as generated in the target using clinical measurements. The model is based on iteratively ray-tracing photons from the target to the measurement plane and vice-versa using a maximum-likelihood expectation-maximization (MLEM) algorithm. No prior assumptions on the source distribution or size are needed. The experimental procedure involves the determination of photon fluence profiles in small fields in air using film measurements and a thin lead (Pb) foil as a build-up material. Blurring effects due to the non-zero electron range and photon scatter are taken into account by convolution with pre-calculated point spread functions (PSF). The method also estimates the appropriate collimator settings that reproduce the source occlusion effect for the given accelerator geometry and measurement set. The accuracy of the method in reconstructing the correct source size and shape is first evaluated by performing MC simulations of the experimental set-up for a range of electron sources and then experimentally by performing measurements on a linear accelerator with known MC beam model parameters.

2. Methods

2.1. The inverse problem

Assuming that the photon fluence distribution can be measured at the bottom of the linear accelerator (exit plane, figure 1(a)), the question we are seeking to answer is how the source fluence distribution can be reconstructed at the top of the target (source plane, figure 1(a)). This question defines an imaging problem to derive an object representation from an initial blurred image. This is referred to as the inverse problem in image reconstruction and iterative methods are commonly applied to address it. In the following, the method will be explained as applied in the source reconstruction problem. In this work we will only consider 1D fluence distributions and the reconstruction is performed on the crossplane and inplane orientations separately.

Figure 1.

Figure 1. (a) Modeling the geometrical source occlusion by the system matrix. Rays passing through the collimation aperture (blue solid lines) and reach the exit plane are assigned '1', while rays completely blocked (red dashed line) are assigned '0' on the corresponding positions on the system matrix. (b) The experimental set-up for the measurement of the photon fluence in air. Dimensions in figures are not to scale.

Standard image High-resolution image

2.2. Extracting the system matrix

As a first step, the relationship between each pixel on the source plane needs to be associated, by geometric means, to each pixel on the exit plane. Let's assume the case of a source pixel j (figure 1(a)). Photons originating from pixel j can potentially reach any of the exit pixels $\left\{i,i+1,\ldots,i+k\right\}$ that are visible through the collimation aperture. Rays that exceed the collimation limits and are incident on a collimation block are assumed to be completely absorbed. This relationship can be expressed numerically by a matrix, commonly known in imaging as the system matrix. Each column of the matrix is associated to a source pixel, while each row to an exit pixel. Thus, in the previous example the column j will be assigned '0' everywhere except rows $\left\{i,i+1,\ldots,i+k\right\}$ , where '1' is assigned. The process is then repeated for all source pixels. The system matrix essentially models geometrically the source occlusion effect, pixel-by-pixel, for the particular accelerator design. In this work the system matrix had a resolution of 0.01 mm on both the source and exit plane.

2.3. The MLEM reconstruction algorithm

After the system matrix is derived it can be used to reconstruct the image of a given source distribution. The source fluence can be iteratively determined for a given measured exit fluence by the MLEM reconstruction technique (Shepp and Vardi 1982, Reader and Verhaeghe 2014). The following steps summarize the process:

  • (i)  
    First, make an arbitrary estimation of the source distribution vector. For example, assume ${{\varphi}_{j}}=1$ , for all j (uniform distribution).
  • (ii)  
    Ray-trace the photons forward to the exit plane to determine the expected fluence distribution vector on the exit plane, $\boldsymbol{q}$ . To do so, multiply the system matrix $\mathbf{A}$ with the latest source distribution vector estimation $\boldsymbol{\varphi }$ .
    Equation (1)
  • (iii)  
    Derive the exit fluence profile correction vector $\mathbf{k}$ by dividing, pixel-by-pixel, the actual fluence measurements $\mathbf{m}$ by the latest estimation of the exit fluence profile $\mathbf{q}$
    Equation (2)
  • (iv)  
    Ray-trace the correction ratios back to the source plane to determine the expected fluence corrections, $\mathbf{c}$ , on the source plane. To do so, multiply the transpose system matrix ${{\mathbf{A}}^{T}}$ with the latest profile correction $\mathbf{k}$ on the exit plane
    Equation (3)
  • (v)  
    Normalize each source pixel correction, cj, to the number of exit pixels contributing to this pixel by dividing element-by-element with the sensitivity image, $\mathbf{s}={{\mathbf{A}}^{\mathbf{T}}}\mathbf{1}$
    Equation (4)
  • (vi)  
    Finally, derive a new estimation of the source distribution vector $\boldsymbol{\varphi }^{{\mathbf{new}}}$ by multiplying, pixel-by-pixel, the latest source estimation $\boldsymbol{\varphi }$ with the normalized source profile correction vector ${{\mathbf{c}}^{n}}$
    Equation (5)
    Steps (ii)–(vi) are repeated iteratively until some stopping criterion is met. In this work, we assume that the algorithm converges if the FWHM of the source has not changed more than 2% in the last 30 iterations. In a condensed form the MLEM algorithm can be written as:
    Equation (6)
    where vector–vector multiplications and divisions are performed element-by-element.

