Characterization of large area avalanche photodiodes in X-ray and VUV-light detection

L M P Fernandes1, F D Amaro1, A Antognini2, J M R Cardoso1, C A N Conde1, O Huot3, P E Knowles3, F Kottmann4, J A M Lopes1, L Ludhova3,4, C M B Monteiro1, F Mulhauser3, R Pohl2,4, J M F dos Santos1, L A Schaller3, D Taqqu4 and J F C A Veloso5

Show affiliations


The present manuscript reviews our R&D studies on the application of large area avalanche photodiodes (LAAPDs) to the detection of X-rays and vacuum ultraviolet (VUV) light. The operational characteristics of LAAPDs manufactured by Advanced Photonix Inc. were investigated for X-ray detection at room temperature. The optimum energy resolution obtained in four LAAPDs investigated was found to be in the range 10-18% for 5.9 keV X-rays. The observed variations are associated with dark current differences between the several prototypes. LAAPDs have demonstrated high counting rate capability (up to about 105/s) and applicability in diverse areas, mainly low-energy X-ray detection, where LAAPDs selected for low dark current may achieve better performance than proportional counters. LAAPDs were also investigated as VUV photosensors, presenting advantages compared to photomultiplier tubes. X-rays are often used as a reference in light measurements; this may be compromised by the non-linearity between gains measured for X-rays and VUV-light. The gain was found to be lower for X-rays than for VUV light, especially at higher bias voltages. For 5.9 keV X-rays, gain variations of 10% and 6% were measured relative to VUV light produced in argon ( ~ 128 nm) and xenon ( ~ 172 nm) for gains of about 200. The effect of temperature on the LAAPD performance was investigated for X-ray and VUV-light detection. Gain variations of more than -4% per oC were measured for 5.9 keV X-rays for gains above 200, while for VUV light variations are larger than -5% per oC. The energy resolution was found to improve with decreasing temperature, what is mainly attributed to dark current. The excess noise factor, another contribution to the energy resolution, was experimentally determined and found to be independent of temperature, increasing linearly with gain, from 1.8 to 2.3 for a 50-300 gain range. The LAAPD response under intense magnetic fields up to 5 Tesla was investigated. While for X-ray detection the APD response practically does not vary with the magnetic field, for 172 nm VUV light a significant amplitude reduction of more than 20% was observed.

Keywords

Photon detectors for UV, visible and IR photons (solid-state)

X-ray detectors

Spectrometers

PACS

85.60.Dw Photodiodes; phototransistors; photoresistors

85.60.Bt Optoelectronic device characterization, design, and modeling

Subjects

Electronics and devices

Optics, quantum optics and lasers

Dates

Issue 08 (August 2007)

Received 14 February 2007, accepted for publication 26 July 2007

Published 17 August 2007



Related review articles

What's this?
View review articles related to this research to gain an insight into the key trends in this subject area. Related review articles are selected based on PACS/MSC codes, and are no more than three years old.

  1. Advances in solid state photon detectors
  2. Nonlocal impact ionization and avalanche multiplication

View by subject




Export








Please login to access our web services, or create an account if you don't yet have one.

You must have cookies enabled in your web browser to be able to login.

Username
Password

Forgotten your password? Get a new one here.