This site uses cookies. By continuing to use this site you agree to our use of cookies. To find out more, see our Privacy and Cookies policy.

Spectrophotometric studies of ultra low loss optical glasses I: single beam method

and

Published under licence by IOP Publishing Ltd
, , Citation K C Kao and T W Davies 1968 J. Phys. E: Sci. Instrum. 1 1063 DOI 10.1088/0022-3735/1/11/303

0022-3735/1/11/1063

Abstract

A single beam spectrophotometer has been developed for the measurement of spectral attenuation coefficients of very low loss optical glasses in the wavelength range 400-1000 nm. The instrument has an accuracy to ±0·00005 cm−1 when sample thicknesses of 20 cm are used. The basic stability of the instrument and the sources and effects of systematic errors have been investigated in detail. Some results are presented in the form of attenuation curves for four different low-loss optical glasses.

Export citation and abstract BibTeX RIS