2.4. Experimental set-up

The experimental set-up used in this work can be seen in figure 1(b). To measure the photon fluence at the exit plane of the linear accelerator dose profile measurements were performed in air using radiochromic film. The measurements were performed on a Varian Novalis Tx linear accelerator (Varian Medical Systems, Palo Alto, CA, USA) at the SRS mode (6 MV). The contribution of backscattered radiation was minimized by placing the films on a 5 cm block of styrofoam. Monte Carlo simulations were performed to verify that the styrofoam block does not change the dose profile distribution. To reduce the impact of contaminant electrons incident on the surface and the blurring due to electron range, a Pb foil of 2 mm thickness was used as a build up material. The thickness of the build-up was chosen based on the average electron energy reaching the phantom surface. The Source to Surface Distance (SSD) was set to 105 cm. In order to maximize the sensitivity to the primary x-ray source and reduce the impact of other beam parameters or scatter sources, the measurements were performed solely in the smallest field size of $0.5\times 0.5$ cm2. The field size was defined by the secondary collimators (jaws) as projected to a SSD of 100 cm. The top surface of the inplane (Y) and crossplane (X) jaws reside at 28.0 cm and at 36.7 cm respectively from the top of the target.

2.5. Blurring in the Pb foil

Even though the high density Pb foil reduces the blurring, due to the smaller electron range, it does not eliminate the effect. Furthermore, photon scattering in the phantom will contribute in penumbra broadening. This blurring can be modeled by a point spread function, $\text{PSF}(x)$ . The relative dose in water after the Pb foil, ${{\left({{D}_{w}}\right)}_{pb}}(x)$ , can then be expressed as a convolution operation between the relative photon fluence, $\Phi(x)$ and the $\text{PSF}(x)$ (equation (7)).

Equation (7)

To extract the optimum PSF, an optimization procedure based on MC simulations was followed. First, a Pearson VII functional form was chosen for parameterizing the PSF (equation (8)).

Equation (8)

As a second step, MC simulations using the accurate accelerator model were performed to calculate the relative photon fluence, $\Phi(x)$ , and relative dose in water after the Pb foil, ${{\left({{D}_{w}}\right)}_{pb}}(x)$ . The values of γ and n varied in a systematic manner following a brute-force approach. For each set of estimated parameter values ($\hat{\gamma},\hat{n}$ ) the dose profile was deconvolved with the PSF(x; $\hat{\gamma},\hat{n}$ ), using a maximum likelihood algorithm, to extract an estimation of the photon fluence, $\rm{\hat \Phi}\left(x;\hat{\gamma},\hat{n}\right)$ . The optimum set of parameter values ${{\left(\hat{\gamma},\hat{n}\right)}_{\text{opt}}}$ were determined by minimizing the mean square relative error (MSRE) between estimated and MC calculated photon fluence (equation (9)).

Equation (9)

The above procedure was also performed for other PSF models, including Gaussian and Lorentzian functions, which presented inferior performance than the Pearson VII. The sensitivity of the extracted PSF to the electron source size used in the MC simulation was evaluated by repeating the above procedure for electron source FWHM values of 0.5, 1.0, 1.5 and 2.0 mm and calculating the average PSF and standard deviation (1 σ). The blurring effect was inherently included in the model by convolving each column of the system matrix with the average PSF.

2.6. Determining the collimator jaw position

The exact jaw position at the time of measurements defines the projected field size and thus affects the system matrix. In order to determine the jaw position for a nominal projected field size of $0.5\times 0.5$ cm2 at a SSD of 100 cm, the collimator position varied from 0.4 to 0.6 cm with a step of 0.01 cm. For each jaw position the system matrix was re-calculated and the source reconstruction repeated. The jaw position for which the extracted system matrix minimized the mean local error between calculated and measured dose profiles in the 90%–10% dose range was selected as the closest to the true value.

