Modeling indirect detectors for performance optimization of a digital mammographic detector for dual energy applications

, , , , , , , , and

Published under licence by IOP Publishing Ltd
, , Citation N Martini et al 2015 J. Phys.: Conf. Ser. 574 012075 DOI 10.1088/1742-6596/574/1/012075

1742-6596/574/1/012075

Abstract

Dual Energy imaging is a promising method for visualizing masses and microcalcifications in digital mammography. The advent of two X-ray energies (low and high) requires a suitable detector. The scope of this work is to determine optimum detector parameters for dual energy applications. The detector was modeled through the linear cascaded (LCS) theory. It was assumed that a phosphor material was coupled to a CMOS photodetector (indirect detection). The pixel size was 22.5 μm. The phosphor thickness was allowed to vary between 20mg/cm2 and 160mg/cm2 The phosphor materials examined where Gd2O2S:Tb and Gd2O2S:Eu. Two Tungsten (W) anode X-ray spectra at 35 kV (filtered with 100 μm Palladium (Pd)) and 70 kV (filtered with 800 pm Ytterbium (Yb)), corresponding to low and high energy respectively, were considered to be incident on the detector. For each combination the contrast- to-noise ratio (CNR) and the detector optical gain (DOG), showing the sensitivity of the detector, were calculated. The 40 mg/cm2 and 70 mg/cm2 Gd2O2S:Tb exhibited the higher DOG values for the low and high energy correspondingly. Higher CNR between microcalcification and mammary gland exhibited the 70mg/cm2 and the 100mg/cm2 Gd2O2S:Tb for the low and the high energy correspondingly.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.

Please wait… references are loading.
10.1088/1742-6596/574/1/012075