2.7. Film measurements

The dose profile measurements were performed using Gafchromic EBT3 film (GAFCHROMIC, International Specialty Products, Wayne, NJ). The film calibration was performed using the red channel in the region of 0–2 Gy and the reflection scanning mode. This calibration protocol has shown to increase the sensitivity and reduce the uncertainties in the low dose regions (Papaconstadopoulos et al 2014b), such as those presented in the profile penumbra. A set of 5 film irradiations were performed. The dimensions of each film piece were $6.35\times 5.08$ cm2. Before each irradiation the jaws were repositioned to a nominal $0.5\times 0.5$ cm2, in order to include mechanical jaw positioning errors in the uncertainty analysis. An Epson Expression 11 000XL (Epson Seiko Corporation, Nagano, Japan) document scanner was used for the scans in reflection mode. The scanning resolution was set to 127 dpi (0.2 mm/pixel). The profiles were re-sampled to a resolution of 0.01 mm using a cubic spline interpolation method.

2.8. Monte Carlo simulations

An accurate model of the Varian Novalis Tx was used for performing the MC simulations of the electron and photon transport in the accelerator using the EGSnrc/BEAMnrc user code (Rogers et al 1995). An elliptical Gaussian distribution was chosen as the electron source incident on the target. The field sizes were defined by the secondary collimators (jaws) as projected to a SSD of 100 cm. The model was commissioned for small fields in previous work (Papaconstadopoulos et al 2014a). The commissioning process resulted at the time in an electron source of FWHM equal to 1.25 mm and 1.10 mm for the crossplane (X) and inplane (Y) orientations respectively. The optimal jaw positions were found to define a projected field side of 4.7 mm and 4.9 mm for the crossplane and inplane orientations respectively. The commissioned model presented a local dose accuracy level within 3% for dose profiles and within 1.5% for output factors compared to measurements in small and large fields. For the dose calculation part the EGSnrc/DOSXYZnrc code was used (Walters et al 2011). To extract the photon fluence distribution a phase-space file was saved at the bottom of the accelerator at a SSD of 105 cm. To simulate the experimental set-up, a Pb foil of 2 mm thickness was included in the accelerator model at a SSD of 105 cm. For all dose calculations the voxel dimensions were set to $1\times 1\times 1$ mm3. For the accelerator and dose calculation simulation part the electron cut-off (ECUT) values were set to 700 keV and 521 keV respectively. The choice of a lower ECUT value for the DOSXYZnrc simulation was made in order to increase the accuracy of dose deposition during electron transport in small voxels. The photon cut-off (PCUT) value was set to 10 keV in all cases.

2.9. Method evaluation

The ability of the method to reconstruct the correct source size and shape was first benchmarked against MC simulations of the experimental set-up of known Gaussian electron sources. The collimating jaws were kept to the commissioned values. The calculated profiles ${{\left({{D}_{w}}\right)}_{Pb}}$ were first re-sampled to a resolution of 0.01 mm and then used as an input to the reconstruction algorithm. The reconstructed FWHM and jaw positions were then directly compared to the expected values.

Since some widening of the electron source may exist due to electron scattering in the target, the photon source distribution at a depth of 0.2 mm from the top of the target was also reported. The depth was chosen based on the results reported by Sterpin et al (2011) and taking into account that the incident energy was higher in this work (6.1 instead of 5.5 MeV).

In order to evaluate the accuracy of the reconstruction method, the reconstructed source distribution was directly compared to the Gaussian electron source distribution that was used as an input to the MC calculation. The reconstructed source was also compared to the photon source distribution. In addition, a Gaussian fit was performed on the reconstructed source and the normalized root mean square error, RMSE (%), between the fitted Gaussian and reconstructed source was reported. The RMSE was calculated in the 100–5% intensity region and normalized to the mean value. The above process was repeated for electron sources of FWHM equal to 0.5, 1.0, 1.5 and 2.0 mm. At this step, the incident electron source on the target is known to be Gaussian during the MC simulations. Thus, the RMSE values evaluate the performance of the method in reconstructing the expected functional form. The reconstruction was also repeated for different SSD levels of 105, 125 and 150 cm with the PSFs recalculated on each SSD level. The electron source FWHM was set to the commissioned values ($1.25\times 1.10$ mm).

As a second step the measured film profiles were used as an input to the algorithm. The reconstructed source was compared to the electron and photon sources that were previously determined during MC model commissioning for small fields, by inspecting the normalized RMSE (%) between the reconstructed source and a Gaussian fit, the FWHM and the Full Width at Tenth of Maximum (FWTM). It should be noted that in this case possible deviations from a Gaussian functional form may be attributed not only to the limitations of the algorithm, but also to the proximity of the source to a Gaussian distribution.

Finally, the reconstructed FWHM of the source and of the jaw positions were used as input source parameters to the MC beam model. The calculated dose profiles of the experimental set-up were directly compared to the respective measurements in the crossplane and inplane orientations.

2.10. Uncertainty analysis

The uncertainties related to the reconstructed source distribution and the FWHM and FWTM metrics can be classified in 3 main components: (i) Jaw positioning, related to possible misestimation of the actual jaw position by the algorithm, which would result in an erroneous system matrix. The total standard deviation (1 σ) in jaw positioning was estimated to be the jaw displacement that resulted in a mean local error between calculated and measured profiles less than 10% in the 90–10% dose region. In this work this displacement was found to be about 0.2 mm. (ii) Experimental, related to measurement uncertainties. The total experimental uncertainty was derived by summing in quadrature the standard deviation (1 σ) of the 5 repeated film measurements and the film calibration fitting uncertainties. The mechanical accelerator jaw repositioning is also included in this component. (iii) PSF, related to possible misestimation of the PSF due to source variations. The PSF uncertainty was calculated as the standard deviation (1 σ) of the PSFs extracted for source sizes of 0.5, 1.0, 1.5 and 2.0 mm.

Each uncertainty component was propagated to the source as follows: first a deviation from the mean of a specific effect (e.g. jaw position) was sampled from a Gaussian distribution, while the rest remained constant (e.g. experimental and PSF). The Gaussian distribution had a standard deviation equal to the effect's estimated standard deviation. For each sampled point the source was reconstructed using the MLEM algorithm. Another point was then sampled and the source reconstructed again. The procedure continued until the standard deviation of all reconstructed FWHM values appeared not to vary more than 1%. The total uncertainty was calculated by randomly sampling a deviation from the mean of all effects (e.g. jaw position, experimental and PSF) at the same time and repeating the previous procedure.

3. Results

The calculated dose profiles after 2 mm of Pb, before and after deconvolution with the respective PSF, are presented in figure 2 for the crossplane and inplane orientation. In order to evaluate the accuracy of the kernel in debluring the profile from the electron and photon scattering effects, the profiles are directly compared to the expected photon fluence at a SSD of 105 cm. The calculations were performed using the MC accelerator model for the commissioned values of the electron source FWHM and jaw positions. Figure 3 presents the average PSF for electron sources of FWHM equal to 0.5, 1.0, 1.5 and 2.0 mm along with the 1 standard deviation uncertainty level. The PSFs were calculated using the methodology presented in section 2.5. The average PSF had a FWHM (δFWHM) equal to 0.88 (0.05) mm and 0.85 (0.13) mm at the crossplane and inplane orientations respectively.

Figure 2.

Figure 2. Monte Carlo dose calculations of the crossplane (a) and inplane (b) profiles after 2 mm of Pb in air ($\Box $ ) and after 2 mm of Pb in air deconvolved using the respective PSF kernel ($\bigtriangleup $ ). The dose profiles are directly compared to the calculated photon fluence (x).

Standard image High-resolution image
Figure 3.

Figure 3. The average point spread function after 2 mm of Pb (solid line) for the crossplane (a) and inplane (b) orientations. The standard deviation (1 σ) due to source variations is presented with the dashed lines.

Standard image High-resolution image

The reconstructed sources using the MC calculated dose profiles of the experimental set-up as an input are presented in figure 4. The calculations were performed for a range of typical electron Gaussian sources of FWHM equal to 0.5–2 mm. The respective photon sources at a depth of 0.2 mm in the target are also presented. The Gaussian fits to the reconstructed source distributions exhibited a RMSE (%) in the range of 2.3%–5.4% and of 2.2%–4.1% for the crossplane and inplane orientations respectively. The relative intensity differences between the reconstructed sources and the electron and photon sources (figure 5) presented a local agreement within 10% in all cases with the major discrepancies occurring in regions of high intensity gradient and for source sizes of FWHM equal to 0.5 and 1 mm.

Figure 4.

Figure 4. Reconstructed source distributions using the MC calculated dose profiles of the experimental set-up as input. Calculations were repeated for electron sources of FWHM equal to 0.5, 1.0, 1.5 and 2.0 mm. Reconstructed sources are directly compared to the expected Gaussian electron (•) and photon ($\bigtriangleup $ ) sources.

Standard image High-resolution image
Figure 5.

Figure 5. Relative local intensity difference between the reconstructed source distribution and the respective MC electron source (symbols) and photon source (lines) at a depth of 0.2 mm in the target for FWHM of 0.5, 1.0, 1.5 and 2.0 mm.

Standard image High-resolution image

The ability of the MLEM algorithm to reconstruct the FWHM of the expected source and field size is presented in table 1. The reconstructed FWHM agrees within 0.12 mm and 0.10 mm to the electron source FWHM (incident on the target) and photon source FWHM (at 0.2 mm depth in the target) respectively. The photon source appears broader than the electron source by 0.02–0.04 mm. The reconstructed jaw positions reproduce the expected values with an accuracy better or equal to 0.2 mm.

Table 1. Reconstructed FWHM of the source and reconstructed field size (crossplane (X)  ×  inplane (Y)) using the MC calculated dose profiles as input.

e source (mm2) γ source (mm2) MLEM source (mm2) Field (mm2)
$0.5\times 0.5$ $0.54\times 0.54$ $0.54\times 0.56$ $4.7\times 4.9$
$1.0\times 1.0$ $1.03\times 1.03$ $1.10\times 1.08$ $4.6\times 4.9$
$1.5\times 1.5$ $1.53\times 1.53$ $1.52\times 1.52$ $4.7\times 4.9$
$2.0\times 2.0$ $2.03\times 2.02$ $2.04\times 2.12$ $4.8\times 4.7$

Note: The expected FWHM of the electron source (incident on the target) and photon source (at 0.2 mm depth in the target) are presented. The expected field size was set to the commissioned values ($4.7\times 4.9$ mm2).

To evaluate the accuracy of the method to reproduce the expected dose profile, the reconstructed dose profiles are directly compared to the MC calculated dose profiles that were used as an input (figure 6). Discrepancies are observed mainly for the smallest source size (FWHM  =  0.5 mm) and in the tail region with a local dose difference reaching 17.7% in the 90–10% dose region.

Figure 6.

Figure 6. Reconstructed dose profiles (solid lines) and MC calculated dose profiles of the experimental set-up (dashed lines). Calculations were repeated for sources of FWHM equal to 0.5, 1.0, 1.5 and 2 mm.

Standard image High-resolution image

Table 2 evaluates the sensitivity of the method to the choice of the exit plane location. The reconstruction was performed for different SSD values of 105, 125 and 150 cm. The MC electron source and field size were set in this case to the commissioned values of $1.25\times 1.10$ mm2 and $4.7\times 4.9$ mm2 respectively. The reconstructed source FWHM agreed within 0.04 mm and the reconstructed jaw settings within 0.2 mm.

Table 2. Reconstructed FWHM of the source and reconstructed field size (crossplane (X)  ×  inplane (Y)) for SSD values of 105, 125 and 150 cm.

SSD (cm) MLEM source (mm) Field (mm2)
105 $1.22\times 1.08$ $4.6\times 4.9$
125 $1.20\times 1.08$ $4.8\times 5.0$
150 $1.24\times 1.12$ $4.8\times 5.0$

Note: The expected source and field size were set to the commissioned values ($1.25\times 1.10$ mm2 and $4.7\times 4.9$ mm2).

The reconstructed source using the film dose profile measurements at the Varian Novalis accelerator is presented in figure 7. In the same figure the Gaussian electron source and the respective photon source, as they were determined during commissioning, are also presented. The reconstructed source exhibited a FWHM of 1.22 mm (±0.12) and 1.21 mm (±0.11) at the crossplane and inplane orientation, respectively. The RMSE (%) of a Gaussian fit was found to be 2.4% and 2.7% for the crossplane and inplane orientations, respectively. The relative local intensity differences between reconstructed source and photon and electron sources are presented in figures 8(a) and 8(b) in the crossplane and inplane orientations respectively. The reconstructed field side for this set of measurements was 4.4 mm on both orientations.

Figure 7.

Figure 7. MLEM reconstructed source distributions using the film profile measurements as input (solid lines). The Gaussian electron (•) and photon ($\bigtriangleup $ ) sources are also presented as they were determined during commissioning. The 1 standard deviation (1 σ) total uncertainty level is presented with dashed lines.

Standard image High-resolution image
Figure 8.

Figure 8. Relative local intensity difference between the reconstructed source distribution and the MC electron (solid line) and photon Gaussian sources (dashed line) as determined during commissioning.

Standard image High-resolution image

The propagated effects of the jaw positioning, experimental and PSF uncertainties on the reconstructed source are presented in figure 9. Overall, the jaw positioning uncertainties resulted in the most significant source variations, while PSF and experimental uncertainties mainly affected the tail region. Table 3 summarizes the reconstructed source parameters and uncertainty components.

Table 3. FWHM and FWTM values of the reconstructed source using the film profile measurements, the MC electron source (incident on the target) and the photon source (at 0.2 mm depth in the target) as determined during model commissioning.

  FWHMx (mm) FWTMx (mm) FWHMy (mm) FWTMy (mm)
e source 1.25 2.26 1.10 2.00
γ source 1.28 2.32 1.13 2.04
rec source 1.22 2.29 1.21 2.32
${{\sigma}_{\text{total}}}$ (rec source) 0.12 0.27 0.11 0.20
${{\sigma}_{\text{jaw}}}$ (rec source) 0.11 0.21 0.10 0.15
${{\sigma}_{\text{exp}}}$ (rec source) 0.03 0.12 0.04 0.11
${{\sigma}_{\text{psf}}}$ (rec source) 0.04 0.16 0.01 0.07

Note: The total and component uncertainties of the reconstruction are presented at the 1 σ level.

Figure 9.

Figure 9. Uncertainty components of the experimentally reconstructed source (solid lines) using the film measurements. Uncertainties include the effects of jaw positioning ((a),(b)), experimental ((c),(d)) and PSF ((e),(f)) (dashed lines).

Standard image High-resolution image

Finally, the dose profiles were recalculated using the accelerator MC beam model with the electron Gaussian source set to the reconstructed FWHM value of $1.22\times 1.21$ mm2 and field size of $4.4\times 4.4$ mm2 (figure 10). The dose distributions exhibited an agreement with the film measurements of 1.2% in the 90–10% dose region.

Figure 10.

Figure 10. MC calculated crossplane and inplane dose profiles using the MLEM reconstructed source and jaw parameters (•) compared to the film measurements (lines).

Standard image High-resolution image

4. Discussion

Measuring the photon fluence in air using a 2 mm Pb foil as build-up material proved to be challenging and some penumbra broadening still appeared in the dose region below 40%. For that purpose, the use of a PSF function to account for the electron blurring and photon scattering in the build-up material appeared to be important. In fact, performing the MLEM reconstruction without using the PSF, resulted in an over-estimation of the source FWHM up to 0.42 mm.

The PSFs, extracted by the MC calculations and the optimization procedure (section 2.5), presented minimal sensitivity to the source size selected during the simulations (figure 3). This result ensures that the method can be applied to any accelerator of the same geometry without prior knowledge of the source. The observed variations in the PSF distributions resulted in only small variations of the reconstructed source as can be seen in figures 9(e) and (f). Thus, a pre-calculated average PSF can be provided to the user or even included in the system matrix. In the latter case the blurring is inherently included in the model and the end user only needs to input the measured dose profiles. It should be emphasized that different accelerator designs may present different PSFs. Further research will be needed to verify the applicability of a universal PSF for all accelerator designs.

The MLEM reconstruction using the MC calculated dose profiles with known electron Gaussian sources reproduced the expected Gaussian shape with RMSE values up to 5.4% and 4.1% in the crossplane and inplane orientations respectively. The maximum deviations were observed for the smallest source with FWHM of 0.5 mm. The discrepancy is also depicted in the tail and shoulder regions of the reconstructed dose profiles upon convergence of the MLEM algorithm. The difficulty in reconstructing the source shape in this case could potentially be attributed to the steep change of the dose gradient.

The reconstructed source appears to overestimate the FWHM in most of the cases relative to the expected electron Gaussian source incident on the target (figures 4, 5 and table 1). This can be partly explained by the electron blurring occurring in the target. The photon source at 0.2 mm depth in the target indeed appeared broader up to 0.04 mm. This result is in agreement with the work by Sterpin et al (2011) that reported a broadening up to 0.06 mm for a similar incident energy. This finding implies that the reconstruction algorithm actually reconstructs a distribution closer to the photon source and not to the electron source, even though both are in close agreement to each other.

A second reason may be related to the observation that the overestimation of the source was often coupled with an underestimation of the projected field side. In fact in the cases that the correct field side was reproduced (FWHM of 0.5 mm and 1.5 mm in table 1), the MLEM reconstructed intensity distribution presented an exceptional agreement with the photon intensity distribution (figures 4 and 5) and the FWHM was within 0.02 mm to the expected value. The overall accuracy in the jaw position estimation was found to be 0.2 mm, a result which agrees with the initially estimated precision. The above observations emphasize the need for accurate reconstruction of both the source distribution and jaw position in order to properly predict the source occlusion effect. It should be noted that for the largest source size (FWHM  =  2 mm) the uncertainties in jaw positioning estimation increased and different combinations of source size/jaw positions could still provide acceptable agreement between reconstructed and expected dose profiles.

The experimentally reconstructed source reproduced a Gaussian shape with RMSE (%) values up to 2.7%. The results are within the range of the RMSE (%) values reported previously using the MC simulations of known Gaussian electron sources. Relative to the previously commissioned electron source the reconstructed FWHM was  −0.03 mm lower and  +0.11 mm higher in the crossplane and inplane orientations respectively. Relative to the corresponding photon source the reconstructed FWHM was  −0.06 mm lower and  +0.08 mm higher in the crossplane and inplane orientations. In both cases, the reconstructed source agrees within the estimated total uncertainty level (1 σ), even though only marginally for the inplane orientation (figure 7). More importantly, if the reconstructed source FWHM and field size are used as an input to the MC accelerator beam model, an excellent agreement was observed between measured and calculated dose profiles (figure 10).

The uncertainty component analysis, presented in figure 9 and table 3, indicates that the estimation of the jaw position is the major source of uncertainty during reconstruction. A 0.2 mm misestimation of the collimator position would result in about 0.10–0.11 mm misestimation of the source FWHM. This, in turn, would result in output factor variations of about 1.5–3.0% for typical source sizes in the range of 0.8–1.4 mm and penumbra width (80–20%) variations of 3–4%. This level of accuracy still competes with most small field detectors that are currently used in the clinic. The development of independent methods for estimating the correct jaw positions would greatly improve the performance of the source reconstruction algorithm. Uncertainties due to PSF variations with the source and experimental measurements, including mechanical jaw positioning, were less significant in the prediction of the FWHM. However, these effects presented a much more significant impact on the tail regions of the source distribution as illustrated in the reported FWTM uncertainties (table 3).

5. Conclusion

In this work, we investigated the performance of a MLEM-based source reconstruction technique for clinical linear accelerators using small field photon fluence profiles. The use of a high density build-up material along with an appropriate PSF were found important for extracting accurately the photon fluence. The PSFs exhibited overall a minimal dependence on the source. The model was able to reconstruct the electron source with an overall accuracy of 0.12 mm for typical source sizes with FWHM ranging from 0.5 to 2 mm that were tested in simulations. Experimentally, the method exhibited an overall accuracy level of 0.11 mm with respect to a previously commissioned MC model, with the most significant uncertainty attributed to the estimation of the jaw position at the time of measurements. The results of this study indicate that MLEM-based approaches could be a powerful and practical tool for the direct reconstruction of the source distribution, without any prior assumptions. In the future, such techniques could become part of a quality assurance procedure that would evaluate potential source variations through time and machine usage or between different accelerator types.

Acknowledgments

PP would like to acknowledge Dr A Ahnesjö, Dr A Reader and S Lee for important conversations on the subject as well as Compute Canada/Calcul Quebec for providing computing resources. PP gratefully acknowledges the financial support by the A S Onassis Public Benefit Foundation in Greece and by the CREATE Medical Physics Research Training Network grant of the Natural Sciences and Engineering Research Council (Grant number: 432290). IL acknowledges research funding from the Research Institute of the McGill University Health Centre, the Montreal General Hospital Foundation, and the Phil Gold Fellowship.

Please wait… references are loading.
10.1088/0031-9155/61/3/1078